Agilent Technologies Computer Accessories G6600 90006 User Manual

Agilent 355 Sulfur and 255 Nitrogen  
Chemiluminescence Detectors  
Operation and Maintenance  
Manual  
Agilent Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Warnings  
English  
This symbol on the instrument indicates that the user should refer to the man-  
ual for operating instructions.  
WARNING  
Any operation requiring access to the inside of the equipment, could result in  
injury. To avoid potentially dangerous shock, disconnect from power supply  
before opening the equipment.  
WARNING  
WARNING  
WARNING  
For continued protection against fire hazard replace fuse with same type and  
rating.  
This symbol indicates that to comply with European Union Directive  
2002/96/EC for waste electrical and electronic equipment (WEEE), the Ana-  
lyzer should be disposed of separately from standard waste.  
This is a safety Class I product. It must be wired to a mains supply with a protective  
earthing ground incorporated into the power cord. Any interruption of the protective  
conductor, inside or outside the equipment, is likely to make the instrument dangerous.  
Intentional interruption is prohibited.  
WARNING  
If this instrument is used in a manner not specified by Agilent, the protection provided by  
the instrument may be impaired.  
WARNING  
WARNING  
High voltage is present in the instrument when the power cord is connected, even if the  
main power switch is in the standby mode. To avoid potentially dangerous shock, discon-  
nect the power cord before removing the side panels.  
Operation and Maintenance Manual  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized  
by using the instrument in a well-ventilated area and by venting the exhaust of the vac-  
uum pump to a fume hood. The ozone generator should be turned off when the instrument  
is not in use.  
WARNING  
Burner temperature Is extremely hot. Do not touch. Allow to cool before servicing.  
WARNING  
WARNING  
WARNING  
Hydrogen is an extremely flammable gas. Use appropriate care when handling. Inspect  
all connections with a suitable leak detector.  
Oxygen rich environments can promote combustion and even result in spontaneous com-  
bustion under conditions of high pressure and exposure to contamination. Use only oxy-  
gen rated components and ensure that components are oxygen clean prior to use with  
pure oxygen.  
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxi-  
dant sensors or burst their connective tubing.  
WARNING  
4
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Español  
Cualquier operación que requiera acceso al interior del equipo, puede causar  
una lesión. Para evitar peligros potenciales, desconectarlo de la alimentación  
a red antes de abrir el equipo.  
WARNING  
WARNING  
WARNING  
WARNING  
WARNING  
Para protección contínua contra el peligro de fuego, sustituir el fusible por  
uno del mismo tipo y características.  
Este símbolo en el instrumento indica que el usuario debería referirse al man-  
ual para instrucciones de funcionamiento.  
Esto es un producto con clase I de seguridad. Debe conectarse a una red que disponga de  
tierra protectora en el cable de red. Cualquier interrupción del conductor protector,  
dentro o fuera del equipo, puede ser peligroso. Se prohibe la interrupción intencionada.  
Si este instrumento se usa de una forma no especificada por Agilent, puede desactivarse  
la protección suministrada por el instrumento.  
Operation and Maintenance Manual  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Français  
Chaque opération à l'intérieur de l'appareil, peut causer du préjudice. Afin  
d'éviter un shock qui pourrait être dangereux, disconnectez l'appareil du  
réseau avant de l'ouvrir.  
WARNING  
WARNING  
WARNING  
WARNING  
Afin de protéger l'appareil continuellement contre l'incendie, échangez le fus-  
ible par un fusible du même type et valeur.  
Le symbol indique que l'utilisateur doit consulter le manuel d'instructions.  
Ceci est un produit de Classe de sécurité I. L'instrument doit être branché sur  
l'alimentation secteur par un fil de secteur prévu d'une prise de masse. Chaque  
interruption du conducteur protégeant, à l'intérieur ou á l'extérieur de l'appareil peut  
rendre l'instrument dangereux. Interruption intentionnelle est interdite.  
Si l'instrument n'est pas utilisé suivant les instructions de Agilent, les dispositions de  
sécurité de l'appareil ne sont plus valables.  
WARNING  
6
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Deutsch  
Vor dem Öffnen des Gerätes Netzstecker ziehen!  
WARNING  
WARNING  
WARNING  
WARNING  
Für kontinuierlichen Schutz gegen Brandgefahr dürfen bei Sicherungswech-  
sel nur Sicherungen der gleichen Stärke verwendet werden!  
Dieses Symbol auf dem Gerät weist darauf hin, dass der Anwender zuerst das  
entsprechende Kapitel in der Bedienungsanleitung lesen sollte.  
Dies ist ein Gerät der Sicherheitsklasse I und darf nur mit einem Netzkabel mit  
Schutzleiter betrieben werden. Jede Unterbrechung des Schutzleiters au erhalb oder  
innerhalb des Gerätes kann das Gerät elektrisch gefährlich machen. Absichtliches  
Unterbrechen des Schutzleiters ist ausdrücklich verboten.  
Wenn das Gerät nicht wie durch die Firma Agilent, vorgeschrieben und im Handbuch  
beschrieben betrieben wird, können die im Gerät eingebauten Schutzvorrichtungen  
beeinträchtigt werden.  
WARNING  
Operation and Maintenance Manual  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Italiano  
Qualsiasi intervento debba essere effettuato sullo strumento può essere  
potenzialmente pericoloso a causa della corrente elettrica. Il cavo di alimen-  
tazione deve essere staccato dallo strumento prima della sua apertura.  
WARNING  
WARNING  
WARNING  
WARNING  
Per la protezione da rischi da incendio in seguito a corto circuito, sostituire I  
fusibili di protezione con quelli dello stesso tipo e caratteristiche.  
Il simbolo sullo strumento avverte l'utilizzatore di consultare il Manuale di  
Istruzioni alla sezione specifica.  
Questo strumento è conforme alle specifiche per I prodotti in Classe I - Il cavo di  
alimentazione dalla rete deve essere munito di "terra". Qualsiasi interruzione del cavo di  
terra all'interno ed all'esterno dello strumento potrebbe risultare pericolòsa. Sono  
proibite interruzioni intenzionali.  
Se questo strumento viene utilizzato in maniera non conforme alle specifiche di Agilent,  
le protezioni di cui esso è dotato potrebbero essere alterate.  
WARNING  
8
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Dutch  
Iedere handeling binnenin het toestel kan beschadiging veroorzaken. Om ied-  
ere mogelijk gevaarlijke shock te vermijden moet de aansluiting met het net  
verbroken worden, vóór het openen van het toestel.  
WARNING  
WARNING  
WARNING  
WARNING  
Voor een continue bescherming tegen brandgevaar, vervang de zekering door  
een zekering van hetzelfde type en waarde.  
Het symbool geeft aan dat de gebruiker de instructies in de handleiding moet  
raadplegen.  
Dit is een produkt van veiligheidsklasse I. Het toestel moet aangesloten worden op het  
net via een geaard netsnoer. Bij onderbreking van de beschermende geleider, aan de  
binnenzijde of aan de buitenzijde van het toestel, kan gebruik het toestel gevaarlijk  
maken. Opzettelijke onderbreking is verboden.  
Indien het toestel niet gebruikt wordt volgens de richtlijnen van Agilent, gelden de  
veiligheidsvoorzieningen niet meer.  
WARNING  
Operation and Maintenance Manual  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
The following symbols are used on the equipment:  
Caution - Refer to manual for  
operating instructions  
Caution - Risk of electrical  
shock.  
Caution - Hot surface.  
Atención - Ver documentación Atención - Riesgo de sacudidas Atención - Superficie caliente.  
pertinente.  
eléctricas.  
Attention - Consultez les  
ocuments d'accomagnement. électrique.  
Attention - Risque de choc  
Attention - Surface brûlante.  
Vorsicht - Heisse Oberfläche.  
Pericolo - Superficie rovente.  
Voorzichtig - Heet oppervlak.  
Vorsicht - Siehe beiliegende  
Unterlagen.  
Vorsicht - Risiko eines  
Elektroschocks.  
Pericolo - Vedi  
documentazione allegata.  
Pericolo - Pericolo di scosse  
elettriche.  
Voorzichtig - Raadpleeg di  
bijehorende documentatie.  
Voorzichtig - Hoge spanning,  
levensgevaar.  
10  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2
12  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
14  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
16  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors  
Operation and Maintenance Manual  
1
Introduction  
This manual will guide you in the installation, operation, and troubleshooting  
of the Agilent 355 Sulfur Chemiluminescence Detector (SCD) and the Agilent  
255 Nitrogen Chemiluminescence Detector (NCD). This manual is intended for  
use with the Agilent 355 SCD or 255 NCD with the Dual Plasma Burner and  
Controller.  
This operation and service manual has some important conventions, such as  
the use of boxed warnings. This information is deliberately set out from the  
text for emphasis and should be followed to assure operator safety and proper  
instrumental operation.  
If you are installing the 355 SCD or 255 NCD yourself, follow the installation  
procedures described in the following sections. If your instrument is already  
installed, turn to the Operation section to begin.  
17  
Agilent Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
18  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors  
Operation and Maintenance Manual  
2
System Description  
19  
Agilent Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Specifications  
Technical Information — 355 SCD  
Sensitivity*  
Typical < 0.5 pg S/second (signal to noise 3.3:1)  
g S/g C > 2 x 107  
Typical Selectivity  
Linearity  
>104  
Precision and Stability†  
<2% RSD 2 hours  
<5% RSD 72 hours  
Ozone flow through the Post Ozone Restrictor  
Typical reaction cell pressure  
20-30 mL/min at 3-6 psig  
4 - 8 Torr RV5 Oil Sealed Pump  
6 - 12 Torr Dry Piston Pump  
250-400 Torr operating  
800 °C  
Typical Burner Pressure  
Typical Burner Temperature  
Typical Air Flow Rate  
65 SCCM recommended  
3-10 SCCM recommended with FID adapter  
40 SCCM recommended  
0-1V, 0-10V  
Typical Hydrogen Flow Rate  
Signal Output Ranges  
Typical time to reach 800 °C from room  
temperature  
10 min typical(120 VAC, 60 Hz)  
Typical safety shroud outside temperature  
Recorder output  
<65 °C at 800 °C Burner temperature typical  
0-1 V or 0-10 V  
* Burner temperature 800 °C, 80 SCCM air, and 60 SCCM hyrdrogen, test compound dimethyl sulfide in toluene.  
† Based on thiopene in benzene at 1 ppm mass sulfur, 1 µL injection split 1:10, 30 m, 0.32 mm ID, 1 µm thick CP  
wax (n=10 for 2.5 hours; n=42 for 72 hours).  
Subject to change without notice.  
Technical Information — 255 NCD  
Sensitivity  
<3 pg N/second (signal to noise 3:1) in both N  
and nitrosamine modes  
Selectivity  
Linearity  
g N/g C > 2 to 107 in N mode (selectivity in  
nitrosamine mode is matrix-dependent)  
>104  
20  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Repeatability*  
<1.5% RSD 8 hours (~ the same in N and  
nitrosamine)  
<2% RSD 18 hours (~3% RSD in nitrosamine  
mode over 21 hours)  
Gas flow through Ozone Generator  
Typical reaction cell pressure  
20-30 mL/min at 3-6 psig (inlet pressure)  
4 - 8 Torr RV5 Oil Sealed Pump  
6 - 12 Torr Dry Piston Pump  
130 - 150 torr operating  
900 °C  
Typical Burner Pressure  
Typical Burner Temperature  
Typical Hydrogen Flow Rate  
Typical Oxygen Flow Rate  
Signal Output Ranges  
6-10 SCCM  
8 - 15 SCCM  
0-1V, 0-10V, 0 - 10V  
* Burner temperature 950 °C, 11 SCCM oxygen, and 6 SCCM hydrogen; 25 ppm N as nitrobenzene in toluene;  
0.2 µL injection on column (HP 19095-121Z), n=7 for 3 hours; n=13 for 18 hours and n=10  
n-dipropylnitrosamine in toluene at 4 µg/mL, 0.2 µL injection on column (HP 19095-121Z).  
