Fluke Sewing Machine fluke123 User Manual

®
Fluke 123  
Industrial ScopeMeter  
Users Manual  
4822 872 00743  
January 1997, Rev. 4, 7/99  
© 1997, 1999 Fluke Corporation. All rights reserved. Printed in the Netherlands.  
All product names are trademarks of their respective companies.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SERVICE CENTERS  
To locate an authorized service center, visit us on the World Wide Web:  
or call Fluke using any of the phone numbers listed below:  
+1-888-993-5853 in U.S.A. and Canada  
+31-402-678-200 in Europe  
+1-425-356-5500 from other countries  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of Contents  
Chapter  
1
Title  
Declaration of Conformity ..............................................................................................  
Unpacking the Test Tool Kit ..........................................................................................  
Safely Using the Test Tool ............................................................................................  
Page  
1
2
4
Using The Test Tool ....................................................................................................  
7
Goal of this Chapter.......................................................................................................  
Powering the Test Tool..................................................................................................  
Resetting the Test Tool .................................................................................................  
Changing Backlight........................................................................................................  
7
7
8
9
Reading the Screen....................................................................................................... 10  
Making Selections in a Menu......................................................................................... 11  
Looking at the Measurement Connections .................................................................... 12  
Input A ...................................................................................................................... 12  
Input B ...................................................................................................................... 12  
COM ........................................................................................................................ 12  
i
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Displaying an Unknown Signal with Connect-and-View™ ............................................. 13  
Making Measurements................................................................................................... 14  
Freezing the Screen....................................................................................................... 16  
Holding a Stable Reading .............................................................................................. 16  
Making Relative Measurements ..................................................................................... 17  
Selecting Auto/Manual Ranges ...................................................................................... 18  
Changing the Graphic Representation on the Screen.................................................... 18  
Changing the Amplitude ............................................................................................ 18  
Changing the Time Base ........................................................................................... 18  
Positioning the Waveform on the Screen .................................................................. 19  
Smoothing the Waveform .......................................................................................... 20  
Displaying the Envelope of a Waveform.................................................................... 21  
TrendPlotting a Waveform.............................................................................................. 22  
Starting a TrendPlot™ function ................................................................................. 22  
Changing the TrendPlot Reading .............................................................................. 23  
Turning Off the TrendPlot Display ............................................................................. 23  
Acquiring the Waveform ................................................................................................. 23  
Making a Single Acquisition....................................................................................... 23  
Recording Slow Signals over a Long Period of Time ................................................ 25  
Selecting AC-Coupling .............................................................................................. 26  
Reversing the Polarity of the Displayed Waveform ................................................... 26  
Triggering on a Waveform.............................................................................................. 27  
Setting Trigger Level and Slope ................................................................................ 27  
Selecting the Trigger Parameters.............................................................................. 28  
Isolated Triggering..................................................................................................... 29  
Triggering on Video Signals ...................................................................................... 29  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Contents (continued)  
Triggering on a Specific Video Line .......................................................................... 31  
Saving and Recalling a Setup or a Screen .................................................................... 32  
Saving Screens......................................................................................................... 32  
Recalling Screens..................................................................................................... 33  
Deleting Screens ...................................................................................................... 33  
Using a Printer............................................................................................................... 34  
Using FlukeView® Software ........................................................................................... 36  
2
Maintaining the Test Tool ........................................................................................... 37  
About this Chapter......................................................................................................... 37  
Cleaning the Test Tool .................................................................................................. 37  
Storing the Test Tool ..................................................................................................... 37  
Charging the Ni-Cd Battery Pack .................................................................................. 38  
Keeping Batteries in Optimal Condition......................................................................... 39  
Replacing and Disposing of the Ni-Cd Battery Pack ..................................................... 40  
Using and Adjusting 10:1 Scope Probes ....................................................................... 41  
Calibrating the Test Tool ............................................................................................... 43  
Parts and Accessories................................................................................................... 43  
Service Manual......................................................................................................... 43  
Standard Accessories............................................................................................... 43  
Optional Accessories ................................................................................................ 46  
3
Tips and Troubleshooting .......................................................................................... 47  
Goal of this Chapter....................................................................................................... 47  
Using the Tilt Stand ....................................................................................................... 47  
Resetting the Test Tool ................................................................................................. 48  
Changing the Information Language ............................................................................. 48  
Changing the Display .................................................................................................... 49  
Adjusting the Screen Contrast .................................................................................. 49  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Setting the Grid Display............................................................................................. 49  
Changing Date and Time ............................................................................................... 50  
Saving Battery Life ......................................................................................................... 51  
Setting the Power Down Timer.................................................................................. 51  
Changing the Auto Set Options...................................................................................... 52  
Using Proper Grounding................................................................................................. 53  
Solving Printing and Other Communication Errors......................................................... 54  
4
Specifications............................................................................................................... 55  
Introduction .................................................................................................................... 55  
Dual Input Oscilloscope.................................................................................................. 56  
Vertical....................................................................................................................... 56  
Horizontal .................................................................................................................. 57  
Trigger ....................................................................................................................... 57  
Advanced Scope Functions....................................................................................... 58  
Dual Input Meter............................................................................................................. 59  
Input A and Input B.................................................................................................... 59  
Input A ....................................................................................................................... 60  
Advanced Meter Functions........................................................................................ 61  
Miscellaneous................................................................................................................. 62  
Environmental ................................................................................................................ 63  
Safety............................................................................................................................. 64  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Declaration of Conformity  
Sample tests  
Declaration of Conformity  
Standards used:  
for  
Fluke 123  
EN 61010.1 (1993)  
Safety Requirements for Electrical Equipment for  
Measurement, Control, and Laboratory Use  
ScopeMeter® test tool  
Manufacturer  
EN 50081-1 (1992)  
Electromagnetic Compatibility.  
Generic Emission Standard:  
EN55022 and EN60555-2  
Fluke Industrial B.V.  
Lelyweg 1  
7602 EA Almelo  
The Netherlands  
EN 50082-2 (1992)  
Electromagnetic Compatibility.  
Generic Immunity Standard:  
IEC1000-4 -2, -3, -4, -5  
Statement of Conformity  
Based on test results using appropriate standards,  
the product is in conformity with  
Electromagnetic Compatibility Directive 89/336/EEC  
Low Voltage Directive 73/23/EEC  
The tests have been performed in a  
typical configuration.  
This Conformity is indicated by the symbol  
,
i.e. “Conformité européenne”.  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Note  
Unpacking the Test Tool Kit  
When new, the rechargeable Ni-Cd battery pack  
is not fully charged. See Chapter 2.  
The following items are included in your test tool kit. (see  
Figure 1.):  
#
Description  
Model  
123  
l
123/SCC  
1
2
Industrial ScopeMeter Test Tool  
Ni-Cd Battery Pack (installed)  
l
l
l
3
Power Adapter/Battery Charger  
Shielded Test Leads (red and gray) with black ground leads  
Test Lead (black)  
l
l
4
l
l
5
l
l
6
Hook Clips (red and gray)  
l
l
7
Alligator Clips (red, gray, and black)  
Banana-to-BNC Adapter(s) (Black)  
Fluke 123 Users Manual (this book)  
Product Registration Card with Envelope  
Shipment box  
l
l
8
l (1x)  
l (2x)  
l
9
l
l
l
10  
11  
12  
13  
14  
l
Optically Isolated RS-232 Adapter/Cable  
FlukeView® ScopeMeter® Software for Windows®  
Hard Carrying Case  
l
l
l
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Unpacking the Test Tool Kit  
Figure 1. ScopeMeter Test Tool Kit  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
See explanation in  
manual  
Equal potential  
inputs  
Safely Using the Test Tool  
Attention  
Disposal information  
Earth  
Carefully read the following safety  
information before using the test tool.  
Recycling  
Conformité  
Safety Precautions  
information  
Européenne  
Specific warning and caution statements, where they  
apply, will be found throughout the manual.  
Double Insulation  
(Protection Class)  
UL 3111 listed  
UL 1244 listed  
A Caution identifies conditions and actions that may  
damage the test tool.  
UL1244  
Warning  
A Warning identifies conditions and actions that pose  
hazard(s) to the user.  
Should this test tool be used with AC  
coupling selected, or manual operation of  
amplitude or time base ranges, the measuring  
results displayed on the screen may not be  
representative of the total signal. This can  
result in the presence of dangerous voltages  
of more than 42V peak (30V rms), not being  
detected. To guarantee user safety, all  
signals should first be measured with DC  
coupling selected and in fully automatic  
mode. This ensures that the full signal  
content is measured.  
Symbols used on the test tool and in this manual are  
explained in the next table.  
Warning  
To avoid electrical shock, use only Fluke  
power supply, Model PM8907 (Power  
Adapter/Battery Charger).  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Safely Using the Test Tool  
The terms ‘Isolated’ or ‘Electrically floating’ are used in  
this manual to indicate a measurement in which the test  
tool COM (common, also called ground) is connected to a  
voltage different from earth ground.  
Warning  
Do the following to avoid electrical shock or  
fire if a test tool  
common input is  
connected to more than 42V peak (30V rms):  
The term “Grounded” is used in this manual to indicate a  
measurement in which the test tool COM (common) is  
connected to an earth ground potential. For more  
information about proper grounding, see Chapter 3.  
Use only test leads and test lead  
adapters supplied with the test tool. (or  
equivalents as specified in the accessory  
list, see Chapter 2.)  
The test tool COM (common) inputs (red INPUT A shield,  
gray INPUT B shield, and black 4-mm banana jack) are  
connected internally via self-recovering fault protection.  
The input connectors have no exposed metal and are fully  
insulated to protect against electrical shock. The black  
4 mm banana jack COM (common) can be connected to a  
voltage above earth ground for isolated (electrically  
floating) measurements and is rated up to 600V rms  
above earth ground.  
Do not use conventional exposed metal  
banana plug connectors.  
Use only one common connection to the  
test tool.  
Remove all test leads that are not in use.  
Use 600V (or more) rated and marked test  
lead adapters. The maximum allowable  
input voltage is 600V.  
Connect the power adapter to the AC  
outlet before connecting it to the test  
tool.  
Do not insert metal objects in the power  
adapter connector.  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
If Safety-Precautions are Impaired  
Use of the test tool in a manner not specified may  
impair the protection provided by the equipment.  
Before use, inspect the test leads for mechanical damage  
and replace damaged test leads!  
Whenever it is likely that safety has been impaired, the  
test tool must be turned off and disconnected from the line  
power. The matter should then be referred to qualified  
personnel. Safety is likely to be impaired if, for example,  
the test tool fails to perform the intended measurements  
or shows visible damage.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 1  
Using The Test Tool  
Goal of this Chapter  
This Chapter provides a step-by-step introduction to the  
test tool. The introduction does not cover all of the  
capabilities of the test tool but gives basic examples to  
show how to use the menus perform basic operations.  
Powering the Test Tool  
Follow the procedure (step 1 to 3) in Figure 1-1 to power  
the test tool from a standard ac outlet. See Chapter 2 for  
battery power instructions.  
Turn the test tool on.  
The test tool powers up in its last setup configuration.  
Figure 1-1. Powering the Test Tool  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Resetting the Test Tool  
If you want to restore the test tool settings as delivered from  
the factory, do the following:  