Physical Specifications  
Power requirements  
355 SCD Detector  
115 VAC, 50/60 Hz, 1400 W  
100 VAC, 50/60 Hz, 1400 W  
220-240 VAC, 50/60 Hz, 650W  
255 NCD Detector  
115 VAC, 50/60 Hz, 1400 W  
100 VAC, 50/60 Hz, 1400 W  
220-240 VAC, 50/60 Hz, 650W  
Dual Plasma Controller  
100-120 VAC, 50/60 Hz, 200 W  
220-240 VAC, 50/60 Hz, 200 W  
Dimensions and weight  
Detector  
Height: 16.0 in (40.6 cm)  
Width: 9.2 in (23.4 cm)  
Depth: 21.8 in (55.3 cm)  
355 SCD Weight: 34.0 lbs (15.0 kg)  
255 NCD Weight: 37.5 lbs (17.0 kg)  
Operation and Maintenance Manual  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Dual Plasma Controller  
Height: 5.0 in (12.7 cm)  
Width: 9.5 in (24.1 cm)  
Depth: 12.5 in (31.8 cm)  
Weight: 9.9 lbs (4.5 kg)  
Dual Plasma Burner  
Height: 12.3 in (31.2 cm)  
Diameter: 4.0 in (10.2 cm)  
Weight: 1.9 lbs (0.9 kg)  
Oil Sealed Vacuum Pump (RV5)  
Height: 10.3 in (26.1 cm)  
Width: 6.0 in (15.2 cm)  
Depth: 16.9 in (43.0 cm)  
Weight: 47.3 lbs (21.5 kg)  
or  
Oil Free Dry Piston Pump  
Height: 12.0 in (30 cm)  
Width: 9.0 in (22.9 cm)  
Depth: 14.0 in (35.6 cm)  
Weight: 29.9 lbs (13.6 kg)  
Installation Category  
Pollution Degree  
II  
2
Ambient Temperature  
Relative Humidity  
50 - 104 °F (10-40 °C)  
Up to 95%, noncondensing  
Intended for indoor use only  
2,000 m (6,562 ft)  
Normal Operating Environment  
Maximum Altitude  
Mains Supply Voltage  
Fluctuation not to exceed 10% of nominal  
voltage  
22  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Theory of Operation  
Sulfur Chemiluminescence Detector  
The Agilent model 355 Sulfur Chemiluminescence Detector (SCD) is a  
sulfur-selective detector for gas chromatography. Operation of the SCD is  
based on the chemiluminescence (light-producing reaction) from the reaction  
of ozone with sulfur monoxide (SO) produced from combustion of the analyte:  
Sulfur compound (analyte) SO + H O + other products  
2
SO + O SO + O + hη (<300–400 nm)  
3
2
2
A vacuum pump pulls the combustion products into a reaction cell at low  
pressure, where excess ozone is added. Light produced from the subsequent  
reaction is optically filtered and detected with a blue-sensitive photomultiplier  
tube and the signal is amplified for display or output to a data system. Figure 1  
is a pneumatic flow diagram that illustrates how the components of the system  
are integrated.  
The Detector has an enclosed, dedicated (Dual Plasma) Burner designed to  
enhance production of the SO intermediate. This Dual Plasma Burner mounts  
in the detector port of the GC. A Dual Plasma Controller provides temperature  
control and gas-flow regulation to operate the Dual Plasma Burner.  
The Agilent model 355 SCD provides high sensitivity (<0.5 pg S/sec), with  
linear and equimolar response over four orders of magnitude (per Sulfur  
atom) while maintaining high selectivity over common solvents. The Agilent  
SCD is compatible with most commercially available gas chromatographs.  
Nitrogen Chemiluminescence Detector  
Operation of the Agilent model 255 Nitrogen Chemiluminescence Detector is  
based on the chemiluminescence or light-producing reaction of ozone with  
nitric oxide formed from combustion. Reacting nitric oxide with ozone results  
in the formation of electronically excited nitrogen dioxide. The excited  
nitrogen dioxide emits light, a chemiluminescence reaction, in the red and  
infrared region of the spectrum. The light emitted is directly proportionally to  
the amount of nitrogen in the sample,  
NO + O NO NO + hη (>800 nm)  
3
2
2
The light (hη) emitted by the chemical reaction is optically filtered and  
detected by a photomultiplier tube. The signal from the photomultiplier tube is  
amplified for display or output to a data system. Organic compounds  
containing nitrogen react to form nitric oxide, carbon dioxide, and water.  
H2/O2  
---------------  
R-N + O  
NO + CO2 + H2O  
2
Δ
Operation and Maintenance Manual  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Dual Plasma Controller  
The Dual Plasma Controller provides all operational parameters of the Dual  
Plasma Burner except for the Detector base temperature. The Detector base  
temperature is controlled by circuitry in the GC. Parameters monitored or  
regulated by the Controller include Burner temperature, Burner temperature  
set-point, hydrogen and oxidant flow rates, and Burner pressure. The  
temperature set-point, actual pressure, oxidant and hydrogen flow rates are  
displayed by rotation of a 4-position switch. Power, valve operation,  
temperature within set-point range and fault conditions are indicated with  
LED illumination on the front display panel.  
The Dual Plasma Controller incorporates several safety features. The safety  
circuitry detects faults such as power loss, vacuum loss, thermocouple failure,  
heater element failure, broken ceramic tube, or high temperature. When a fault  
is detected, the Fault LED illuminates and hydrogen and oxidant flow is  
stopped by normally-closed solenoid valves.  
Dual Plasma Burner with the 355 SCD  
The Dual Plasma Burner is based on the same chemistry and basic principles  
of earlier SCD Burner designs. A key difference, however, is the addition of a  
second “flame” or “plasma,” the lower is oxygen-rich and the upper is  
hydrogen-rich. The Burner consists of a tower assembly that contains an outer  
sheath for burn protection, a heating element, thermocouple, and combustion  
tubes. Conversion of sulfur containing compounds to SO occurs within the  
ceramic reaction chamber housed in the Burner assembly. Agilent also  
provides a Flame Ionization Detector (FID) option for the simultaneous  
collection of hydrocarbon and sulfur chromatograms for some GCs.  
Dual Plasma Burner with the 255 NCD  
Compounds eluted from the GC column are combusted in the Dual Plasma  
Burner first by an oxygen rich flame (plasma) followed by catalytic combustion  
on a Noble metal screen. For hydrocarbons, this two stage combustion  
technique results in complete conversion of the matrix to products, such as  
carbon dioxide and water, which do not chemiluminesce with ozone. Nitrogen  
atoms in a compound are converted into nitric oxide and potentially other  
nitrogen oxide species. In the second stage, the catalyst is used to convert  
other nitrogen oxide species to nitric oxide, resulting in a high efficiency of  
conversion.  
24  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Description of Major Components  
Dual Plasma Burner  
The Dual Plasma Burner consists of a tower assembly that contains an outer  
sheath for burn protection, a heating element, thermocouple, and combustion  
tubes. In the SCD, conversion of sulfur containing compounds to SO occurs  
within the ceramic reaction chamber housed in the Burner assembly and  
potentially interfering hydrocarbons are oxidized to CO and H O, with air as  
2
2
the oxidant. In the NCD, oxygen is used as the oxidant.  
A fitting is located on top of the Burner. The vacuum line from the Detector  
box is connected directly to the top of this fitting and H2 is input to the longer  
side of this fitting. The air inlet is connected to the base of the Burner.  
The Burner is mounted onto the GC by a model-specific mounting kit (see  
www.Agilent.com/chem or contact Agilent for the most current information).  
The GC column is inserted into the Burner using a 1/32" knurled nut and fused  
silica adapter ferrule.  
A cross-section illustration of the Dual Plasma Burner for the 355 SCD is  
shown in Figure 1 and for the 255 NCD is shown in Figure 2.  
Operation and Maintenance Manual  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Figure 1 Cross-Section of the Dual Plasma Burner for the 355 SCD  
26  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Figure 2 Cross-Section of the Dual Plasma Burner for the 255 NCD  
Dual Plasma Controller  
The Dual Plasma Controller powers the Dual Plasma Burner and supplies its  
gases. Hydrogen and oxidant should be provided at 25 psig (1.72 bar) and the  
Controller is plugged into an appropriate AC electrical outlet.  
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxi-  
dant sensors or burst their connective tubing.  
WARNING  
The parameters monitored or regulated by the Controller include Burner  
Operation and Maintenance Manual  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
temperature, hydrogen and oxidant flow rates, and Burner pressure. The  
temperature, actual pressure, oxidant and hydrogen flow rates are selected for  
display by rotation of a 4-position control knob. Power, valve operation,  
temperature within set-point range and fault conditions are indicated with  
LED illumination on the front display panel.  
The Dual Plasma Controller incorporates several safety features. The safety  
circuitry detects faults such as power loss, vacuum loss, thermocouple failure,  
heater element failure, broken ceramic tube, or high temperature. When a fault  
is detected, the Fault LED illuminates and hydrogen and oxidant flow is  
stopped by normally-closed solenoid valves.  
Ozone Generator  
The SCD and NCD produce ozone by corona discharge using a clean,  
pressurized air or oxygen source. Use of oxygen should increase ozone  
production and, hence, Detector response. High voltage to the ozone generator  
is applied only when the reaction cell pressure is less than 100 torr in the SCD  
and less than 200 torr in the NCD. Gas flow through the ozone generator is  
controlled by a pressure regulator and flow restrictors.  
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized by  
using the instrument in a well ventilated area, changing the chemical trap regularly, and  
venting the exhaust of the vacuum pump. The ozone generator should be turned off when  
the instrument is not in use to reduce maintenance requirements.  
WARNING  
Chemiluminescence Reaction Cell and Photomultiplier Tube  
Sulfur monoxide (formed in the Burner) and ozone (produced in the ozone  
generator located in the Detector) are mixed in the reaction cell. The cell is  
designed such that the reaction between SO and O occurs directly in front of  
3
the photomultiplier tube (PMT). A UV band pass filter (300 - 400 nm) located  
between the reaction cell and the PMT selectively transmits the light emitted  
by the SO/O reaction. Efficient combustion in the ceramic tubes coupled with  
3
the UV band pass filter eliminates interference from non-sulfur containing  
analytes (e.g. nitric oxide, olefins, etc.) which also undergo chemiluminescent  
reactions with ozone. A background signal is typically present as a result of  
ozone-wall interactions and low level sulfur contamination of Detector gases.  
This background signal can be used as a troubleshooting aid (see Section 10).  
28  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
10  
5
9
4
3
6
6
2
7
1
8
1. Ozone Generator  
6. Vacuum Line  
7. Particulate Filter  
2. High Voltage Transformer  
3. Photomultiplier Tube Socket  
8. Pre-Ozone Restrictor  
9. Post-Ozone Restrictor  
10. Pressure Transducer  
4. Photomultiplier Housing  
5. Reaction Cell  
Figure 3 355 SCD Left Side  
Operation and Maintenance Manual  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
18  
17  
IN OFST  
RV1  
OUT OFST  
RV2  
11  
I
AMP  
TP10  
1
TP8  
TP12  
TP2  
TP13  
OUT  
TP5 TP4TP6 TP7  
TP2  
TP3  
TP4  
TP3  
ALCO  
13  
TP5  
TP1  
TP12  
12  
HV/100  
TP11  
TP8  
TP13  
TO  
GND  
FRONT  
TP1  
PANEL  
TP14  
TP9  
TP9  
USA  
TP10  
TP6  
TP7  
TP16 TP15  
TP11  
SIEVERS RESEARC
ELECTRONICS  
16  
RV1  
HV  
LOAD  
P4  
NEUT  
P5  
PUMP  
OUT  
WHT  
OZONE  
PUMP  
LOAD  
P6  
NEUT  
P7  
OZONE  
GEN  
WHT  
15  
LINE  
P2  
NEUT  
P3  
AC  
WHT  
14  
11. Amplifier Cable  
12. HV Cable  
15. Fuses  
16. Pressure Regulator  
17. Transfer line  
13. PMT Amplifier  
14. EMI Filter  
18. Front panel display  
Figure 4 355 SCD Right Side  
30  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Figure 5 255 NCD Left Side View  
Operation and Maintenance Manual  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Figure 6 255 NCD Right Side View  
Pressure Transducer  
Vacuum in the reaction cell is measured using a pressure transducer. The  
pressure of the reaction cell can be monitored from the front panel and will  
typically range from 5 to 10 torr depending on the type of vacuum pump used,  
condition of the chemical (ozone) trap, ceramic tube position and the  
condition of the combustion chamber. The range of response is from 0 to 760  
torr.  
Vacuum Pump  
There are two choices of vacuum pumps for the 355 SCD and 255 NCD. A  
two-stage, oil-sealed rotary vacuum pump is used to produce an operating  
pressure between 3 and 10 torr in the reaction cell. The oil-free Dry Piston  
pump produces a vacuum between 5 and 12 torr. All vacuum pumps serve the  
same purpose:  
32  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Collection and transfer of the combustion gases from the Burner to the  
reaction cell.  
Transfer of ozone from the ozone generator to the reaction cell.  
Reduction of non-radiative collisional quenching of the emitting species in  
the chemiluminescent reaction cell.  
The higher vacuum produced by the oil-sealed rotary pump facilitates a  
shorter residence time in the reaction chamber, and therefore reduces the  
incidence of collisional relaxation of the excited SO . This results in slightly  
2
lower detection limits using the oil-sealed pump as compared to the oil-free  
pump.  
Chemical Trap  
To protect the vacuum pump from exposure to ozone, a chemical trap is  
located at the inlet of the vacuum pump. This trap contains a catalyst that  
converts ozone to diatomic oxygen. This trap is a consumable part and should  
be replaced every 90 days of continuous Detector usage.  
Oil Coalescing Filter  
The oil-sealed rotary vacuum pump is operated with the gas ballast partially  
open to aid in the elimination of water produced in the Burner and transferred  
to the pump. As a result of the open gas ballast and the high flow rates of  
gases, oil vaporized in the pump can escape through the pump exhaust. To  
minimize oil loss, an oil coalescing filter is placed on the pump exhaust to trap  
vaporized oil and to return this oil to the vacuum pump oil reservoir. This is  
not necessary for use with the oil-free pump. A replaceable element in the  
filter is a consumable part and should be replaced every 90 days of continuous  
use.  
FID Adapter (Optional)  
The Agilent 355 Sulfur Chemiluminescence Detector is designed to mount into  
most GC detector ports and operate as a stand-alone sulfur detector. For some  
applications it is desirable to monitor both sulfur and hydrocarbon  
components using a single column without splitting. For this reason, Agilent  
offers a few adapter kits to mount the Dual Plasma Burner onto a Flame  
Ionization Detector for the simultaneous collection of FID and SCD  
chromatograms. During dedicated SCD operation, 100% of the column effluent  
passes through the Burner to the Detector. During simultaneous detection  
approximately 10-20% of the FID exhaust gases are drawn into the Burner  
through a restrictor, which reduces sulfur sensitivity to approximately 1/10 of  
the signal observed in a dedicated SCD Burner.  
NCD Reaction Cell  
The model 255 NCD reacts nitric oxide from the Burner and ozone from the  
Ozone Generator in the Chemiluminescence Reaction Cell. The reaction occurs  
Operation and Maintenance Manual  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
directly in front of the photomultiplier tube (PMT). A red cut-off filter between  
the reaction cell and the PMT selectivity transmits the light from the nitric  
oxide and ozone reaction.  
Efficient combustion in the ceramic tubes and the red cut-off filter eliminate  
interference from non-nitrogen containing analytes (sulfur dioxide, alkenes,  
olefins) that have chemiluminescence reactions with ozone.  
A background signal is typically present as a result of ozone-wall interactions  
and low level nitrogen contamination from carrier and detector gases. The  
background signal is a useful troubleshooting aid.  
NCD Photomultiplier Tube and Cooler  
In the NCD, a red-sensitive PMT detects emissions from the nitric oxide and  
ozone reaction. A thermoelectric cooler cools the PMT 40 °C below ambient  
temperature to approximately -15 to -20 °C to reduce background noise from  
the PMT. The cooler operates whenever the Detector is connected to AC power.  
34  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
25 psig max  
25 psig max  
S
S
Mass  
flow  
NC  
NC  
F.C.  
F.C.  
Mass  
flow  
Air  
H2  
P
Dual Plasma Controller  
Dual  
Plasma  
Burner  
355 SCD  
Transfer  
Line  
Column  
P
Reaction  
Cell  
PMT  
Gas Chromatograph  
Vent  
Figure 7 Schematic for 355 SCD  
Operation and Maintenance Manual  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
25 psig max  
25 psig max  
S
S
Mass  
flow  
NC  
NC  
F.C.  
F.C.  
Mass  
flow  
O2  
H2  
P
Dual Plasma Controller  
Dual  
Plasma  
Burner  
255 NCD  
Transfer  
Line  
Column  
P
Reaction  
Cell  
PMT  
Gas Chromatograph  
Vent  
Figure 8 Schematic for 255 NCD, in Nitrogen Mode  
36  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
25 psig max  
S
S
Mass  
flow  
NC  
NC  
F.C.  
F.C.  
Mass  
flow  
O2*  
P
*Or O2, He, N2  
Dual Plasma Controller  
Dual  
Plasma  
Burner  
255 NCD  
Transfer  
Line  
Column  
P
Reaction  
Cell  
PMT  
Gas Chromatograph  
Vent  
Vacuum Pump  
Figure 9 Schematic for 255 NCD, in Nitrosamine Mode  
Operation and Maintenance Manual  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
38  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Overview  
Installation and start-up of the Agilent 355 SCD or 255 NCD by a qualified  
Agilent Service technician is recommended. If you choose to install the  
detector yourself, carefully read all of this chapter prior to installation of the  
instrument.  
Although every reasonable safeguard against shipping damage has been taken,  
product damage may still occur due to excessive mishandling. If obvious  
damage has occurred during shipment, contact Agilent. Shipping materials for  
the 355 SCD or 255 NCD should be saved. If the instrument must be returned  
to the factory, it must be packed in the original carton to reduce the chance for  
damage during shipment. Replacement shipping containers can be purchased  
from Agilent.  
Substituting parts or performing unauthorized modification to the instrument may result  
in a safety hazard.  
WARNING  
WARNING  
Hydrogen is a flammable gas. Perform periodic leak tests to verify there are no leaks in  
the hydrogen lines and connections. Before making any connections, shut off the hydro-  
gen supply. Connect or cap all fittings at all times when using hydrogen.  
40  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Step 1: Selecting a Location  
The instrument should be placed on a clean, unobstructed surface  
approximately 22" (55 cm) deep by 10" (24 cm) wide that can support 44  
pounds (19.9 kg) in addition to existing equipment. Figure 10 illustrates the  
relationship between the major system components. To facilitate proper heat  
dissipation, an additional 1-2" (2.5-5.0 cm) should be available at the rear and  
on both sides of the instrument. Distance between the SCD and the GC is  
limited by the length of gas and heater lines.  
Position the Controller on top of the Detector box or in another convenient  
location near both the Detector and GC. The Controller requires supplies of  
hydrogen and oxidant (air for the SCD and oxygen for the NCD). The length of  
the gas inlet lines limit the distance between the Controller and the Burner to  
approximately 1 meter.  
There are two pump options for the Detector, each with different size  
requirements. The Edwards RV5 oil-sealed vacuum pump requires a 6.3"  
(16cm) by 18.5" (47cm) area on a shelf or a nearby floor and a minimum height  
clearance of 22" (58 cm). The pump weight is 52 lbs (24 kg). The Dry Piston  
Pump (oil-free) requires a 9" (22 cm) by 14" (35 cm) area on a shelf or a nearby  
floor and a minimum height clearance of 12" (30 cm). The pump weight is  
30 lbs (13.6 kg). The distance to the 355 Detector for both pump options is  
limited by the power cord connection—6 ft (2 m) for the oil-free piston pump or  
8 ft (2.4 m) for the oil-sealed vacuum pump.  
Consider placing the oil-sealed pump over a plastic or metal container to capture any oil  
leaks or spills.  
NOTE  
Power Requirements  
See Chapter 2 for voltage and power requirements. Grounded outlets are  
required.  
Operation and Maintenance Manual  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Figure 10 Drawing of the Detector with Dual Plasma Burner and Controller  
Environmental Considerations  
The instrument should be operated in an environment which is comfortable  
for human habitation with reasonably constant temperature and humidity.  
Operation of the instrument at elevated temperatures (>30 °C) may result in  
an increased background noise from the photomultiplier tube.  
Combustion Gas Requirements  
The Agilent 355 SCD and 255 NCD both require a hydrogen source; the 355  
SCD requires an air source, and the 255 NCD requires an oxygen source.  
Oxygen may be used for the ozone generator in either Detector to obtain a  
modest increase in sensitivity; the Dual Plasma Burner was designed for use  
42  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
with air as the oxidant in the SCD and oxygen as the oxidant in the NCD.  
For the SCD, detector gases must be sulfur free (<1ppb) for proper Detector  
operation. In general, bottled air is preferred to “house” air from a compressor  
because compressors tend to generate lower quality air. In addition, pressure  
fluctuations induced by the air generator may be detrimental to Detector  
performance.  
For the NCD and GC, use gases that are low in nitrogen and total  
hydrocarbons. Use gases with an "instrument" or "chromatographic" purity  
rating. Agilent recommends gases with a purity range of 99.995% to 99.9995%.  
Do not use oil-pumped supplied air since it contains large concentrations of  
hydrocarbons.  
Use of appropriate traps on carrier, hydrogen and oxidant sources to improve  
support gas quality, such as a sulfur trap gas purifier, is recommended. The  
oxidant used to supply the ozone generator should be dry to prevent internal  
corrosion. Use of a moisture trap or dryer is recommended.  
Two-stage pressure regulators  
The combustion gases must be supplied to the Controller at a pressure of  
25 psig or less. Use a two-stage pressure regulator rather than a single-stage  
regulator to eliminate pressure surges. High-quality, stainless-steel  
diaphragm-type regulators are recommended. On/Off valves at the regulators  
are useful but are not essential. Mount the valves at the outlet fitting of the  
pressure regulators.  
If you use gases with a two-stage regulator up-stream, it will be sufficient to  
use single-stage regulators at the Detector.  
Particle Filtration  
Up-stream, inline particle filtration of 10 micron or better is required to  
prevent damage to Controller and Burner components.  
Ozone is a hazardous gas and a strong oxidant. Minimize exposure to ozone by using the  
instrument in a well-ventilated area and venting the exhaust of the vacuum pump to a  
fume hood. Turn off the ozone generator when the instrument is not in use.  
WARNING  
Supply Tubing for Combustion Gases  
Use preconditioned and cleaned copper or stainless steel tubing to supply  
gases to the 355 SCD and 255 NCD. Do not use ordinary copper tubing since it  
contains oils and other contaminants. Plastic tubing is not recommended since  
it is permeable to oxygen and other contaminants that can damage columns  
and detectors or cause elevated background. Teflon tubing may be acceptable  
in some clean environments.  
Operation and Maintenance Manual  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Secure all gas cylinders to an immovable structure or permanent wall. Store compressed  
gases in accordance with all safety codes.  
WARNING  
WARNING  
Wear eye protection when using compressed gas to avoid possible eye injury.  
Venting Gases  
During normal operation of the GC with the SCD, NCD, FID, other detectors  
and split/splitless inlet purge, some of the carrier gas and sample vents  
outside the GC. In addition, the vacuum pump will vent a small amount of  
ozone and other combustion products. If the components of the sample are  
toxic or noxious, vent the exhaust from the GC outlets to a fume hood. Also,  
vent the exhaust from the vacuum pump to a fume hood.  
44  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Step 2: Unpack and Inspect the Instrument  
Before unpacking boxes, inspect them for signs of physical damage. If damage  
is observed, photographs should be taken in order to make a claim with the  
carrier should any equipment damage be found. Check contents of boxes  
against the shipping documents. Contact Agilent as soon as possible should a  
discrepancy be found.  
Required Installation Tools  
The following tools are recommended for installation of both the 355 SCD and  
255 NCD:  
One adjustable wrench, with adjustment up to 1 inch (2.5 cm)  
One 1/4"-5/16" open-end wrench  
Two 3/8"-7/16" open-end wrenches  
Two 3/4"-9/16" open-end wrenches  
One 5/8" open-end wrench  
One 1/4" Nutdriver with hollow shaft (for FID restrictor)  
One Phillips head screwdriver  
One pair cotton gloves (for hand protection)  
One small flat-head screwdriver  
Operation and Maintenance Manual  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Step 3: Set Up the Vacuum Pump  
Initial connections require access to the rear of the Detector and the vacuum  
pump. Follow the installation instructions for either the RV5 Edwards pump  
or the oil-free dry piston pump, depending on your configuration.  