ô
í
Turn the test tool off.  
Press and hold.  
Press and release.  
The test tool turns on, and you should hear a double  
beep, indicating the Reset was successful.  
÷
Release.  
Now look at the display; you will see a screen that looks  
like Figure 1-2.  
Figure 1-2. The Screen After Reset  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Changing Backlight  
1
Changing Backlight  
After power-up, the screen has a high bright display.  
To save battery power, the screen has an economic  
brightness display when operated on the battery pack (no  
power adapter connected).  
To change the brightness of the display , do the following:  

ô
Dim the backlight.  
Brighten the backlight again.  
The high brightness increases when you connect the  
power adapter.  
Note  
Using dimmed display lengthens maximum  
battery power operation time by about one hour.  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Reading the Screen  
The screen is divided into three areas: Reading area,  
Waveform area, and Menu area. Refer to Figure 1-3  
during the following.  
Reading area (A): Displays the numeric readings.  
Because only input A is on, you will see the input A  
readings only.  
Waveform area (B): Displays the input A waveform. The  
bottom line displays the ranges/div and the power  
indicator (line or battery). Because only input A is on, you  
will see the input A waveform only.  
Note  
When battery powered, the battery indicator  
informs you about the condition of the battery  
Figure 1-3. The Screen Area’s  
from full to empty:  
.
Menu area (C): Displays the menu that provides choices  
available through the blue function keys.  
When you change a setup, a part of the screen is used to  
display the choices. The area displays one or more menus  
with choices accessed with the arrow keys:  
.
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Making Selections in a Menu  
1
Figure 1-4 shows the basic navigation of the test tool.  
Making Selections in a Menu  
Subsequently follow steps to ÷to open a menu and to  
choose an item.  

Press the SCOPE MENU key to  
open the Scope menu.  
Note  
Pressing the SCOPE MENU key a second time  
closes this menu and resumes normal  
measurement. This toggling enables you to  
check the menu without destroying your settings.  
Figure 1-4. Basic Navigation  
Note  
ô
í
÷
Use the blue arrow keys to  
highlight the item.  
When you do not change an item with the blue  
arrow keys, repeatedly pressing  
enables  
you to step through a menu without changing the  
setup of the test tool.  
Press the blue ‘ENTER’ function  
key to confirm selection  
Press ‘ENTER’ until you return to  
normal mode.  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Looking at the Measurement  
Connections  
Look at the top of the test tool. The test tool provides two  
4-mm safety shielded banana jack inputs (red input A and  
gray input B) and a safety 4-mm banana jack input (COM).  
(See Figure 1-5.)  
Input A  
Figure 1-5. Measurement Connections  
You can always use the red input A for all single input  
measurements possible with the test tool.  
Input B  
For measurements on two different signals you can use  
the gray input B together with the red input A .  
COM  
You can use the black COMmon as single ground for low  
frequency measurements, and for Continuity, Ohm (),  
Diode, and Capacitance measurements.  
Warning  
To avoid electrical shock or fire, use only one  
COM  
(common) connection, or ensure that  
all connections to COM  
potential.  
are at the same  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Displaying an Unknown Signal with Connect-and View™  
1
Displaying an Unknown Signal with  
Connect-and View™  
The Connect-and-View™ function enables hands-off  
operation to display complex unknown signals. This  
function optimizes the position, range, time base, and  
triggering and assures a stable display on nearly all  
waveforms. If the signal changes, the setup will track  
these changes.  
To enable the Connect-and-View™ function, do the  
following:  
Connect the red test lead from red input A to the  
unknown signal to be measured .  
Perform an Auto Set.  
In the next example, the screen displays 1.411” in large  
numbers and “-0.103” in smaller numbers. A scope trace  
gives a graphical representation of the waveform.  
Figure 1-6. The Screen After an Auto Set  
The trace identifier (A) is visible on left of the waveform  
area. The zero icon (-) identifies the ground level of the  
waveform.  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Making Measurements  
The reading area displays the numeric readings of the  
chosen measurements on the waveform that is applied to  
the input jack.  
First connect the red shielded test lead from input A,  
and the gray shielded test lead from input B to the  
signals to be measured. Connect the short ground  
leads to the same ground potential. (See Figure 1-7.)  
Note  
For Ohm (), continuity, diode, and capacitance  
measurements, use the red shielded test lead  
from input A and the black unshielded ground  
lead from COM (common). (See Figure 1-7.)  
Figure 1-7. Measurement Setup  
ô
í
Highlight Hz.  
To choose a frequency measurement for Input A, do the  
following:  

Open the INPUT A menu.  
Select Hz.  
Observe that Hz is now the main reading. The former  
main reading has now moved to the smaller secondary  
reading position. (See Figure 1-8.)  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Making Measurements  
1
To choose also a Peak-to-Peak measurement for Input B,  
do the following:  
Highlight PEAK-PEAK.  
Accept the pk-pk measurement.  
ø
ù
Open the INPUT B menu.  

Now, you will see a screen like Figure 1-8.  
Highlight ON.  
ô
Turn Input B on. Observe that the  
highlight jumps to the present  
main measurement.  
í
Highlight PEAK...  
÷
Open the PEAK submenu.  
û
Figure 1-8. Hz and Vpp as Main Readings  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Freezing the Screen  
Holding a Stable Reading  
You can freeze the screen (all readings and waveforms)  
at any time.  
The Touch Hold® function captures and freezes the next  
stable measurement result. A beep indicates that a stable  
measurement has been made.  

Freeze the screen. HOLD  
appears at the bottom of the  
reading area.  
Use the following procedure for the Touch Hold function:  

ô
Open the INPUT A menu.  
ô
Resume your measurement.  
TOUCH HOLD OFF appears on  
bottom of the screen.  
í
÷
Measure the signal.  
Wait until an audible beep: now  
you have a stable display.  
BEEP)))  
The screen continues to update  
with valid readings (and beeps)  
as longs as you maintain the  
measurement connections.  
Because no special keys accompany the Touch Hold  
function, you can use this function for hands-free  
measurements.  
û
Return to normal measurement.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Making Relative Measurements  
1
Making Relative Measurements  
Zero Reference displays the present measurement result  
with respect to the defined value. This feature is useful  
when you need to monitor the measured value in relation  
to a known good value.  