Exhaust gases from the pump should be vented to a fume hood to eliminate any potential  
hazard.  
WARNING  
Installing the Edwards RV5 Pump Oil-Sealed Vacuum Pump  
The normal surface temperature of the pump body at ultimate vacuum (operation) at  
WARNING  
ambient temperature of 20 °C is 50 °C to 70 °C. If you use the pump at a high ambient tem-  
perature, the temperature of the pump body may exceed 70 °C, and you must fit suitable  
guards to prevent contact with hot surfaces. For more information refer to the pump oper-  
ating manual.  
Do not operate the vacuum pump with the oil level below the minimum oil level mark or  
above the maximum oil level mark.  
WARNING  
1 Install the chemical trap mounting bracket to the top of the pump with the  
screws and provided wrench. The chemical trap is a white plastic cylinder  
approximately 1 foot (30 cm) in length by 1.5 inches in (3.8 cm) diameter  
fitted with a barbed fitting on both ends (see Figure 11).  
2 Remove the protective cap from the pump inlet port.  
3 Install the metal conical screen and black centering o-ring into the pump  
inlet port (see Figure 13).  
4 Attach the aluminum 1/2-inch barbed adapter to the pump inlet port with  
the metal clamping ring.  
5 Slide the 2 inches Tygon vacuum hose over the pump inlet barbed adapter  
and secure with a hose clamp.  
6 Slide another hose clamp over the 2 inches Tygon vacuum hose.  
7 Remove the plastic caps from the barb fittings on the ends of the chemical  
trap.  
8 Press the trap into the mounting bracket and the 2 inches of Tygon vacuum  
hose (see Figure 11 and Figure 12).  
9 Tighten the hose clamp.  
10 Locate the Oil Return Line Kit.  
46  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
11 Remove the drain plug and the bonded seal from the oil mist filter. The  
bonded seal looks like a metal washer with a black inner o-ring.  
12 Install the bonded seal to the oil mist filter drain adapter. The drain adapter  
looks like a drain plug with a small plastic nozzle.  
13 Screw the black drain adapter and bonded seal into the oil mist filter.  
14 Remove the plastic protective cover from the pump exhaust port. Place the  
centering o-ring on the pump exhaust port (see Figure 13).  
15 Place the oil mist filter onto the o-ring on the pump exhaust port. Install the  
filter so that the gray half is above the white half of the filter (see Figure 11  
and Figure 12).  
16 Position the oil mist filter so the drain adapter points toward the gas ballast  
inlet (see Figure 11 and Figure 12).  
17 Fit the clamping ring onto the adapter and oil mist filter and hand tighten.  
18 Turn the gray plastic gas ballast knob counterclockwise to position II. Press  
the knob down against the spring and continue to turn counterclockwise  
until the knob is free. Remove knob from the pump. Confirm that the  
spring is still in place.  
19 Locate the tall aluminum gas ballast control assembly, and install the small  
o-ring into the groove on the shaft.  
20 Insert the gas ballast control assembly into the pump, pressing down  
against the spring, and turn it clockwise until the nozzle on the assembly is  
directly over the mark on the top of the pump.  
21 Cut approximately 3/4 of the black silicone oil return line and insert the  
steel restrictor approximately half-way into the line. If needed, use a small  
screwdriver or other small tool to aid in positioning the restrictor.  
22 Fit one end of the oil return line to the drain adapter on the oil mist filter.  
Fit the other end of the line to the nozzle on the gas ballast adapter. Ensure  
that the tubing is not tight and has no tight bends. Secure the line at each  
end, using the black hose clips provided.  
23 Add oil to the pump via either of the two oil fill caps. The oil level should be  
between one-third and one-half when viewed in the oil sight glass. Replace  
the oil fill cap prior to operation of the pump.  
Do not operate the vacuum pump with the oil level below the minimum oil level mark or  
above the maximum oil level mark.  
WARNING  
24 Place a hose clamp over the black heat shrink end of the 6’ Tygon vacuum  
hose, and connect the hose to the barbed fitting labeled Exhaust on the  
back of the Detector. Tighten the hose clamp securely.  
25 The SCD should be placed near the GC and be accessible from the rear in  
order to connect the electrical power and the recorder signal cable.  
Operation and Maintenance Manual  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
26 The vacuum pump should be placed within approximately 3 feet of the  
Detector (elevation not important), in order to connect the vacuum hose  
from the back of the Detector to the chemical trap. A hose clamp should be  
placed over the hose, and the hose should be connected to the straight end  
of the chemical trap. Tighten the hose clamp securely.  
Vent the exhaust gas from the vacuum pump to fume hood to eliminate any potential haz-  
ard.  
WARNING  
27 Place the vacuum pump in an adequately ventilated area or connect an  
exhaust line (not provided) to the outlet located at the top of the mist filter.  
Attach an aluminum 1/2-inch barbed adapter to the outlet with the metal  
clamping ring and another centering o-ring. Secure the exhaust line with a  
hose clamp (not provided) to the aluminum adapter. Route the exhaust line  
to a fume hood or other suitable discharge location.  
28 Shape the exhaust line to capture excess water (see Figure 14). Do not allow  
water to condense in the exhaust line and drip back into the coalescing  
filter. Alternatively, install a water trap using a vacuum flask after the  
coalescing filter to capture excess water and prevent water from dripping  
back into the coalescing filter.  
It is imperative that water does not condense in the exhaust line and fall back into the  
mist filter. It is recommended that transparent 3/8 inch (0.95 cm) id tubing be used as an  
exhaust line.  
WARNING  
48  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Chemical  
Trap  
Ballast control  
Oil return line  
Oil mist  
filter  
Mounting  
Bracket  
Power switch  
EDWARDS  
Edwards  
Mode selector  
Figure 11 RV5 Oil-Sealed Vacuum Pump and Associated Traps (Front Side)  
Chemical Trap  
Ballast Control  
Barbed VacuumInle  
Oil Return  
Line  
Oil Mist  
Filter  
EDWARDS  
Power  
Plug  
5
Figure 12 RV5 Oil-Sealed Vacuum Pump and Associated Traps (Back Side)  
Operation and Maintenance Manual  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Gas ballast  
Inlet &valve  
Pump exhaust  
Retractable handle  
Chemical  
Trap  
Oil fill caps  
Pump inlet  
Figure 13 RV5 Oil- Sealed Vacuum Pump and Associated Traps (Top)  
Sometimes water condensation and accumulation are visible in the exhaust line. This is  
normal. However, do not allow water to continue to accumulate after approximately one  
week of operation. Significant water accumulation may indicate improper pump operation,  
an improperly vented exhaust line, or a Burner leak. Water accumulation in the exhaust line  
can cause damage to the pump, especially if allowed to fall back into the pump. Contact  
Agilent for advice if water accumulation continues to occur.  
NOTE  
50  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Figure 14 RV-5 Oil-S ealed Vacuum Pump Exhaust Line  
Setting the Gas Ballast Position (RV5)  
Set the mode selector halfway between the High Vacuum mode, the small 6  
symbol (see Figure 14) and the High Throughput mode. Do not set the mode  
selector to the High Throughput mode, the large 6 symbol.  
The RV5 vacuum pump and the Oil Drain Kit with ballast flow control ensure  
the vacuum pump operates continuously with a gas ballast flow. The purpose  
of the ballast control is to sweep ambient air into the pump oil. The air purges  
the water (created from the combustion of air and hydrogen in the Burner)  
and the oil (vaporized by the pump) into the oil coalescing filter. The filter  
separates the oil from the water, vents the air and water, and returns the oil to  
the vacuum pump.  
The Oil Drain Kit with the ballast control continuously returns trapped oil in  
the oil mist filter to the vacuum pump. This feature reduces oil loss from the  
pump and minimizes the need to refill the pump with oil.  
The Oil Drain Kit with the ballast control supplied is configured so that the gas  
ballast flow rate is equivalent to that with the gas ballast control on the pump  
in position II. For most applications, there is no need to change the gas ballast  
flow rate. If required, the gas ballast flow can be adjusted using the following  
procedure.  
1 The restrictor plate on top of the aluminum ballast control has three  
screws. Remove the three screws that secure the restrictor plate. Do not  
dismantle the assembly (see Figure 16).  
2 The restrictor plate has circular indentations. The position of the  
indentations with respect to the indentation on the side of the oil return  
assembly identifies the gas ballast flow setting. Turn the restrictor to the  
required position:  
Operation and Maintenance Manual  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
To select no gas ballast flow (not recommended), turn the restrictor  
plate so that none of the indentations are aligned with the indentation  
on the side of the oil return assembly.  
To select low gas ballast flow, turn the restrictor plate so that the single  
indentation on the restrictor plate is aligned with the indentation on the  
side of the oil return assembly.  
To select high gas ballast flow (the setting the ballast control is shipped  
from the factory), turn the restrictor plate so that the two indentations  
are aligned with the indentation on the side of the oil return assembly.  
Installing the Welch Dry Piston Vacuum Pump  
The Welch Dry Piston Pump may be used as a direct replacement for an  
oil-sealed vacuum pump for the NCD or SCD. This pump produces all the  
advantages of an oil-free pump with little or no loss in instrument  
performance, however, operating Detector and Burner pressures will typically  
be a few Torr higher than those obtained with the oil-sealed pump.  
To install the dry piston vacuum pump, follow these steps:  
1 Remove the pump from its packaging and place on an accessible work  
surface with the power cord side nearest to you and the pump "INLET" port  
to your right. Refer to Figure 15 as a visual aid to these instructions.  
2 Verify that the voltage of the pump matches that of the Detector.  
3 Open the plastic bag containing the Dry Piston Pump Kit. Locate the brass  
elbow and brass barb (3/8" NPT). Remove their plastic protective caps and  
wrap three or four turns of Teflon tape (not supplied) onto the threaded  
connections. Screw the barb fitting into the elbow and tighten using  
wrenches. Similarly, remove the plastic cap from the pump's inlet fitting  
and screw elbow into it. Tighten the elbow so that the barb ends up pointed  
parallel to the floor and pointed toward the power cord side of the pump.  
4 Locate the mounting brackets, 4 small screws and washers and spring clips.  
Use 2 screws and washers to attach the spring clips, one to each bracket.  
5 Using the supplied Torx wrench, remove two Torx screws from the pump,  
the right corner screw located closest to you and the closest screw to you  
located immediately to the left of the pump's handle.  
6 Attach the brackets to these locations using the Torx screws as showing in  
7 Attach the short piece of ½" ID hose tubing supplied in the Dry Piston  
Pump Kit to the brass barb and use one of the supplied hose clamps to  
tighten it onto the barb.  
8 Place a second hose clamp over the other end of the plastic tube. Insert the  
barb on the elbow end of a Chemical Trap into the open end of the short  
plastic hose that was just attached to the brass barb on the pump.  
9 Rotate the Chemical Trap so that it is retained by the spring clips.  
52  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
10 Tighten the hose clamp located on the plastic tube and the elbow barb end  
of the Chemical Trap.  
11 Remove the plastic cap from the inlet end of the Chemical Trap. Taking the  
clear end of the vacuum hose and another hose clamp attach the vacuum  
hose to the inlet end of the Chemical Trap and tighten the hose clamp.  
12 Attach the black end of the vacuum hose to the barb fitting located on the  
back of the Detector box and tighten the last hose clamp onto it to make this  
connection.  
13 Using three or four wraps of Teflon® tape around NPT fitting of the  
supplied muffler or user supplied exhaust fitting to the pump's "EXHAUST"  
port. Any exhaust tubing attached to the pump exhaust should be  
positioned to avoid the accumulation of water from condensation of water  
vapor.  
Vent the exhaust from the vacuum pump to a fume hood or other exhaust system to elimi-  
nate any potential hazard.  
WARNING  
14 Place the vacuum pump where it will be located when the pump is  
operated, within the length of the power cord to be attached to the female  
connector on the back of the Detector.  
15 Make sure that the On/Off switch on the pump is in the on position.  
Operation Notes (Welch Pump)  
1 The pump is designed to start against atmospheric pressure. Therefore, if  
the pump is turned off inadvertently, it may be necessary to allow the  
Detector pressure to rise before restarting the pump.  