ô
Open the INPUT A menu.  
Open the METER A OPTIONS  
submenu.  
Figure 1-9. Making a Relative Measurement  
(2x)  
í
Jump to ZERO REF.  
Highlight ON.  
The relative measurement now becomes the main  
÷
û
reading, while the former main measurement has moved  
to the smaller secondary reading position. (See Figure  
1-9.)  
Activate the relative  
measurement.  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Changing the Time Base  
Selecting Auto/Manual Ranges  

ô
Increase the number of periods.  
Press  
to automatically adjust the position, range,  
time base, and triggering. This assures a stable display on  
nearly all waveforms. The bottom line shows the range,  
the time base for both inputs, and the trigger information.  
Decrease the number of periods.  
Available settings are from 20 ns/div to 5 s/div in normal  
mode.  
Press  
a second time to select the manual range.  
MANUAL appears at the bottom of the reading area.  
Changing the Graphic Representation  
on the Screen  
From Auto range, you can use the light-gray rocker keys  
to change the graphic representation on the screen  
manually.  
Changing the Amplitude  

ô
Enlarge the waveform.  
Reduce the waveform.  
Available settings are from 5 mV/div to 500 V/div when  
using the test leads.  
Observe that AUTO at the bottom of the reading area  
disappears to indicate that the continuous Auto Set  
function is not valid anymore.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Changing the Graphic Representation on the Screen  
1
Positioning the Waveform on the Screen  
Considerable flexibility is offered in moving the  
waveform(s) around the screen.  

Press until you have left any open  
menu. Observe that the following  
main menu appears on bottom of  
the screen.  
ô
í
Choose A MOVE.  
Position the waveform of INPUT  
A on the screen.  
Waveform positioning is demonstrated in Figure 1-10.  
Figure 1-10. Positioning the Waveform  
Observe that the trigger identifier ( ) moves horizontally  
on the screen.  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Smoothing the Waveform  
To smooth the waveform, do the following:  

ô
Open the SCOPE INPUTS menu.  
Open the SCOPE OPTIONS  
submenu.  
í
÷
û
Jump to WAVEFORM MODE.  
Highlight SMOOTH.  
Accept waveform smooth.  
Figure 1-11. Smoothing the Waveform  
You can use waveform smooth to suppress noise without  
loss of bandwidth. Waveform samples with and without  
smoothing are shown in Figure 1-11.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Changing the Graphic Representation on the Screen  
1
Displaying the Envelope of a Waveform  
The test tool records the envelope (minimum and  
maximum) of the live waveforms A and B.  
Repeat the first three actions of ‘Smoothing the  
Waveform’, and then do the following:  
÷
û
Highlight ENVELOPE.  
Start monitoring the envelope of  
the waveform.  
The screen shows the resultant envelope in a gray  
waveform. See Figure 1-12.  
You can use ENVELOPE to observe variations in time or  
amplitude of input waveforms over a longer period of time.  
Figure 1-12. Displaying the Envelope of a Waveform  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
TrendPlotting a Waveform  
The TrendPlot™ function plots the digital readings as a  
function of time. Date and time stamp shows the time of  
the most recent change in a MIN or MAX reading.  
Starting a TrendPlot™ function  

ô
Open the INPUT A menu.  
Start TRENDPLOT.  
The test tool records the minimum (MIN) reading as the  
main (upper displayed) measurement of input A. The date  
and time stamp appear below the MIN reading. (See  
Figure 1-13.)  
The test tool also continuously logs all readings to  
memory and displays these as graphs. Automatic vertical  
scaling and horizontal time compression resizes the  
TrendPlot to fit on the screen. The TrendPlot is built up on  
the screen from left to right until the screen is full. The  
automatic time scaling then compresses this information  
to about half the screen.  
Figure 1-13. TrendPlot Reading  
Note  
When a new minimum value is detected , a beep  
occurs and the new minimum is displayed.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Acquiring the Waveform  
1
Changing the TrendPlot Reading  
Acquiring the Waveform  
To toggle the TrendPlot reading between MIN (minimum),  
MAX (maximum), and AVERAGE, do the following:  
Making a Single Acquisition  
To catch single events, you can perform a single shot.  
(One time screen update.) To set up the test tool for a  
single shot on the input A waveform, do the following:  
í
÷
Change MIN into MAX reading.  
Change MAX into AVG reading.  

Connect the probe to the signal to be measured.  
Open the SCOPE INPUTS menu.  
Note that the date and time stamp now updates  
continuously to indicate the most recent change in a  
reading.  
ô
Open the SCOPE OPTIONS  
submenu.  
Turning Off the TrendPlot Display  
û
Turn TrendPlot off again.  
í
÷
Highlight SINGLE SHOT.  
(2x)  
Accept the set up for a Single  
Shot.  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
û
ø
ù
Wait appears on bottom of the  
screen to indicate that the test  
tool is waiting for a trigger.  
Run appears on bottom of the  
screen when the single  
acquisition is triggered.  
Hold appears on bottom of the  
screen when the single  
acquisition has been completed.  
The test tool will now have a screen like Figure 1-14.  
To perform a next single acquisition, do the following:  
Wait for another single acquisition  
trigger.  
Figure 1-14. Making a Single Acquisition  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Acquiring the Waveform  
1
Recording Slow Signals over a Long Period of  
Time  
The roll mode function supplies a visual log of waveform  
activity and is especially useful when you measure lower  
frequency waveforms.  

ô
Open the SCOPE INPUTS menu.  
Open the SCOPE OPTIONS  
submenu.  
í
÷
Highlight ROLL MODE.  
Start Recording.  
Figure 1-15. Recording Waveforms over Longer  
Period of Time  
(2x)  
û
Freeze Recording.  
The waveform moves accross the screen from right to left  
like a normal chart recorder. Observe that during  
recording no measurements are made. (See Figure 1-15.)  
Observe that the measurement values are only displayed  
after is pressed. (See figure 1-15.)  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Selecting AC-Coupling  
Reversing the Polarity of the Displayed  
Waveform  
Use AC-coupling when you wish to observe a small AC  
signal that rides on a DC signal.  
To invert the input A waveform, do the following:  

Open the SCOPE INPUTS menu.  

Open the SCOPE INPUTS menu.  
ô
í
Highlight AC.  
ô
í
÷
Select NORMAL (of INPUT A).  
Highlight INVERT.  
(4x)  
Accept AC-coupling for INPUT A.  
(3x)  
Accept inverted waveform  
display.  
For example, a negative-going waveform is displayed as  
positive-going, providing a more meaningful viewing  
perspective in some cases. An inverted display is  
identified by trace identifier A on left of the waveform  
area.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Triggering on a Waveform  
1
Triggering on a Waveform  
í
÷
Adjust the Trigger Level  
continuously. Observe the trigger  
icon on the second time division  
line indicates the trigger level.  
Triggering tells the test tool when to begin displaying the  
waveform. You can select which input signal should be  
used, on which edge this should occur, and you can  
define the condition for a new update of the waveform.  
Finally you can tell the test tool to trigger on video signals.  
Trigger on either positive Slope or  
negative Slope of the chosen  
waveform.  
The bottom line of the waveform area identifies the trigger  
parameters being used. Trigger icons on the screen  
indicate the trigger level and slope. (See Figure 1-16.)  
Setting Trigger Level and Slope  

Perform an AUTO SET.  
For quick operation, use the AUTO SET key to  
automatically trigger on nearly all signals. To optimize  
trigger level and slope manually, do the following:  

Press until you have left any open  
menu.  
ô
Enable the arrow keys for Trigger  
Level and Slope adjustment.  
Figure 1-16. Screen with all Trigger Information  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Selecting the Trigger Parameters  
î
Accept all trigger selections and  
return to normal measurement.  
To trigger on the input A waveform, with automatic screen  
update, and to configure the auto range triggering for  
waveforms from 1 Hz, do the following:  
Note  
Setting the automatic triggering to >1Hz will slow  
down the auto range.  

ô
Open the SCOPE INPUTS menu.  
Open the TRIGGER submenu.  
TRIG:A appears in gray text on bottom of the screen  
when no trigger is found.  
Note  
Gray text in a menu or button bar indicates that  
the function is disabled or the status is not valid.  
í
÷
û
ø
ù
Highlight Input ‘A’.  
Select Input ‘A’.  
Highlight FREE RUN.  
Select FREE RUN.  
Highlight >1 Hz  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Triggering on a Waveform  
1
Isolated Triggering  
Triggering on Video Signals  
Apply an interlaced video signal to the red input A.  
Use the optically isolated trigger probe (ITP120, optional)  
to trigger on an external source, and to isolate the test tool  
from a trigger waveform. See Figure 1-17.  
To trigger on a random video line, continue from point ô  
of the previous example as follows:  
To choose the isolated trigger probe, select ‘EXT’ in point  
÷of the previous example. Trigger level is fixed and is  
TTL compatible.  
í
÷
Highlight VIDEO on A.  
Open the VIDEO TRIGGER  
submenu.  
û
ø
ù
î
Highlight PAL.  
Figure 1-17. Isolated Triggering  
Select PAL.  
Highlight RANDOM.  
Select RANDOM.  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Highlight POSITIVE.  
ü
ê
Accept the video trigger  
selections .  
Trigger level and slope are now fixed. (See Figure 1-18.)  
Positive video is indicated as a +” icon on bottom of the  
screen.  
Figure 1-18. Measuring Video Signals  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Triggering on a Waveform  
1
Pressing  
selects the line number function.  
Triggering on a Specific Video Line  
To view a specific video line in more detail you can select  
the line number. To measure on a selected video line,  
continue from point øof the previous example as follows:  
To choose line 135, do the following:  

ô
Enable video line selection.  
Select number 135.  
ù
Highlight SELECT  
î
ü
ê
Select SELECT  
Highlight POSITIVE.  
Accept the video trigger  
selections .  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Note  
Saving and Recalling a Setup or a  
Screen  
From this point the screen is frozen until you  
leave the SAVE&PRINT menu again.  
You can save Screens and Setups to memory, and recall  
them again from memory. Two Screen memories and ten  
Setup memories are available. Save Screens when you  
want to use the present screen image for future reference.  
Save a Setup when you need the present operating  
configuration more often for your measurements.  
ô
Open the SAVE SCREEN  
submenu.  
Note  
Because the navigations for Setups and Screens  
are identical, only saving and recalling Screens  
are explained in this Section.  
Note that free memory locations are indicated by an open  
square (p) before the memory number.  
Saving Screens  
í
Highlight SCREEN 2.  
To save a screen, do the following:  

Open the SAVE&PRINT menu.  
÷
Save the actual screen.  
Recalling Screens  
To recall a screen, do the following:  

ô
Open the SAVE&PRINT menu.  
Highlight SCREENS RECALL.  
Observe that SCREENS SAVE is  
already highlighted.  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Saving and Recalling a Setup or a Screen  
1
Deleting Screens  
To delete all screen memories, do the following:  
í
Open the RECALL SCREEN  
submenu.  

Open the SAVE&PRINT menu.  
Highlight SCREENS DELETE.  
ô
í
Open the DELETE SCREEN  
submenu.  
Already filled memory locations  
are indicated with a closed  
square (n).  
÷
û
Highlight SCREEN 2.  
Filled memory locations are  
indicated with a closed square.  
÷
û
Highlight BOTH SCREENS.  
Delete all screen memories.  
View the saved screen.  
The image is presented as a picture that can no longer be  
changed.  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Using a Printer  
To print a (graphic) hard copy of the present screen, you  
need to use one of the following:  
l
l
The Optically Isolated RS-232 Adapter/Cable  
(PM9080) to connect a serial printer to the OPTICAL  
PORT of the test tool. See Figure 1-19.  
The Print Adapter Cable (PAC91, optional) to connect  
a parallel printer to the OPTICAL PORT of the test  
tool. See Figure 1-20.  
Figure 1-19. Connecting a Serial Printer  
Figure 1-20. Connecting a Parallel Printer  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using The Test Tool  
Using a Printer  
1
This example covers how to set up the test tool to print on  
a HP Deskjet printer with a baudrate of 9600 baud:  
Now you are ready to print.  
To print a live screen, do the following:  

Open the SAVE&PRINT menu.  
Observe that the screen is  
freezed.  
ù
î
Open the SAVE&PRINT menu.  
Start printing.  
ô
Open the PRINTER SETUP  
submenu.  
To print a recalled screen, do the following:  
Start printing.  
ù
A message that indicates that the test tool is printing  
appears on bottom of the screen.  
í
÷
û
ø
Highlight DESKJET.  
Select DESKJET.  
Highlight 9600.  
Accept the print selections.  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Using FlukeView® Software  
To connect the test tool to a computer for using the  
FlukeView software for Windows® (SW90W), do the  
following:  
l
Use the Optically Isolated RS-232 Adapter/Cable  
(PM9080) to connect a computer to the OPTICAL  
PORT of the test tool. See Figure 1-21.  
Figure 1-21. Connecting a Computer  
For all information relating to installing and using the  
FlukeView ScopeMeter software, see the SW90W Users  
Manual.  
A Software & Cable Carrying Case Kit is optional available  
as model number SCC 120.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 2  
Maintaining the Test Tool  
About this Chapter  
Cleaning the Test Tool  
This chapter covers basic maintenance procedures that  
can be performed by the user. For complete service,  
disassembly, repair, and calibration information, see the  
Service Manual. You will find the part number of the  
Service Manual in the section ‘Parts and Accessories’ in  
this manual.  
Clean the test tool with a damp cloth and a mild soap to  
avoid abrasion of text on the test tool. Do not use  
abrasives, solvents, or alcohol.  
Storing the Test Tool  
If you are storing the test tool for an extended period of  
time, charge the Ni-Cd battery pack before storing. It is  
not necessary to remove the battery pack.  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Charging the Ni-Cd Battery Pack  
At delivery, the Ni-Cd batteries may be empty and must be  
charged for 4 hours (test tool is off) to fill them completely.  
When fully charged, the batteries typically provide 4 hours  
of use at full brightness and 5 hours at normal brightness.  
When battery powered, the battery indicator on the bottom  
of the screen informs you about the condition of the  
battery. The battery symbols are:  
.
The battery symbol indicates that there are typically  
five minutes of operating time left.  
Use the setup as shown in Figure 2-1 to charge the battery  
pack and power the instrument.  
Turn the test tool off to charge the batteries more quickly.  
Note  
Figure 2-1. Charging the Battery Pack  
The test tool uses a trickle charge for the  
batteries, so no damage can occur if you leave it  
charging for long periods, e.g., through the  
weekend.  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintaining the Test Tool  
Keeping Batteries in Optimal Condition  
2
To refresh the battery pack, do the following:  
Keeping Batteries in Optimal Condition  

Be sure that the test tool is line powered.  
Open the USER OPTIONS menu.  
Always operate the test tool on batteries until an  
-icon  
appears on the bottom line of the screen. This indicates  
that the battery level is low and that the Ni-Cd batteries  
need to be recharged.  
ô
Open the BATTERY REFRESH  
submenu.  
Frequent charging of the batteries when they are not  
completely empty can reduce the operating time for the  
test tool.  
You can refresh the battery pack at any time. This battery  
refresh cycle fully discharges and charges the battery pack.  
A complete refresh cycle takes about 12 hours and should  
be done at least four times a year.  
í
÷
Highlight START REFRESH.  
Start the refresh cycle.  
Note  
Note  
Be sure not to disconnect the Power Adapter  
during the complete refresh cycle. Doing so will  
interrupt the refresh cycle.  
After start of the refresh cycle, the screen will be  
black.The backlight is on during discharging  
within the refresh cycle.  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Replacing and Disposing of the Ni-Cd  
Battery Pack  
Warning  
To avoid electrical shock, remove the test  
leads and probes before replacing the battery  
pack.  
Note  
This instrument contains Nickel-Cadmium  
batteries. Do not dispose of this battery pack with  
other solid waste. Used batteries should be  
disposed of by a qualified recycler or hazardous  
materials handler. Contact your authorized  
FLUKE Service Center for recycling information.  
To replace the battery pack, do the following: (See Figure  
2-2.)  
1. Disconnect the test leads and probes both at the  
source and at the test tool.  
Figure 2-2. Replacing the Battery Pack  
2. Power the test tool with the Power Adapter.This  
ensures information stored in memories will not be lost.  
4. Lift the battery access cover away from the test tool.  
5. Take the battery pack out of the battery compartment.  
6. Remove the battery plug from the connector.  
3. Locate the battery access cover on the bottom rear.  
Loosen the screw with a flat-blade screwdriver.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintaining the Test Tool  
Using and Adjusting 10:1 Scope Probes  
2
7. Install a new battery pack.  
Note  
Ensure that the battery pack is placed in the  
battery compartment as shown in Figure 2-2. Use  
only the Fluke BP120 Ni-Cd battery pack.  
8. Reinstall the battery cover and secure the screw.  
Using and Adjusting 10:1 Scope Probes  
You need to adjust the red and gray scope probes  
(PM8918, optional) for optimal response.  
Warning  
To avoid electrical shock use the BB120  
Banana-to-BNC adapter (delivered with the  
test tool) to connect a 10:1 scope probe to  
the input of the test tool.  
To adjust probes, do the following:  
Connect the 10:1 scope probe from the gray input B  
jack to the red input A jack. Use the red 4-mm  
banana adapter (delivered with the probe) and the  
banana-to-BNC adapter (BB120). See Figure 2-3.  
Figure 2-3. Adjusting Scope Probes  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  