2 The only user maintenance required is the periodic replacement of the  
piston sleeves and seals, e.g., after 8 months of continuous operation.  
Consult the pump’s manual for further details.  
3 Although, the pump is not damaged by exposure to ozone, use of the  
Chemical Trap is highly recommended. Refer to the pump manual and seal  
replacement kit instructions for more detailed information.  
4 High line voltage will cause the pump to overheat and trip its thermal fuse.  
5 When new, or after changing seals, the pump may require a short break-in  
period (2-4 hours).  
Operation and Maintenance Manual  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Figure 15 The Welch Dry Piston Pump  
Screws  
Indentation  
(Low Ballast)  
Indentation  
(High Ballast)  
Restrictor Plate  
Indentation  
(Alignment Guide)  
Figure 16 Oil Drain Kit with Ballast Control  
6 Turn the switch on the vacuum pump to the On position.  
54  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Note the position of the oil level in the window after operating the pump for  
several hours. For the next several days of operation, recheck the oil level  
daily. If the oil level is increasing, water is accumulating in the oil reservoir.  
Confirm that the water vapor is properly being expelled from the oil mist  
filter. If the oil level is dropping, excess air flow through the pump is forcing  
oil out of the vacuum pump. Turn the restrictor plate so that the single  
indentation on the restrictor plate is aligned with the indentation on the  
side of the oil return assembly. After adjustment of the gas ballast, allow the  
system to operate for an additional day and check the oil level again.  
It is normal for the oil to appear foamy from air in the oil when viewing the oil  
in the oil level window. The purpose of the ballast control is to sweep ambient  
air into the pump oil. Ensure the oil level when the pump is operating is not  
above the “Full” mark on the pump.  
Operation and Maintenance Manual  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Step 4: Connect the Power Cord  
Connect the pump power cord to the female socket on the back of the Detector  
(see Figure 17). The pump has an On/Off switch located on the electrical motor  
and this switch should be turned On. Do not connect to the AC power supply at  
this point in the installation procedure.  
AIR INLET  
RECORDER  
OUTPUT  
10V  
EXHAUST  
1V  
100mV  
FUSE 250 VOLTS  
10 AMPS  
VACUUM  
PUMP  
Voltage Label  
Figure 17 SCD (230 V Unit) Rear Panel Diagram  
56  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Step 5: Install the Dry Compressed Air or O2 Supply  
Connect a 1/8" OD Teflon (PFA) line fitted with a 1/8" brass Swagelok nut from  
the AIR INLET at the rear of the Detector (see Figure 17) to a supply of dry  
compressed air or oxygen. The air regulator located inside the front door of the  
Detector should be set to approximately 3-6 psi.  
Operation and Maintenance Manual  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Step 6: Install the Signal Output Cables  
Signal output cables are available from Agilent as standard equipment and can  
be used with most data systems. Confirm that the output cable supplied is  
correct for your system. A standard cable fitted with two crimp lug connectors  
is supplied for use with most integrators, recorders or data systems. Attach  
the BNC connector end of the recorder cable to the matching output  
connector, labeled RECORDER OUTPUT, on the back of the SCD (see  
Standard Cable Connection  
The standard recorder cable is connected to the integrator by installing the  
red crimp-lug connector to the signal terminal (+) and the black crimp-lug  
connector to the negative terminal (-). No additional ground connection is  
required.  
HP 3390 Series Integrator Cable Connection  
The keyed-edge connector on the HP 3390 series cable is attached to the signal  
input of HP 3390 series integrators. Note that the connector can only be  
installed one way.  
HP 3396 Integrator Cable Connection  
The jack plug on the HP 3396 cable is attached to the analog signal input  
connector at the rear of the integrator. This cable also works with the Agilent  
35900 controller.  
HP 5890 GC Analog Input Board  
A cable made specifically for this input board is available. This board is used  
when the SCD signal is input to ChemStation.  
Agilent 6890 GC Analog Input Board  
A cable made specifically for this input board is available. This board is used  
when the SCD signal is input to ChemStation.  
58  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Step 7: Install the Dual Plasma Burner  
Remove the cover plates from the Detector area of the GC to expose the hole  
into the oven through which a Detector is normally mounted. If the GC has  
more than one available Detector position, pick the most convenient one.  
Prepare the GC by cutting the inside liner and top liner per Figure 17. Note  
that mounting fastener patterns will vary by GC manufacturer. Make sure the  
notch in the inside liner is on the right, in order to accommodate the geometry  
of the Burner as it sits in the shroud and mounting plate.  
Figure 18 Dimensions of GC Liner Cut-Outs  
Align the Burner mounting plate with the mounting screw holes on the GC.  
Clear the hole into the GC oven of interfering insulation, and then secure the  
mounting plate onto the oven with the screws provided. Attach the Burner's  
heated base connector to the GC's temperature control circuit. Consult your  
GC's operation or service manual to confirm proper connection of the 100 ohm  
RTD sensor and the cartridge heaters.  
Occasionally geometric design changes will occur within or between GC models. If the  
mounting plate provided does not match up with the top of the GC, contact Agilent for  
additional information.  
NOTE  
Position the Dual Plasma Burner (column end down) into the tapered fitting of  
the heated base, with the lower hydrogen line and pin aligned with the slot in  
the heated base. The Burner should be secure when properly positioned.  
Operation and Maintenance Manual  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Step 8: Install the Dual Plasma Controller  
Position the Controller such that the gas lines from the Burner can be easily  
attached to the back of the Controller. Connect the Controller to both a  
hydrogen source and an oxidant source, per Step 1. Connect the gas supplies  
to the 1/8" bulkhead unions marked "Oxidizer Inlet" and "Hydrogen Inlet."  
Clean copper tubing (1/8" OD) is recommended.  
Connect the two 1/16" gas delivery lines (provided) from the Burner to the  
appropriate fitting on the rear of the Controller marked Oxidizer or Hydrogen  
in the Outlets (to Burner) area. Use a 1/4" open-end wrench to tighten the nut  
onto the bulkhead unions, using a 9/16" wrench to back-up the union.  
Connect the yellow thermocouple plug from the Burner to the thermocouple  
jack at the rear of the Controller. The thermocouple connector will only fit one  
way, with the iron terminal (the one with the two grooves) down. Connect the  
Burner heater line to the two pin locking connector on the back of the  
Controller.  
The Controller can be used for either 100/115 or 230 VAC. The correct voltage  
for your Controller is set at the factory; however, this should be verified upon  
installation. The voltage selection is shown through a small window on the  
power entry module (see Figure 19). If the voltage setting is incorrect, check to  
make sure that the proper fuses are installed. Contact Agilent for a  
replacement fuse set.  
Do not change the voltage selector without changing the main power fuse. Operating the  
Controller box without the proper fuse may damage the electronics.  
WARNING  
Figure 19 Dual Plasma Controller Rear Panel  
60  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Step 9: Install Column Connections  
The Burner operates under reduced pressure and there will be a slight vacuum  
on the end of the column. If a higher outlet pressure for the column outlet is  
desired, fused silica capillary restrictors may be attached to the end of the  
analytical column (both capillary and packed) prior to making the Detector  
connection.  
Capillary Columns  
Place the column nut over the end of the capillary column. Place the  
appropriate fused silica adapter ferrule onto column. Remove a few  
centimeters from the end to remove any particles that may have entered the  
column. Insert the GC column into the Burner by 108-109 mm from the upper  
end of the nut (114-115 mm if measured from the flat bottom of the nut). Do  
not force the column. The pathway is narrow and may take several tries to seat  
it correctly. Tighten the column nut finger tight, or until sealed using a 7/16"  
open-end wrench to back-up the hydrogen inlet fitting to prevent it from  
slipping.  
108-109 mm  
114-115 mm  
Figure 20 Measuring Column Insertion  
Packed Columns and Columns with an Outside Diameter > 0.8 mm  
Connect a short piece (0.5 m or less) of deactivated fused silica tube, for  
example 0.32 mm internal diameter, to the Detector end of the column. Follow  
the procedure for capillary column connection as above.  
Operation and Maintenance Manual  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Step 10: Install the Transfer Line  
Connect the black transfer line (extending from the side of the Detector) to the  
top connector on the Burner and tighten with a 3/8" open-end and 7/16"  
open-end wrench (backing up the union on top of the Burner to prevent its  
position form slipping).  
62  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors  
Operation and Maintenance Manual  
4
Front Panel Controls and Initial Startup  
63  
Agilent Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Detector Front Panel Controls  
As illustrated in Figure 21, the front panel is divided into three sections: signal  
control, display output control, and power control. Each section is described  
below. All the front panel LEDs are red in the ON mode and darkened in the  
OFF mode except for POWER. This LED toggles green for ON and red for  
STANDBY. In STANDBY mode, the front panel display and high voltage to the  
photomultiplier tube (PMT) are turned off. The red LED in the STANDBY mode  
serves as a reminder that power is still supplied to the instrument.  
Figure 21 Front Panel Controls  
Power Controls  
ON: Turns on front panel displays and high  
voltage to the PMT.  
STANDBY: Darkens panel and turns off high  
voltage to the PMT. Does not shut off main  
power.  
Turns pump ON (red) and OFF (dark).  
Functions in both ON and STANDBY modes.  
64  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Turns ozone generator ON or OFF.  
Functions in both ON and STANDBY modes if the  
pump is producing sufficient vacuum.  
Display Output Controls  
Controls the range of the front panel display  
between high sensitivity (0.1-200mV) and low  
sensitivity (0-2V). Does not affect recorder  
output signal.  
Displays the output signal of the SCD in  
millivolts. (Note: Changing the output range of  
the SCD by adjusting the recorder switch on the  
back panel does not change the range of the  
display.)  
Displays the pressure of the chemiluminescence  
reaction chamber in Torr (mm Hg)  
Signal Controls  
The SCD and NCD use analog amplifier circuitry for the measurement of the  
current produced by the photomultiplier tube. Two Front Panel Controls are  
used to adjust the output signal of the Detector.  
Changes the gain of the amplifier by a factor of  
100. For high sensitivity measurement of sulfur  
or nitrogen compounds (measurement of low  
levels of sulfur or nitrogen compounds), the  
attenuation should be operated in the 1 position.  
For samples containing high levels (ppm to  
percent) of sulfur or nitrogen compounds, use  
the 100 position.  
Operation and Maintenance Manual  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
The baseline signal from the SCD and NCD can be  
adjusted from 0 to 1% of the full scale recorder  
output using the offset control. (e.g., +10 mV to  
-10 mV for a 1 V full scale setting). This control  
can be used to offset the background signal.  
Dual Plasma Controller Controls  
The Agilent Dual Plasma Controller provides easy access to basic settings.  
Figure 22 Dual Plasma Controller Front Panel  
Oxidizer and  
Hydrogen Control  
Knobs  
Allow you to adjust the oxidizer and hydrogen flow  
rate.  
Temperature Control  
Knob  
Allows you to adjust the Burner temperature.  
Selector Control Knob Changes the display to show the current value or  
set point for each setting (temperature in °C,  
pressure in Torr, oxidizer in sccm, or hydrogen in  
sccm).  
66  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Controller Status  
Lights  
Illuminate to reflect current status. The Power LED  
indicates power is on; the Heater LED indicates  
temperature set point has been obtained; the Valves  
LED indicates the solenoid valves open; and the  
Fault LED indicates a drop in pressure or excessive  
details).  
Initial Startup  
Vacuum Test  
With the 1/8" Valco cap on the black PFA transfer line still in place and the air  
supply to the ozone generator off, set the internal air regulator to 0 psi (fully  
counterclockwise), and power on the SCD or NCD by plugging the power cord  
into the house power supply. The SCD and NCD will power up in STANDBY  
mode, with the pump and the ozone generator OFF. For this test, the desired  
mode is power ON, vacuum pump ON, ozone generator OFF, display range in  
the low sensitivity mode and PRESSURE signal displayed. Press the following  
controls in the order shown in order to start the vacuum pump and to view the  
reaction cell pressure.  
Press until green ON LED is lit  
Press until red LED is lit  
Press to display reaction cell pressure in torr  
After the pressure has stabilized ( 5 minutes), record the pressure below.  
~
Typical pressure in the reaction cell should be 1-2 torr with the oil-sealed  
pump, or 5-8 torr for the oil-free pump.  
Operation and Maintenance Manual  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Reaction cell pressure, with air to ozone  
generator OFF (0 psi setting on internal  
regulator) and transfer line capped: ___________________ torr  
Tighten connections if necessary and check to make sure pressure stabilizes in  
the expected region. If proper pressure is not obtained contact Agilent for  
assistance. If the reaction cell pressure is within the expected range, record  
the value, reset the internal air regulator to 3-6 psi, turn the pump OFF and  
proceed with the recorder test.  
Recorder Test  
The standard 355 SCD and 255 NCD recorder output configuration is 1 volt full  
scale. In addition, an output range can be set with the recorder output  
selection switch located on the back panel (Figure 11). Use a small  
screwdriver to adjust the switch to the proper position for your integrator or  
data system.  
To check that the recorder cable has been properly connected to the integrator  
or data system, set the integrator to a high sensitivity setting (e.g. attenuation  
of x 1) and plot the background signal. Use the Output Offset to decrease the  
Detector baseline. If the recorder cable is connected correctly, the baseline will  
shift in response to changes in the Output Offset. If the polarity of the  
connection is incorrect, a negative response will be observed when the  
baseline is increased, and vice versa. Switch the polarity of the wires to  
correct this problem. If no response is observed, re-check the signal  
connections and repeat the test. Contact Agilent if there is no data system  
response after completion of this test.  
68  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Detector Interface Setup  
Initial Checkout  
Careful attention to eliminating leaks in the Detector interface will lead to  
better Detector sensitivity and easier troubleshooting if problems develop.  
1 Check that the gas connections have been made correctly and that they are  
tight.  
2 Plug the 3-prong connector for power on the Controller into a 100-volt,  
50/60 Hz, 115-volt, 50/60-Hz or 230-volt, 50/60 Hz AC outlet.  
3 Connect the oxidant delivery line to the oxidant inlet nut on the Burner.  
Leave the heater cable disconnected and the thermocouple plugged in.  
4 Connect the black PFA transfer line from the Detector to the top of the  
Burner (see Figure 1 and/or Figure 2).  
5 Turn on the Controller, the Detector power, and the vacuum pump.  
6 Let the system pump down to a stable pressure ( 5 minutes).  
~
7 Read the pressure from the LCD display on the Controller. It should be <30  
torr, with higher pressures observed for packed or megabore columns due  
to the higher column flow. Record this pressure for reference.  
Pressure of Burner with column flow but without hydrogen  
and oxidant flow to Burner:_________ torr  
The reaction cell pressure on the Detector should be less than 5 torr  
(10 with oil-free pump).  
8 If pressure readings are not within the above specifications, then there are  
leaks in the system. Leak check all positive pressure points (swage nuts,  
tees, crosses) with SNOOP or a leak Detector. Tighten fittings if necessary  
and recheck the pressure. Confirm that the Burner fittings (oxidant inlet  
nut, tee, and hydrogen nut) are strongly hand-tightened. If necessary,  
tighten the gas delivery lines and transfer line. If pressure readings are still  
high, contact Agilent for service.  
Do not overtighten! Overtightening the fittings on the ceramic tubes can cause the tubes  
to break. Vespel ferrules that are greatly distorted by overtightening cannot be used  
again.  
WARNING  
9 Turn off the Controller and the vacuum pump.  
10 Reconnect and/or turn on the hydrogen and oxidant supplies to the  
Controller.  
Adjust the hydrogen and oxidant inlet pressures to 25 psig (1.7 bar).  
Refer to Table 1 on page 75 for typical operating conditions.  
Operation and Maintenance Manual  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Monitoring Oxidizer and Hydrogen Flow with the Dual Plasma Controller  
Hydrogen will flow to the Burner only if the temperature is above 325 °C and  
the pressure of the Burner is <575 torr.  
Turn ON the vacuum pump at the Detector and the power to the Dual Plasma  
Controller. The power LED should illuminate. When the temperature reaches  
approximately 325 °C, the valve’s LED should illuminate, indicating hydrogen  
and oxidant flow.  
Use the Controller display knobs to select display of oxidizer or hydrogen flow.  
Hydrogen gas is explosive and must be handled carefully. Keep away from sources of  
ignition.  
WARNING  
The Burner is hot. Do not touch the Burner. Let the Burner cool before performing any  
operations involving the Burner.  
WARNING  
70  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors  
Operation and Maintenance Manual  
5
Operation  
71  
Agilent Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Start-Up Procedure  
1 Turn on the GC and set the carrier flow rate.  
2 Turn on the hydrogen and oxygen to the Dual Plasma Controller.  
3 Turn the Detector from STANDBY to ON.  
4 Press the PUMP button until the red LED illuminates. Make sure the  
vacuum pump is running.  
5 Turn on the oxygen to the ozone generator and set the regulator inside the  
front door of the Detector to between 3 and 6 psig.  
6 Allow the system to evacuate for approximately 1 minute before turning on  
the ozone at the front panel. A properly functioning system will show a  
background signal deflection between the ozone on and ozone off.  
7 Turn ON the Dual Plasma Controller using the ON/OFF switch on the rear  
panel of the Controller. As the Burner heats, the SCD or NCD background  
will increase and then slowly decrease. The hydrogen and oxidant flow will  
automatically start when the temperature of the Burner rises above  
approximately 325 °C.  
8 Check the LCD display of Dual Plasma Controller to confirm that all  
pressure, temperature, and signal responses are within the desired  
specifications.  
The Burner is hot. Do not touch the Burner. Let the Burner cool before performing any  
operations involving the Burner.  
WARNING  
72  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Detector Operation  
Detector Stability and Response  
The time required for system stabilization varies depending on the application,  
system cleanliness, presence of active sites and other factors. Useful results  
could be generated within 30 minutes of start-up, especially with a previously  
operated system. A longer stabilization time is likely to be required upon  
changing critical system components, such as the combustion tubes or the GC  
column. In addition, gas flow rates may drift initially as thermal equilibrium is  
reached due to changes in gas viscosity with temperature. Therefore, it is good  
practice to monitor gas flow rates and adjust them accordingly.  
Even though a system may not be fully stabilized, sample injections can be  
made within minutes of instrument start-up. Whether the results are useful  
largely depends on application. Typically, an elevated baseline will initially be  
observed, which will diminish upon successive programmed runs.  
After stabilization has been reached, the system should exhibit good  
short-term and long-term precision. Of course this also depends on the  
application and concentration of components being measured. As an example,  
analysis of thiophene in benzene at the 1 ppm sulfur level yielded 1.4% RSD  
(n=10) over about 2 hours and 3.6% RSD (n=42) over about 96 hours. As  
expected, carbon disulfide at a lower concentration of 90 ppb sulfur yielded  
2.6% and 10.4% RSD, respectively.  
Column Bleed  
Accumulation of column bleed causes silicon dioxide to build up in the Burner.  
This silicon dioxide creates active sites that are detrimental to performance. In  
many cases, the choice of column can be optimized for a particular  
application. Column bleed can be minimized by the use of oxygen traps on the  
carrier gas, low-bleed columns, and lowest possible maximum oven  
temperature.  
Coking  
Contamination from some sample matrices can reduce sensitivity. Crude oils  
containing volatile metal complexes may contaminate ceramic tubes. The  
incomplete combustion of certain hydrocarbon-containing compounds leaves  
behind coke deposits on the tubes. Coke deposits may be removed from the  
Burner by reducing the hydrogen flow rate. The Dual Plasma Burner is much  
less susceptible to coke formation than other designs.  
Hydrogen Poisoning  
Hydrogen poisoning of the ceramic tubes occurs when there is no oxidizer flow  
through the ceramic tubes. The result is extremely reduced, or no response.  
Hydrogen poisoned tubes can not be reconditioned and should be discarded.  
Operation and Maintenance Manual  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
             
Contaminated Gases  
The use of clean gases for the 355 SCD is essential for optimal performance.  
High purity gases (99.999% pure or better) are advised. Sulfur and other  
contaminants from gases may accumulate in the column and bleed out over  
time desensitizing the tubes and causing elevated baselines. The use of sulfur  
traps is highly recommended for all gases.  
Fluctuating Pressures  
Fluctuations in pressure, especially from gas generators, will affect Detector  
response. It is therefore recommended that only bottled gases equipped with a  
dual stage regulator or appropriate steps to ensure stable pressure supplies  
are used.  
74  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Typical Operating Conditions  
The Controller is calibrated at the factory for flow rates to deliver gas in sccm  
units. The following table summarizes the typical operating conditions:  
Table 1 Typical Operating Conditions  
Condition  
SCD  
NCD  
Detector Pressure (Torr)  
4-8 (6-12*)  
4-10  
Dual Plasma Controller  
Pressure (Torr)  
300-400  
100-250  
Burner Temperature (°C)  
Hydrogen Flow Rate (sccm)  
Oxidant Flow Rate (sccm)  
Background Noise (mV)  
* Oil-free pump  
800  
900-950  
4-6  
40-50  
60-65 (air)  
0.3-2.0  
8-12 (oxygen)  
0.3-1.0  
Thermocouple lifetime at 950 °C is diminished.  
The recommended conditions should yield satisfactory results for most  
applications and should be used as a typical starting point for any method  
development. Like any detector, however, there are optimum conditions which  
may very somewhat from the recommended conditions. In optimization of  
conditions for the Dual Plasma Burner and Controller the following guidelines  
should be considered:  
High flow rates of hydrogen and oxidant can release enough heat at high  
temperature to vaporize combustion tubes and cause blockages downstream  
where the materials condense. High flow rates will eventually cause the  
pressure in the Burner to exceed its fault cut-off limit of about 600 Torr. For  
this reason, do not exceed the recommended flow rate by more than about  
25%.  
A higher hydrogen to oxidant ratio may initially show higher response but  
later yield a reduced response because of the accumulation of contaminants,  
such as soot or other active species, that reduce the Detector response.  
Operating the Burner at higher temperatures will place more demand on the  
heater, thermocouple and seal materials, effectively shortening their lifetime.  
In general, when making any parameter change, keep in mind that the system  
may require time to reach equilibrium.  
Operation and Maintenance Manual  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Detection Limits  
The following table lists the detection limits which can be expected for typical  
chromatographic conditions, assuming proper operation of the Detector and  
chromatographic systems.  
Table 2 Expected Detection Limits for Chromatographic Conditions  
Type of Injection  
Volume  
Column  
Detection Limit Per  
Compound as Sulfur  
Liquid, Split 1:10  
1 µL  
1 µL  
Capillary 0.32 mm ID 50 parts per billion  
Capillary 0.32 mm ID 0.5 parts per million  
Capillary 0.32 mm ID 5 parts per billion  
Capillary 0.32 mm ID 0.5 parts per billion  
Capillary 0.32 mm ID 50 parts per billion  
Liquid, Split 1:100  
Liquid, On-column or splitless  
Liquid, Splitless  
1 µL  
10 µL  
1 cm3  
1 cm3  
1 cm3  
10 cm3  
Gas, Split 1:10  
Gas, Direct on-column  
Gas, Direct on-column  
Gas, Direct on-column  
0.53 mm ID  
Packed  
5 parts per billion  
5 parts per billion  
0.5 parts per billion  
Packed  
76  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Instrument Shut-Down  
Daily Shutdown  
1 Toggle off the ozone generator.  
2 Turn off the air regulator (counter-clockwise), located inside Detector door.  
3 Toggle power to "stand-by."  
4 Leave the vacuum pump and Dual Plasma Controller operating at all times.  
Complete Shutdown  
1 Toggle off the ozone generator.  
2 Turn off the air regulator (counter-clockwise), located inside Detector door.  
3 Toggle power to STANDBY.  
4 Turn off power to the Dual Plasma Controller.  
5 After 15 minutes, toggle off the vacuum pump, so the Burner cools and  
moisture is removed from the system.  
6 Turn off gases to unit.  
Operation and Maintenance Manual  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Special Operating Modes  
Using the 255 NCD in Nitrosamine Mode  
By default, the 255 NCD is configured to detect nitrogen. To change from  
nitrogen to nitrosamine mode, first turn off and unplug the Controller. Remove  
the cover, find jumper P6 and the positions labelled High Setpoint and Low  
Setpoint located on the printed circuit board near the left front of the  
Controller. Move the jumper position from the High Setpoint to the Low  
Setpoint position. This changes the temperature control range to 350-500 °C to  
be used for nitrosamine analysis. Refer to see Figure 9 on page 37 for a  
schematic drawing of the 255 NCD in nitrosamine mode.  