ô
Open the SCOPE INPUTS menu.  
Open the PROBES submenu.  
ù
Adjust the trimmer screw in the  
probe housing to give an  
optimum square wave.  
í
÷
Highlight PROBE AC ADJUST.  
î
Return to normal mode.  
Open the PROBE AC ADJUST  
submenu.  
û
ø
Highlight ADJUST 10:1 PROBE.  
A square wave appears on the  
screen.  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintaining the Test Tool  
Calibrating the Test Tool  
2
Calibrating the Test Tool  
Parts and Accessories  
You can ask for the model identity (version and calibration  
data) at any time. To display the identity, do the following:  
Service Manual  
Ordering Number: 4822 872 05375  

ô
Open the USER OPTIONS menu.  
Standard Accessories  
Open the  
VERSION&CALIBRATION  
submenu.  
The next tables list the user-replaceable parts for the  
various test tool models. To order replacement parts,  
contact your nearest service center.  
The screen gives you information about the model number  
with software version, the calibration number with latest  
calibration date, and the latest battery refresh date.  
í
Return to normal mode.  
Recalibration must be carried out by qualified personnel  
only. Contact your local Fluke representative for  
recalibration.  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Standard Accessories (cont)  
Item  
Ordering Code  
Ni-Cd Battery Pack (installed)  
BP120  
Power Adapter/Battery Charger, available models:  
Universal Europe 230V, 50Hz  
PM8907/801  
PM8907/803  
PM8907/804  
PM8907/806  
PM8907/807  
PM8907/808  
North America 120V, 60Hz  
United Kingdom 240V, 50Hz  
Japan 100V, 60Hz  
Australia 240V, 50Hz  
Universal 115V/230V *  
* The 230V rating of the PM8907/808 is not for use in North America.  
A line plug adapter complying with the applicable National  
Requirements may be provided to alter the blade configurations for  
a specific country.  
Set of two Shielded Test Leads (Red and Gray), designed for use only  
with the Fluke ScopeMeter 120 series test tool.  
STL120  
Set contains the following replaceable part:  
Ground Lead with Alligator Clip (Black)  
5322 320 11354  
TL75  
Set of two Test Leads (Red and Black)  
UL1244  
Set of two Hook Clips (Red and Gray)  
HC120  
AC120  
BB120  
Set of three Alligator Clips (Red, Gray, and Black)  
Set of two Banana-to-BNC Adapters (Black)  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Maintaining the Test Tool  
Parts and Accessories  
2
Standard Accessories (cont)  
Item  
Ordering Code  
Users Manual (English)  
Users Manual (German)  
Users Manual (French)  
Users Manual (Spanish)  
Users Manual (Portuguese)  
Users Manual (Italian)  
Users Manual (Dutch)  
Users Manual (Danish)  
Users Manual (Norwegian)  
Users Manual (Swedish)  
Users Manual (Finnish)  
Users Manual (Chinese)  
Users Manual (Japanese)  
Users Manual (Korean)  
4822 872 00743  
4822 872 00744  
4822 872 00745  
4822 872 00746  
4822 872 00795  
4822 872 00747  
4822 872 00748  
4822 872 00749  
4822 872 00751  
4822 872 00752  
4822 872 00753  
4822 872 00754  
4822 872 00755  
4822 872 00756  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Optional Accessories  
Item  
Ordering Code  
Software & Cable Carrying Case Kit  
Set contains the following parts:  
Optically Isolated RS-232 Adapter/Cable  
Hard Carrying Case  
SCC 120  
PM9080  
C120  
Set of two Banana-to-BNC Adapters (Black)  
FlukeView® ScopeMeter® Software for Windows®  
Set of 10:1 Scope Probes (Red and Gray)  
BB120  
SW90W  
PM8918/002  
UL1244  
Optically Isolated RS-232 Adapter/Cable  
Hard Carrying Case  
PM9080  
C120  
Compact Soft Case  
C125  
Soft Carrying Case  
C789  
Isolated Trigger Probe  
ITP120  
PAC91  
Print Adapter Cable for Parallel Printers  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 3  
Tips and Troubleshooting  
Goal of this Chapter  
This Chapter gives you information and tips on how you  
can make the best use of the test tool.  
Using the Tilt Stand  
The test tool is equipped with a tilt stand, allowing viewing  
from an angle. You can also use the tilt stand to hang the  
test tool at a convenient viewing position. Simply tilt the  
stand and hang the test tool. Typical positions are shown  
in Figure 3-1.  
Figure 3-1. Using the Tilt Stand  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Resetting the Test Tool  
Changing the Information Language  
Perform a Master Reset to make sure that your test tool is  
in the initial settings condition.  
When you operate the test tool, messages appear on the  
bottom of the screen. These messages are always  
displayed in a box, and can in some test tool versions be  
displayed in more languages. In different combinations,  
you can choose from 10 languages: English, French,  
German, Italian, Spanish, Portuguese, Dutch, Japanese,  
Korean, and Chinese.  

ô
í
Turn the test tool off.  
Press and hold.  
Press and release.  
If you want to change the language of the messages into  
e.g. Spanish, do the following:  
The test tool turns on, and you should hear a double  
beep, indicating the Reset was successful.  

ô
Open the USER OPTIONS menu.  
÷
Release.  
Open the LANGUAGE SELECT  
submenu.  
í
÷
Highlight ESPAÑOL.  
Accept ESPAÑOL (Spanish) as  
language.  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Tips and Troubleshooting  
Changing the Display  
3
Setting the Grid Display  
To choose a dotted grid, do the following:  
Changing the Display  
Adjusting the Screen Contrast  


Open the USER OPTIONS menu.  
From the main menu, choose  
CONTRAST.  
ô
í
Highlight GRID TYPE.  
Open the GRID TYPE submenu.  
ô
Adjust the contrast of the screen.  
Note  
The new contrast adjustment of the screen is  
saved in memory until a new adjustment is made.  
÷
û
Choose DOTS.  
Accept the new grid display.  
Use LINES when you need a cross hatch pattern based  
on the horizontal time and vertical divisions on the screen.  
Use DOTS when you need vertical and horizontal division  
dots as added reference points to the screen.  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Changing Date and Time  
ø
ù
î
ü
ê
Choose 04.  
The test tool has a date and time clock. To change the  
date to (e.g.) 19 April, 1996, do the following:  
Jump to DAY.  
Choose 19.  

Open USER OPTIONS menu.  
Jump to FORMAT.  
Choose DD/MM/YY.  
Accept the new date.  
You can change the time in a similar way by opening the  
TIME ADJUST submenu. (steps ôand í.)  
ô
í
Highlight DATE ADJUST.  
Open DATE ADJUST submenu.  
÷
û
Choose 1996.  
Jump to MONTH.  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Tips and Troubleshooting  
Saving Battery Life  
3
Saving Battery Life  
í
Open the submenu.  
When operated on the battery pack (no Power Adapter  
connected), the test tool conserves power by shuting itself  
down. If you have not pressed a key for at least 30 minutes,  
the test tool turns itself off automatically.  
Note  
If the Power Adapter is connected, there is no  
automatic power shutdown.  
÷
û
Highlight AFTER 5 MIN.  
Although automatic power shutdown will not occur if  
TrendPlot is on, the backlight will dim. Recording will  
continue even if the battery pack is low, and retention of  
memories is not jeopardized.  
Accept the new power down time.  
Setting the Power Down Timer  
To extend battery life, the power shutdown time is set to  
30 minutes after the last key press. To set the power  
shutdown to five minutes, do the following:  

ô
Open the USER OPTIONS menu.  
Highlight POWER DOWN ...  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Changing the Auto Set Options  
ø
ù
Highlight UNCHANGED.  
Normally, the Auto Set function captures waveforms from  
15 Hz.  
Accept the new Auto Set  
configuration.  
To configure Auto Set for waveforms from 1 Hz with  
unchanged input coupling, do the following:  
Note  
Setting the Auto Set adjust to 1 Hz will slow  
down the Auto Set response.  