Hydrogen is not used in the nitrosamine mode. Turn off and disconnect the  
hydrogen inlet gas. Using the supplied tee, couple the oxidant outlet with the  
two Burner inlets. Use of oxygen is recommended at 5-10 mL/min because the  
Burner can easily be cleaned by raising the Burner temperature to about  
900-1000 °C in flowing oxygen. Alternatively, helium or argon can be used,  
however, these gases will not permit in situ cleaning at elevated temperature.  
The presence of catalyst in the quartz combustion tube will generally yield the  
highest sensitivity for nitrosamines. However, for some sample types that  
contain potentially interfering nitrogen compounds, such as nicotine, it may be  
desirable to remove the catalyst to obtain better selectivity. This is achieved by  
removing the quartz combustion tube and using a straight 1/16" rod or tube to  
push the catalyst out of the tube (save the catalyst for reuse or recovery as it  
contains 90+% platinum). Replace the combustion tube and optimize the  
burner temperature as desired (in general it is desirable to increase the  
pyrolysis temperature by 50-100 °C when the catalyst is not used). Refer to the  
“Tube Replacement for the NCD” on page 102 for additional instruments with  
regard to tube removal and replacement.  
Using the SCD in High Sensitivity Mode for Nonhydrocarbon Gaseous Samples  
There are circumstances in which it may be desirable to operate the SCD in a  
non-typical manner. These could include the analysis of very low levels of  
sulfur species in a nonhydrocarbon gas matrix, such as helium, carbon  
dioxide, or even hydrogen. It should be possible to measure low ppb or high  
ppt levels of sulfur species.  
In the case of a hydrogen matrix, the sample matrix itself can suffice as the  
lower source of hydrogen for Dual Plasma operation in a non-chromatographic  
mode. For instance, a flow rate of nominally 20 SCCM of the sample, a  
hydrogen calibration gas and a “clean” hydrogen source can be alternately  
introduced to the Burner. The sample and calibration gas would be introduced  
into the normal column connection. The side port of the splitter fitting would  
be plugged and the clean hydrogen source would be plumbed to the lower  
hydrogen inlet port. The upper hydrogen and air flow rates would be adjusted  
to nominally 30 and 65 SCCM, respectively.  
In the case of nonhydrocarbon gaseous samples, the potential for coking is not  
a concern, so a Dual Plasma is not necessary. Some improvement in sensitivity  
78  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
can be achieved by eliminating the lower plasma. This is accomplished by  
plugging the side port of the splitter fitting and teeing the air into the lower  
hydrogen line, much like the configuration used for Nitrosamine analysis as  
shown in Figure 9. A slightly lower air flow rate, such as 40 SCCM and a  
hydrogen flow rate of around 60 SCCM is used. Sensitivity of around 0.1 pg  
S/sec or less should be readily achievable. It is not possible to cover all  
potential applications, so some optimization may be desirable.  
Operation and Maintenance Manual  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
80  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Pump Maintenance  
To maintain optimum performance of the Agilent 355 SCD and 255 NCD,  
routine replacement of the chemical trap (for ozone destruction), oil  
coalescing filter and oil (Edwards oil-sealed pump only) is necessary. Refer to  
Table 3 for the expected life span of each replacement part or material.  
It is beneficial to keep a maintenance log that tracks when maintenance is  
performed and any instrument or operational changes that might impact  
performance. Also keep track of detector flow rates (oxidizer and hydrogen),  
pressures (burner controller and reaction cell in detector), and background  
signal (the difference between ozone “on” and ozone “off”).  
Table 3 Operating Life of Components for Edwards RV5 Vacuum Pump  
Component/Material  
Chemical Trap (RV5)  
Oil Coalescing Filter (RV5)  
Pump oil†  
Operating life*  
~ 3 month  
~ 3 months  
~ 3 months  
* The operating life is based on the total time logged during operation of the Detector with the Burner and the  
ozone generator ON.  
† Pump oil can be purchased from a supplier or directly from Agilent: SAE 10W-30, Multiviscosity Synthetic  
Motor Oil such as, MOBIL 1 or AMSOIL.  
82  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Cleaning the Detector  
You can clean the external housing of the Detector with a damp cloth using  
water or non-abrasive cleaners. Turn off power to the Detector and disconnect  
it from main power prior to cleaning. Do not spray liquids directly on the  
Detector. Wipe dry with a clean, soft cloth.  
No cleaning agents which could cause a hazard as a result of reaction with the  
Burner unit are to be used in cleaning the instrument.  
Contact Agilent to address concerns about the compatibility of specific  
cleaning agents with the Burner unit.  
Operation and Maintenance Manual  
83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Changing the Oil Mist Filter (RV5)  
The oil mist filter on the RV5 pump has two components: the charcoal odor  
filter and the oil coalescing filter element. To replace the filters, disassemble  
the oil mist filter assembly with the 4 mm long-handled allen wrench  
(provided). The smaller charcoal odor filter sits on top of the larger oil  
coalescing filter element. It is recommended to replace the oil coalescing filter  
element after 90 days of continuous use, however replacement of the charcoal  
odor filter is optional. After replacing the filter, re-assemble the filter  
assembly and attach it to the pump flange.  
84  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Reaction Cell Cleaning  
Over time, the reaction cell and UV pass filter (SCD) or IR pass filter (NCD)  
will develop a build-up of material which should be removed for optimum  
sensitivity. The cleaning schedule depends upon Detector use and the nature  
of the analyses; however, it is recommended the cell should be cleaned  
annually. The cleaning procedure requires removal of the photomultiplier  
tube (PMT) from the Detector, and special precautions are required to prevent  
damage to the PMT.  
Exposure of a photomultiplier tube to bright light can result in damage to the  
PMT, even when the high voltage is off. To avoid potential damage, minimize  
light exposure. The black PMT cover included in the Detector accessories  
package is recommended for this procedure. Carefully read all the instructions  
below before attempting to remove the PMT. If you have any questions  
regarding this procedure please contact Agilent.  
To clean the cell and pass filter:  
1 Disconnect the main power cord (if the vacuum pump is operated from an  
independent power outlet, unplug the power to the pump as well).  
Any operation requiring access to the inside of the equipment, could result in injury. To  
avoid potentially dangerous shock, disconnect from power supply before opening the  
equipment.  
WARNING  
2 Remove the right side panel from the Detector and disconnect the high  
voltage and coaxial signal cables from the rear of the PMT housing (See  
3 Turn off all room lights and minimize outside light sources.  
4 Unscrew the PMT socket connector assembly. Carefully pull the socket  
assembly and the PMT out toward the rear of the instrument at a slight  
angle to remove the PMT/socket assembly from the housing.  
5 Immediately place the black PMT cap over the PMT, being careful to avoid  
placing fingerprints on the PMT window. Do not remove the PMT from the  
socket assembly. After the PMT cap is in place, carefully place the PMT and  
socket assembly on a soft surface inside a drawer or other dark location to  
minimize exposure of the PMT to light. The room lights may be turned on at  
this point.  
Operation and Maintenance Manual  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
PMT SOCKET  
PMT HOUSING  
PMT  
FILTER  
HIGH  
VOLTAGE  
SIGNAL  
to power supply  
to amplifier  
PMT SOCKET  
REACTION  
CELL  
Figure 23 Reaction Cell, PMT Housing and PMT Socket  
PRESSURE  
SENSOR  
O-RINGS  
REDUCING  
UNION  
TRANSFER  
LINE  
FROM OZONE  
GENERATOR  
Figure 24 Reaction Cell  
6 Use a 7/64" Allen wrench to remove the three mounting screws from the  
reaction cell. Slowly pull the reaction cell back from the PMT housing. The  
optical filter is located between the reaction cell and the PMT housing.  
Remove the optical filter from the housing by tipping up the back of the  
Detector, if necessary, and allow the filter to fall onto a soft cloth.  
7 Inspect the o-rings and replace them if they show any wear.  
86  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
8 Clean any deposits on the optical filter using a soft cloth or Kimwipe  
dampened with methanol or deionized (DI) water. Do not leave fingerprints  
or fibers on the cleaned filter. Deposits inside the reaction cell can be  
cleaned in the same manner, however, care must be taken to avoid bending  
the ozone inlet tube that extends into the cell.  
9 Reseat the pass filter into the PMT housing. Confirm that the two o-rings  
around the reaction cell are properly seated in the O-ring grooves (see  
Figure 24). Carefully re-align the reaction cell to the PMT housing, making  
sure the o-rings remain properly seated, and secure the cell to the housing  
with the three screws. (The two o-rings must be correctly seated in order to  
obtain a vacuum-tight and light-tight seal.)  
10 To re-install the PMT, minimize all light sources. Remove the PMT cap,  
being careful not to touch the PMT window. Insert the PMT and the socket  
assembly into the PMT housing and screw in the socket assembly until it is  
seated tightly against the housing. The room lights may now be turned on.  
11 Reconnect the PMT high voltage cable to the MHV connector (longer  
connector) and the PMT signal cable to the BNC connector (shorter  
connector) at the back of the socket assembly (see Figure 23).  
The high voltage and signal cables must be attached to the proper connectors on the pmt  
socket; damage to the pmt will occur if the cables are not properly connected.  
WARNING  
12 While the side panel is still removed, reconnect the AC power cord to the  
Detector and switch the front panel power to ON. To ensure that the o-rings  
have been properly sealed, monitor the Detector baseline at "attenuation 1"  
for several minutes (with the room light ON), and after noting the baseline  
signal, turn the room lights OFF. If the baseline signal significantly  
decreases, the o-rings are incorrectly positioned and the PMT must be  
removed and the cell re-installed.  
13 If no change in the baseline is observed, check the system for vacuum leaks  
by turning on the vacuum pump and monitoring the pressure as described  
in Section 7. If the pressure readings are comparable to start-up values, the  
o-rings are properly positioned. If a significantly higher pressure is  
observed, the PMT must be removed and the cell re-installed.  
Operation and Maintenance Manual  
87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Flow Sensor Calibration  
The hydrogen and oxidant flow sensors installed in the Dual Plasma Controller  
have very good repeatability, but significant non-linearity. They have each  
been factory calibrated at the midrange flow rate (50 SCCM) using an NIST  
traceable flow meter. Over the typical operating range for the SCD, the flow  
sensors should produce accuracy of better than 10% of reading. However,  
due to sensor non-linearity at low or high flow extremes, error greater than  
this could be observed.  
If greater accuracy is desired at a particular flow rate range, it is possible to  
re-calibrate the flow sensors. A reference flow meter, appropriate gases, and a  
trim-pot adjustment tool are required for this procedure. With the controller  
on and no gases flowing, the zero calibration points are set with RP6 and RP4  
for the hydrogen and oxidant flow streams, respectively. Connect a reference  
flow meter to the gas outlet(s) on the back of the controller. Connect  
pressurized gas line(s) to the inlets on the back of the controller. Connect  
jumper (JP1) to bypass the pressure fault circuitry. Allow the burner  
temperature to exceed 325 °C in order for the hydrogen and oxidant valves to  
be open. The spans of the sensors are then adjusted for the particular flow  
stream to the desired flow rate as measured with the reference flow meter. The  
span adjustments are set with RP3 and RP2, for hydrogen and oxidant  
respectively, so that the display on the controller matches the value obtained  
with the reference flow meter. Upon completion, remove the reference flow  
meter and return the bypass jumper (JP1) to its original position.  
88  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Detector Sensitivity  
Detector sensitivity is an indicator reflecting the performance characteristics  
of a given system, and is a useful tool to determine when Detector  
maintenance is warranted. Sensitivity is typically reported as a minimum  
detection limit (MDL) as calculated from the following formula:  
0.66×(Am'tN / S)×(PktoPkNoise)  
MDL =  
(Wd  
)×(Signal)  
1/2ht  
Where Am't N/S (amount of nitrogen or sulfur) is the mass of nitrogen or  
sulfur in picograms that reaches the Detector, PktoPkNoise (peak-to-peak  
noise) is the measure of the noise (e.g. in mV), Signal is the height of the peak  
in the same units, and Wd1/2ht is the width of the peak at half height in  
seconds. The constant 0.66 is used in the calculation assuming the MDL S/N =  
3.29.  