Open the User Options menu.  
Highlight AUTOSET ADJUST...  
ô
í
Open the AUTOSET ADJUST  
submenu.  
÷
û
Highlight SIGNALS > 1 Hz.  
Select COUPLING.  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Tips and Troubleshooting  
Using Proper Grounding  
3
Using Proper Grounding  
Incorrect grounding can cause various problems. This  
Section gives you guidelines for proper grounding.  
l
Use the short ground lead(s) when measuring DC or  
AC signals on input A and input B. (See Figure 3-2.)  
Warning  
To avoid electrical shock or fire, use only one  
COM (common) connection, or ensure that all  
connections to COM are at the same  
potential.  
Figure 3-2. Grounding with Short Ground Lead  
l
Use the unshielded black ground lead to COM  
(common) for Ohm (), Continuity, Diode, and  
Capacitance measurements. (See Figure 3-3.)  
Using the unshielded ground lead is also possible for  
single or dual input measurements for waveforms with  
a frequency up to 1 MHz. This may add some hum or  
noise to the waveform display due to the unshielded  
ground lead.  
Figure 3-3. Grounding with Unshielded Ground Lead  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Solving Printing and Other  
Communication Errors  
Battery Testing of Fluke Accessories  
When using battery operated Fluke accessories, always  
check the battery condition of the accessory first on a  
Fluke multimeter .  
RS-232 communication may cause problems. When  
experiencing communication problems, try the following  
remedies:  
l
Make sure that the interface cable is connected to the  
correct port on the printer or computer. If necessary  
use the 9 pin to 25 pin adapter or gender changer.  
l
l
l
Make sure that you have selected the correct printer  
type. (To select printer type, see Chapter 1.)  
Make sure that the baud rate matches with the printer  
or computer. (To set the baud rate, see Chapter 1.)  
Reset the RS-232 parameters to defaults.  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chapter 4  
Specifications  
Safety Characteristics  
Introduction  
The test tool has been designed and tested in accordance  
with Standards ANSI/ISA S82.01-1994, EN 61010.1  
(1993) (IEC 1010-1), CAN/CSA-C22.2 No.1010.1-92  
(including approval), UL3111-1 (including approval) Safety  
Requirements for Electrical Equipment for Measurement,  
Control, and Laboratory Use.  
Performance Characteristics  
FLUKE guarantees the properties expressed in numerical  
values with the stated tolerance. Specified non-tolerance  
numerical values indicate those that could be nominally  
expected from the mean of a range of identical  
ScopeMeter test tools.  
This manual contains information and warnings that must  
be followed by the user to ensure safe operation and to  
keep the instrument in a safe condition. Use of this  
equipment in a manner not specified by the manufacturer  
may impair protection provided by the equipment.  
Environmental Data  
The environmental data mentioned in this manual are  
based on the results of the manufacturer’s verification  
procedures.  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Input Impedance  
Dual Input Oscilloscope  
Vertical  
excluding probes and test leads ................1 M//12 pF  
with BB120.................................................1 M//20 pF  
with STL120.............................................1 M//225 pF  
with PM8918 ............................................10 M//15 pF  
Sensitivity............................................5 mV to 500 V/div  
Frequency Response  
DC Coupled:  
excluding probes and test leads: ...................................  
DC to 20 MHz (-3 dB)  
Display Modes ..............................................A, -A, B, -B  
with STL120 1:1 shielded test leads:.............................  
DC to 12.5 MHz (-3 dB)  
DC to 20 MHz (-6 dB)  
with PM8918 10:1 probe (optional accessory):..............  
DC to 20 MHz (-3 dB)  
Max. Input Voltage A and B  
direct or with test leads ...................................600 Vrms  
with BB120......................................................300 Vrms  
(For detailed specifications, see “Safety”)  
Max. Floating Voltage  
from any terminal to ground............................600 Vrms  
up to 400 Hz  
AC Coupled (LF roll off):  
excluding probes and test leads .............<10 Hz (-3 dB)  
with STL120 ............................................<10 Hz (-3dB)  
with PM8918.............................................<1 Hz (-3 dB)  
Resolution................................................................ 8 bit  
Vertical Accuracy .......................±(1% + 0.05 range/div)  
Max. Vertical Move....................................... ±4 divisions  
Rise Time  
excluding probes and test leads...................... <17.5 ns  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Dual Input Oscilloscope  
4
Horizontal  
Trigger  
Scope Modes .................................. Normal, Single, Roll  
Screen Update...............................Free Run, On Trigger  
Ranges  
Source.............................................................. A, B, EXT  
Normal:  
EXTernal via optically isolated trigger probe ITP120  
equivalent sampling ...................20 ns to 500 ns/div  
real time sampling..............................1 µs to 5 s/div  
Single (real time) .....................................1 µs to 5 s/div  
Roll (real time) .........................................1s to 60 s/div  
(optional accessory)  
Sensitivity A and B  
@ DC to 5 MHz.............................0.5 divisions or 5 mV  
@ 25 MHz.................................................. 1.5 divisions  
@ 40 MHz..................................................... 4 divisions  
Sampling Rate (for both channels simultaneously)  
Equivalent sampling (repetitive signals) ........................  
up to 1.25 GS/s  
Slope................................................... Positive, Negative  
Video on A..........................interlaced video signals only  
Modes ............................................... Lines, Line Select  
Standards...........................PAL, NTSC, PAL+, SECAM  
Polarity.............................................. Positive, Negative  
Sensitivity..........................................0.6 divisions sync.  
Real time sampling:  
1 µs to 5 ms/div...........................................25 MS/s  
10 ms to 60 s/div...........................................5 MS/s  
Time Base Accuracy  
Equivalent sampling .................. ±(0.4% +0.04 time/div)  
Real time sampling .................... ±(0.1% +0.04 time/div)  
Glitch Detection..................40 ns @ 20 ns to 5 ms/div  
200 ns @ 10 ms to 60 s/div  
Glitch detection is always active.  
Horizontal Move........................................... 10 divisions  
Trigger point can be positioned anywhere across the  
screen.  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Advanced Scope Functions  
Display Modes  
Normal....... Captures up to 40 ns glitches and displays  
analog-like persistence waveform.  
Smooth ...... Suppresses noise from a waveform.  
Envelope ... Records and displays the minimum and  
maximum of waveforms over time.  
Auto Set  
Continuous fully automatic adjustment of amplitude, time  
base, trigger levels, trigger gap, and hold-off. Manual  
override by user adjustment of amplitude, time base, or  
trigger level.  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Dual Input Meter  
4
5 MHz to 12.5 MHz.................... ±(30% +25 counts)  
5 MHz to 20 MHz (excl. test leads or probes) .........  
±(30% +25 counts)  
Dual Input Meter  
The accuracy of all measurements is within ± (% of  
reading + number of counts) from 18 °C to 28 °C.  
Add 0.1x (specific accuracy) for each °C below 18 °C or  
above 28 °C. For voltage measurements with 10:1 probe,  
add probe uncertainty +1%. More than one waveform  
period must be visible on the sceen.  
AC coupled with 1:1 (shielded) test leads  
60 Hz (6 Hz with 10:1 probe)..........................-1.5%  
50 Hz (5 Hz with 10:1 probe).............................-2%  
33 Hz (3.3 Hz with 10:1 probe)..........................-5%  
10 Hz (1 Hz with 10:1 probe)...........................-30%  
DC Rejection (only VAC)....................................>50 dB  
Common Mode Rejection (CMRR) ....... >100 dB @ DC  
>60 dB @ 50, 60, or 400 Hz  
Input A and Input B  
DC Voltage (VDC)  
Ranges ........................500 mV, 5V, 50V, 500V, 1250V  
Accuracy...........................................±(0.5% +5 counts)  
Normal Mode Rejection (SMR).......................... >60 dB  
@ 50 or 60 Hz ±1%  
Common Mode Rejection (CMRR) ....... >100 dB @ DC  
>60 dB @ 50, 60, or 400 Hz  
Full Scale Reading.....................................5000 counts  
The reading is independent of any signal crest factor.  
Peak  
Modes ........................ Max peak, Min peak, or pk-to-pk  
Ranges........................ 500 mV, 5V, 50V, 500V, 1250V  
Accuracy:  
Full Scale Reading .................................... 5000 counts  
Max peak or Min peak........................ 5% of full scale  
Peak-to-Peak ................................... 10% of full scale  
Full Scale Reading.......................................500 counts  
True RMS Voltages (VAC and VAC+DC)  
Ranges ........................500 mV, 5V, 50V, 500V, 1250V  
Accuracy for 5 to 100% of range  
Frequency (Hz)  
DC coupled:  
Ranges................. 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz,  
100 kHz,1 MHz, 10 MHz, and 40 MHz  
Frequency Range in Continuous Autoset .....................  
15 Hz (1 Hz) to 30 MHz  
DC to 60 Hz (VAC+DC) ...............±(1% +10 counts)  
1 Hz to 60 Hz (VAC).....................±(1% +10 counts)  
AC or DC coupled:  
60 Hz to 20 kHz ........................±(2.5% +15 counts)  
20 kHz to 1 MHz...........................±(5% +20 counts)  
1 MHz to 5 MHz .........................±(10% +25 counts)  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Accuracy:  
Temperature (TEMP)....with optional temperature probe  
Range ....................................... 200 °C/div (200 °F/div)  
Scale Factor................................ 1 mV/°C and 1 mV/°F  
Accuracy........... as VDC (add temp. probe uncertainty)  
@1 Hz to 1 MHz ............................±(0.5% +2 counts)  
@1 MHz to 10 MHz .......................±(1.0% +2 counts)  
@10 MHz to 40 MHz .....................±(2.5% +2 counts)  
Full Scale Reading ................................. 10 000 counts  
Decibel (dB)  
Duty Cycle (DUTY)  
0 dBV........................................................................ 1V  
0 dBm (600/50).............................................. 1 mW  
referenced to 600or 50Ω  
dB on ......................................VDC, VAC, or VAC+DC  
Full Scale Reading.....................................1000 counts  
Range.......................................................... 2% to 98%  
Frequency Range in Continuous Autoset......................  
15 Hz (1 Hz) to 30 MHz  
Accuracy:  
@1Hz to 1 MHz .............................±(0.5% +2 counts)  
@1 MHz to 10 MHz .......................±(1.0% +2 counts)  
@10 MHz to 40 MHz .....................±(2.5% +2 counts)  
Resolution...........................................................0.1%  
Crest Factor (CREST)  
Range ................................................................ 1 to 10  
Accuracy...............................................±(5% +1 count)  
Full Scale Reading.......................................100 counts  
Pulse Width (PULSE)  
Frequency Range in Continuous Autoset......................  
15 Hz (1 Hz) to 30 MHz  
Phase  
Modes......................................................A to B, B to A  
Range .................................................0 to 359 degrees  
Accuracy......................................±(1 degree +1 count)  
Resolution....................................................... 1 degree  
Accuracy:  
@1 Hz to 1 MHz ............................±(0.5% +2 counts)  
@1 MHz to 10 MHz .......................±(1.0% +2 counts)  
@10 MHz to 40 MHz .....................±(2.5% +2 counts)  
Full Scale Reading ................................. 1000 counts  
Input A  
Ohm ()  
Ranges .......500, 5 k, 50 k, 500 k, 5 M, 30 MΩ  
Accuracy:......................................... ±(0.6% +5 counts)  
Full Scale Reading:  
500to 5 M.........................................5000 counts  
30 M.....................................................3000 counts  
Amperes (AMP).................... with optional current probe  
Ranges ......... same as VDC, VAC, VAC+DC, or PEAK  
Scale Factor ...1 mV/A, 10 mV/A, 100 mV/A, and 1 V/A  
Accuracy....... same as VDC, VAC, VAC+DC, or PEAK  
(add current probe uncertainty)  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Dual Input Meter  
4
Measurement Current..........................0.5 mA to 50 nA  
decreases with increasing ranges  
Open Circuit Voltage .............................................. <4V  
Advanced Meter Functions  
Zero Set  
Set actual value to reference  
Continuity (CONT)  
Fast/Normal/Smooth  
Beep .....................................<(30±5) in 50range  
Measurement Current........................................ 0.5 mA  
Detection of shorts of .......................................... 1 ms  
Meter settling time Fast: 1s @ 1 µs to 10 ms/div.  
Meter settling time Normal: 2s @ 1 µs to 10 ms/div.  
Meter settling time Smooth: 10s @ 1 µs to 10 ms/div.  
Diode  
Measurement Voltage:  
Touch Hold (on A)  
@0.5 mA........................................................... >2.8V  
@open circuit....................................................... <4V  
Accuracy..............................................±(2% +5 counts)  
Measurement Current........................................ 0.5 mA  
Polarity......................................+ on input A, - on COM  
Captures and freezes a stable measurement result.  
Beeps when stable. Touch Hold works on the main  
meter reading, with thresholds of 1 Vpp for AC signals  
and 100 mV for DC signals.  
TrendPlot  
Capacitance (CAP)  
Graphs meter readings of the Min and Max values from  
15 s/div (120 seconds) to 2 days/div (16 days) with time  
and date stamp. Automatic vertical scaling and time  
compression.  
Displays the actual and Minimum, Maximum, or average  
(AVG) reading.  
Ranges .................50 nF, 500 nF, 5 µF, 50 µF, 500 µF  
Accuracy............................................±(2% +10 counts)  
Full Scale Reading .................................... 5000 counts  
Measurement Current............................5 µA to 0.5 mA  
increases with increasing ranges  
Dual slope integrating measurement with parasitic  
serial and parallel resistance cancellation.  
Fixed Decimal Point  
Possible by using attenuation keys.  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Memory  
Miscellaneous  
Number of Screens......................................................2  
Number of User Setups .............................................10  
Display  
Size ....................................72 x 72 mm (2.83 x 2.83 in)  
Resolution ........................................... 240 x 240 pixels  
Waveform Display:  
Vertical..............................................8 div of 20 pixels  
Horizontal ......................................9.6 div of 25 pixels  
Backlight..................Cold Cathode Fluorescent (CCFL)  
Mechanical  
Size.......................232 x 115 x 50 mm (9.1 x 4.5 x 2 in)  
Weight.................................................... 1.1 kg (2.5 lbs)  
including battery pack  
Interface.................................. RS-232, optically isolated  
To Printer..........................supports Epson FX, LQ, and  
HP Deskjet®, Laserjet®, and Postscript  
Power  
External: .............................via Power Adapter PM8907  
Input Voltage .........................................10 to 21V DC  
Power .........................................................5W typical  
Input Connector ..........................................5 mm jack  
Internal:  
Battery Power .................... Rechargeable Ni-Cd 4.8V  
Operating Time...............4 hours with bright backlight  
5 hours with dimmed backlight  
Serial via PM9080 (optically isolated RS-232  
adapter/cable, optional).  
Parallel via PAC91 (optically isolated print adapter  
cable, optional).  
To PC........................ Dump and load settings and data  
Serial via PM9080 (optically isolated RS-232  
adapter/cable, optional), using SW90W (FlukeView®  
software for Windows®).  
Charging Time ...................... 4 hours with test tool off  
12 hours with test tool on  
12 hours with refresh cycle  
Allowable ambient temperature:  
during charging ................. 0 to 45 °C (32 to 113 °F)  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Environmental  
4
Electromagnetic Compatibility (EMC)  
Environmental  
Emission ........................................ EN 50081-1 (1992):  
EN55022 and EN60555-2  
Environmental ....MIL 28800E, Type 3, Class III, Style B  
Temperature  
Immunity ........................................ EN 50082-2 (1992):  
IEC1000-4-2, -3, -4, -5  
Operating................................ 0 to 50 °C (32 to 122 °F)  
Storage.................................-20 to 60 °C (-4 to 140 °F)  
(See also Tables 1 to 3)  
Humidity  
Enclosure Protection.......................... IP51, ref: IEC529  
Operating:  
@0 to 10 °C (32 to 50 °F) .................noncondensing  
@10 to 30 °C (50 to 86 °F) ................................ 95%  
@30 to 40 °C (86 to 104 °F) ................................75%  
@40 to 50 °C (104 to 122 °F) ..............................45%  
Storage:  
@-20 to 60 °C (-4 to 140 °F)...............noncondensing  
Altitude  
Operating.......................................4.5 km (15 000 feet)  
Max. Input and Floating Voltage 600 Vrms upto 2 km,  
linearly derating to 400 Vrms @ 4.5 km  
Storage...........................................12 km (40 000 feet)  
Vibration............................................................. max. 3g  
Shock................................................................ max. 30g  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Safety  
Designed for measurements on 600 Vrms Category III  
Installations, Pollution Degree 2, per:  
ANSI/ISA S82.01-1994  
EN61010-1 (1993) (IEC1010-1)  
CAN/CSA-C22.2 No.1010.1-92 (including approval)  
UL3111-1 (including approval)  
Max. Input Voltage Input A and B  
Direct on input or with leads ..........................600 Vrms  
for derating, see Figure 4-1.  
With Banana-to BNC Adapter BB120.............300 Vrms  
for derating, see Figure 4-1.  
Max. Floating Voltage  
from any terminal to ground............................600 Vrms  
up to 400 Hz  
Figure 4-1. Max. Input Voltage v.s. Frequency  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Specifications  
Safety  
4
The Fluke 123, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, as defined by  
IEC1000-4-3, with the addition of the following tables.  
Trace disturbance with STL120  
Table 1  
No visible disturbance  
E= 3 V/m  
E= 10 V/m  
Frequency range 10 kHz to 27 MHz  
Frequency range 27 MHz to 1 GHz  
50 mV/div to 500 V/div  
50 mV/div to 500 V/div  
500 mV/div to 500 V/div  
50 mV/div to 500 V/div  
Table 2  
Disturbance less than 10% of full scale  
E= 3 V/m  
E= 10 V/m  
Frequency range 10 kHz to 27 MHz  
Frequency range 27 MHz to 1 GHz  
10 mV/div to 20 mV/div  
5 mV/div to 20 mV/div  
50 mV/div to 200 mV/div  
-
(-): no visible disturbance  
Test tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full scale.  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Multimeter disturbance:  
VDC, VAC, and VAC+DC with STL120 and short ground lead.  
OHM, CONT, DIODE, and CAP with STL120, and black test lead to COM.  
Table 3  
Disturbance less than 1% of full scale  
E= 3 V/m  
E= 10 V/m  
Frequency range 10 kHz to 27 MHz  
VDC, VAC, VAC+DC  
OHM, CONT, DIODE  
CAP  
500 mV to 1250V  
500to 30 MΩ  
50 nF to 500 µF  
500 mV to 1250V  
500to 30 MΩ  
50 nF to 500 µF  
Frequency range 27 MHz to 1 GHz  
VDC, VAC, VAC+DC  
OHM, CONT, DIODE  
CAP  
500 mV to 1250V  
500to 30 MΩ  
50 nF to 500 µF  
500 mV to 1250V  
500to 30 MΩ  
50 nF to 500 µF  
Test tool ranges not specified in table 3 may have a disturbance of more than 10% of full scale.  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
Automatic Power Shutdown, 51  
Avoid Electrical Shock, 12  
Battery Replacing, 40  
BB120 Adapters, 44  
Blue Function Keys, 10  
—A—  
BP120 Battery Pack, 2, 41, 44  
Bright Display, 9  
AC120 Alligator Clips, 44  
Accessories, 43  
—B—  
AC-Coupling, 26  
Backlight, 9  
Acquiring the Waveform, 23  
Acquisition Modes, 57  
Adjusting Scope Probes, 41  
Advanced Meter Functions, 61  
Advanced Scope Functions, 58  
Alligator Clips, 2, 44  
Altitude, 63  
Ampere Measurement, 60  
Amplitude, 18  
Banana Jack Inputs, 12  
Banana-to-BNC Adapter, 2, 44  
Bandwidth, 56  
Battery Charger, 2, 44  
Battery Disposing, 40  
Battery Indicator, 10  
Battery Life, 51  
Battery Pack, 2, 38, 44  
Battery Power, 62  
Battery Refresh, 39  
Battery Refresh Date, 43  
—C—  
C120 Hard Case, 2, 46  
C789 Soft Case, 46  
Calibrating the Test Tool, 43  
Capacitance, 14, 53, 61  
Changing the Amplitude, 18  
Changing the Display, 49  
Changing the Graphic Representa-  
tion, 18  
Auto Set, 13, 58  
Auto Set Configuration, 52  
Changing the Time Base, 18  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Charge Time, 62  
Charger, 44  
Charging, 38  
Gray Text, 28  
Grey INPUT B, 12  
Grid Display, 49  
—E—  
Earth Ground, 5  
Electrical Shock, 5  
Electrically Floating, 5  
Electromagnetic Compatibility, 1,  
63  
Cleaning, 37  
Common, 5, 12  
Communication Errors, 54  
Computer, 36  
Ground, 5  
Ground Test Lead, 2, 44  
Grounding Problems, 53  
Emission, 63  
Envelope, 58  
Envelope of a Waveform, 21  
Environmental, 63  
Environmental Data, 55  
Connect-and-View function, 13  
Connecting a Computer, 36  
Continuity, 14, 53, 61  
Contrast, 49  
Crest Factor, 60  
Current Measurement, 60  
—H—  
Hands-Free Measurements, 16  
Hard Carrying Case, 2, 46  
HC120 Hook Clips, 44  
Holding a Stable Reading, 16  
Hook Clips, 2, 44  
—F—  
Horizontal Move, 57  
Humidity, 63  
—D—  
Fast/Smooth, 61  
Fixed Decimal Point, 61  
FlukeView, 2, 36, 46  
Freeze Recording, 25  
Freezing the Screen, 16  
Frequency (Hz), 59  
Frequency Response, 56  
Function Keys, 10  
Date, 50  
DC Voltage (VDC), 59  
Decibel (dB), 60  
Declaration of Conformity, 1  
Deleting Screens, 33  
Dimmed Display, 9  
Diode, 14, 53, 61  
Display, 49, 62  
Hz, 59  
—I—  
Immunity, 63  
Information Language, 48  
Input A, 12  
Input B, 12  
Disposing Batteries, 40  
Duty Cycle, 60  
Input Impedance, 56  
Isolated, 5  
—G—  
Glitch Detection, 57  
Graphic Representation, 18  
Isolated Trigger Probe, 29, 46  
Isolated Triggering, 29  
68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index (continued)  
ITP120, 29, 46  
Printing Errors, 54  
Probe, 41, 46, 56  
—N—  
Ni-Cd Battery Pack, 2, 38, 44  
Product Registration Card, 2  
Pulse Width, 60  
—L—  
Language, 48  
Level, 27  
—O—  
—R—  
Ohm (), 14, 53, 60  
Reading Area, 10, 14  
Reading the Screen, 10  
Recalibrating, 43  
Operating Time, 62  
Optical Interface, 34, 36, 62  
—M—  
Maintenance, 37  
Recalling, 32  
Making Measurements, 14  
Manual, 45  
—P—  
Recalling Screens, 33  
Record a Waveform, 21  
Recording Slow Signals, 25  
Red INPUT A, 12  
Refreshing the Batteries, 39, 43  
Relative Measurements, 17  
Replaceable Parts, 43  
Replacing Batteries, 40  
Resetting, 48  
Resetting the Test Tool, 8  
Reversing the Polarity, 26  
Rise Time, 56  
Roll Mode Function, 25  
RS-232 Adapter/Cable, 2, 34, 36,  
46  
Manual Override, 58  
Max. Floating Voltage, 56, 64  
Max. Input Voltage, 56, 64  
Maximum (MAX) Reading, 23  
Measurement Connections, 12  
Measurements, 14  
Mechanical, 62  
Memory, 62  
Memory Location, 33  
Menu Area, 10  
Meter A Measurement, 14  
Meter B Measurement, 15  
Min Max Reading, 23  
Minimum (MIN) Reading, 22  
Multimeter Disturbance, 66  
PAC91, 46  
Parallel Print Cable, 46  
Parts, 43  
Peak, 59  
Performance Characteristics, 55  
Phase, 60  
PM8907, 44  
PM8918 Probe, 41, 46, 56  
PM9080, 34, 36, 46  
Polarity, 26  
Positioning the Waveform, 19  
Power Adapter, 2, 44, 51  
Power Down Timer, 51  
Powering the Test Tool, 7  
Print Cable, 46  
RS-232 Communication Errors, 54  
Printing, 34  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Fluke 123  
Users Manual  
Specifications, 55  
Stable Reading, 16  
Stand, 47  
—S—  
—U—  
Safety, 64  
Unpacking, 2  
Safety Characteristics, 55  
Safety Precautions, 4  
Safety Requirements, 1  
Sampling Rate, 57  
Saving, 32  
User Manual, 45  
Users Manual, 2  
Using a Printer, 34  
Using FlukeView Software, 36  
STL120 Test Leads, 44  
Storing, 37  
SW90W Software, 2, 36, 46  
—T—  
Saving Screens, 32  
SCC 120, 36, 46  
—V—  
Temperature, Environmental, 63  
Temperature Measurement, 60  
Test Leads, 44  
Tilt Stand, 47  
Time, 50  
Scope Modes, 57  
Scope Probes, 41, 46, 56  
Screen Contrast, 49  
Selecting Trigger Parameters, 28  
Sensitivity, 56  
Vertical Accuracy, 56  
Vibration, 63  
Video Line, 31  
Video on A, 57  
Video Signals, 29  
Time Base, 18  
Serial Printer, 34  
Time Base Accuracy, 57  
Time Base Ranges, 57  
Touch Hold® function, 16, 61  
Trace Disturbance, 65  
TrendPlot™ Function, 22, 61  
Trigger, 57  
Service Manual, 43  
Shielded Test Leads, 2, 44  
Shock, 63  
Single Shot, 23  
Slope, 27, 57  
—W—  
Waveform Area, 10  
—Z—  
Slow Signals, 25  
Trigger Icons, 27  
Smooth, 20, 58  
Trigger Level, 27  
Zero Reference, 17  
Soft Carrying Case, 46  
Software, 46  
Software Version, 43  
Trigger Parameters, 28  
Trigger Sensitivity, 57  
Triggering, 27  
70  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Everpure Water System A 10 User Manual
Exmark Lawn Aerator 00 Higher User Manual
Fisher Price Baby Furniture G2808 User Manual
Ford Automobile Accessories HM02 E 10712484 User Manual
Fujitsu Server BX400 S1 User Manual
Gefen Log Splitter GTB MHDMI13 144 User Manual
GE Microwave Oven Over the Range Instructions Microwave Oven User Manual
GE Range 164D2966P079 User Manual
GE Washer WWSE6260BCC User Manual
Giant Bicycle REIGN X2 User Manual