Before it is released from the factory, each 355 SCD must pass an MDL level of  
<0.5 pg Sulfur/second and each 255 NCD must pass an MDL level of <3.0 pg  
Nitrogen/second. The response from individual detection systems may vary by  
a factor of 2 or 3; however, it is typical for Detectors to perform in the 0.1-0.3  
pg S/second range for the 355 SCD and in the 1-2 pg N/second for the 255 NCD  
when tested at Agilent.  
Operation and Maintenance Manual  
89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Assembling the Dual Plasma Burner for Component Replacement with the  
SCD  
The following procedure can be used to assemble the Dual Plasma Burner for  
use with the SCD or for replacement of Burner components, such as ceramic  
tubes. Refer to Figure 1 on page 26 for proper part nomenclature.  
1 Slide the 0.066" internal diameter (I.D.) double taper ferrule onto the lower  
Burner tube. The tube should extend approximately 2 mm past the end of  
the ferrule.  
2 Insert the lower Burner tube and double taper ferrule into the Burner inlet  
fitting.  
3 Slide the 1/4" Burner adapter over the top of the lower Burner tube all the  
way down to the Burner inlet fitting and screw it onto the fitting  
finger-tight.  
Figure 25 Ferrule Placement on Lower Burner Tube  
4 Slide a 1/4" Swagelok nut over the Burner adapter and then slide a 1/4"  
ferrule over the Burner adapter and position it into the 1/4" nut. Note: If a  
graphite ferrule is used, a small amount of shavings may be created and  
some graphite will be left on the tube surface; this is normal. Avoid allowing  
any shavings to fall inside a tube.  
5 Center the lower Burner tube so that it will slide into the tapered union.  
Insert the lower end of the tapered union fitting into the 1/4" Swagelok nut  
and screw it on finger-tight. If necessary, the brazed H line can be gently  
2
bent out of the way, however, be careful not to stress the brazed (welded)  
connection.  
6 Insert the large ceramic tube into the quartz heater assembly. Position a  
1/4" ferrule (flat end butted up against the top of the swivel nut) onto the  
large ceramic tube. With the ferrule positioned against the swivel nut,  
approximately 0.5 cm of the large ceramic tube should extend outside of the  
nut. Insert the lower Burner tube into the center of the large ceramic tube  
and finger tighten the heater swivel nut onto the tapered union fitting.  
90  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Figure 26 Proper Ferrule Orientation to the Large Ceramic Tube  
Figure 27 Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly  
7 Position the upper ceramic tube into the long axis of the splitter fitting so  
that about 4 mm of it extends past the top of the fitting. Slide the 0.054" ID  
double tapered ferrule over the upper ceramic tube. Gently holding these  
parts so that neither the ferrule nor the upper ceramic tube slips out of  
position, finger-tighten the union fitting onto the splitter fitting.  
Operation and Maintenance Manual  
91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Figure 28 Orientation of the Double Taper Ferrule  
Figure 29 Positioning the Upper Tube in the Union Fitting  
8 Approximately 1.5 cm of the large ceramic tube should extend above the top  
of the quartz heater assembly. Slide a 1/4" Swagelok nut over the large  
ceramic tube and then also slide a 1/4" ferrule over the tube (flat side on  
back of the nut).  
9 Holding the splitter fitting, gently insert the upper ceramic tube into the  
large ceramic tube coaxially, to avoid placing stress on the fragile upper  
ceramic tube. Lower the splitter fitting into place to engage the threads of  
the 1/4" Swagelok nut. Tighten finger-tight.  
10 To begin the final alignment and tightening, use a 7/16" wrench and 5/16"  
wrench to tighten the 1/4" Burner adapter one-quarter turn past  
finger-tight.  
11 Using a 5/8" wrench on the heater swivel nut and a 1/2" wrench on one of  
the flats of the tapered union fitting, tighten the heater swivel nut  
one-quarter turn past finger-tight. Using a 5/16" wrench on the 1/4" Burner  
92  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
adapter, rotate this fitting so that the brazed H line is aligned 180 °  
2
(opposite) from the oxidizer Inlet port.  
Figure 30 Tightening the Heater Swivel Nut  
12 Making sure that the Burner inlet fitting does not loosen, use a 1/2" wrench  
on a flat of the tapered union fitting and 9/16" wrench on the 1/4" Swagelok  
nut of the Burner adapter to tighten the tapered union fitting 1/4" past  
finger-tight.  
13 Rotate the quartz heater assembly so that the thermocouple and heater  
leads are in the same plane and pointed in the same direction as the peg on  
the Burner inlet fitting. Turn the splitter fitting so that H2 inlet port is also  
aligned with the peg on the Burner inlet fitting.  
Figure 31 Proper Alignment of the Burner  
14 Tighten the nut on the splitter fitting one-quarter turn past finger-tight  
using a 9/16" wrench on the 1/4" nut and a 7/16" wrench on the flats of the  
splitter fitting.  
Operation and Maintenance Manual  
93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
15 Carefully bend the H line into position so that the 1/16" Valco nut and  
2
ferrule can be screwed into the side port of the splitter fitting. Tighten the  
connection of the H line to the splitter fitting using a 3/8" wrench on the  
2
vertical flat of the splitter fitting and 1/4" wrench on the Valco nut.  
16 Make sure that no other connections have loosened or moved out of  
alignment, if so, reposition or retighten the fittings as needed.  
17 The assembled Burner is now ready for re-installation on the GC.  
94  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Assembling the Dual Plasma Burner for Component Replacement with the  
NCD  
The following procedure can be used to assemble the Dual Plasma Burner for  
use with the NCD or for replacement of Burner components, such as ceramic  
tubes. Refer to Figure 2 on page 27 for proper part nomenclature. Note that the  
lower section of the NCD Burner is identical to the lower section of the SCD  
Burner.  
While the use of 1/4" ferrules is optional with the SCD, the quartz tube used in the NCD  
NOTE  
Burner is more fragile and the use of 1/4" graphite ferrules is highly recommended.  
1 Slide the 0.066" internal diameter (I.D.) double taper ferrule onto the lower  
Burner tube. The tube should extend approximately 2 mm past the end of  
the ferrule.  
2 Insert the lower Burner tube and double taper ferrule into the Burner inlet  
fitting.  
3 Slide the 1/4" Burner adapter over the top of the lower Burner tube all the  
way down to the Burner inlet fitting and screw it onto the fitting  
finger-tight.  
Figure 32 Ferrule Placement on Lower Burner Tube  
4 Slide a 1/4" Swagelok nut over the Burner adapter and then slide a 1/4"  
ferrule over the Burner adapter and position it into the 1/4" nut.  
5 Center the lower Burner tube so that it will slide into the tapered union.  
Insert the lower end of the tapered union fitting into the 1/4" Swagelok nut  
and screw it on finger-tight. If necessary, the brazed H line can be gently  
2
bent out of the way. Be careful not to stress the brazed (welded) connection.  
6 Find the quartz combustion tube, 1/4" ferrule and quartz heater assembly.  
7 Insert the quartz tube into the quartz heater assembly. Position a 1/4"  
ferrule (flat end butted up against the top of the swivel nut) onto the quartz  
combustion tube. With the ferrule positioned against the swivel nut,  
approximately 0.5 cm of the quartz tube should extend outside the nut.  
Operation and Maintenance Manual  
95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Figure 33 Proper Ferrule Orientation to the Large Quartz Tube  
Figure 34 The Quartz Tube Properly Inserted into the Quartz Heater Assembly  
8 Insert the lower burner tube into the center of the quartz tube and finger  
tighten the heater swivel nut onto the tapered union fitting, then tighten an  
additional 1/4 turn making sure not to break the quartz tube.  
9 To begin the final tightening, use a 7/16" wrench and 5/16" wrench to  
tighten the 1/4" Burner adapter one-quarter turn past finger-tight.  
96  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
10 Using a 5/8" wrench on the heater swivel nut and a 1/2" wrench on one of  
the flats of the tapered union fitting, tighten the heater swivel nut  
one-quarter turn past finger-tight. Using a 5/16" wrench on the ¼" Burner  
adapter, rotate this fitting so that the brazed H line is aligned 180 °  
2
(opposite) from the oxidizer Inlet port.  
11 Making sure that the Burner inlet fitting does not loosen, use a 1/2" wrench  
on a flat of the tapered union fitting and 9/16" wrench on the 1/4" Swagelok  
nut of the Burner adapter to tighten the tapered union fitting 1/4" past  
finger-tight.  
12 Rotate the quartz heater assembly so that the thermocouple and heater  
leads are in the same plane and pointed in the same direction as the peg on  
the Burner inlet fitting. Turn the splitter fitting so that H inlet port is also  
2
aligned with the peg on the Burner inlet fitting.  
Figure 35 Burner Assembly Detail  
Figure 36 Burner Assembly Alignment  
13 The assembled Burner is now ready for re-installation on the GC.  
Operation and Maintenance Manual  
97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Tube Replacement for the SCD  
The Burner has two combustion tubes that require replacement: the upper  
ceramic tube and the Large Ceramic Tube. Generally the tubes should be  
replaced only if sensitivity decreases. The “Troubleshooting” chapter provides  
additional information to assist in determining whether tube replacement may  
be necessary. Refer to Figure 1 on page 26 for proper part nomenclature.  
Follow the instructions below for tube replacement.  
1 Turn off power to the GC and the Controller and let the system cool down  
under vacuum.  
2 Turn off power to the vacuum pump.  
3 Lift the Burner out of the shroud. It is recommended to remove the coil,  
noting the position of the ferrule. In some instances, if the coil can be  
uncoiled, it may be convenient to leave it attached to the Burner.  
4 Disconnect the hydrogen and oxidant lines.  
5 Disconnect the power connector that leads to the GC, if necessary.  
6 Tilt the Burner at an angle, so that when loosening the union fitting the  
upper ceramic tube does not slide down into the large ceramic tube.  
7 Loosen and disconnect the union fitting, and pull the splitter fitting and  
upper ceramic tube out of the Burner.  
8 Slide the upper ceramic tube out of the splitter fitting.  
9 Slide the upper ceramic tube into the splitter fitting, so that approximately  
4 mm of the tube extends beyond the top of the fitting. Then, slide the  
double taper ferrule onto the tube (see Figure 37 and Figure 38 for proper  
positioning). Gently holding these parts so that neither the ferrule nor the  
upper ceramic tube slip out of position, finger-tighten the union fitting onto  
the splitter fitting.  
Figure 37 Orientation of the Double Taper Ferrule  
98  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Figure 38 Positioning the Upper Tube in the Union Fitting  
10 Holding the splitter fitting, gently insert the upper ceramic tube into the  
large ceramic tube coaxially, to avoid placing stress on the fragile upper  
ceramic tube. Lower the splitter fitting into place to engage the threads of  
the 1/4" Swagelok nut. Tighten finger-tight.  
11 If you do not need to replace the Large Ceramic Tube, proceed to step 19.  
12 Remove the Tapered Union Fitting from the bottom of the Burner.  
13 Slide the Large Ceramic Tube out of the Burner and remove it from the  
Quartz Heater Assembly.  
14 Insert the new large ceramic tube into the quartz heater assembly. Position  
a 1/4" ferrule (flat end butted up against the top of the swivel nut) onto the  
large ceramic tube. With the ferrule positioned against the swivel nut,  
approximately 0.5 cm of the large ceramic tube should extend outside of the  
nut. Insert the lower Burner tube into the center of the large ceramic tube  
and finger tighten the heater swivel nut onto the tapered union fitting.  
Figure 39 Proper Ferrule Orientation to the Large Ceramic Tube  
Operation and Maintenance Manual  
99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Figure 40 Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly  
15 Approximately 1.5 cm of the large ceramic tube should extend above the top  
of the quartz heater assembly. Slide a 1/4" Swagelok nut over the large  
ceramic tube and then also slide a 1/4" ferrule over the tube (flat side on  
back of the nut).  
16 To begin the final alignment and tightening, use a 7/16" wrench and 5/16"  
wrench to tighten the 1/4" Burner adapter one-quarter turn past  
finger-tight.  
17 Using a 5/8" wrench on the heater swivel nut and a 1/2" wrench on one of  
the flats of the tapered union fitting, tighten the heater swivel nut  
one-quarter turn past finger-tight. Using a 5/16" wrench on the 1/4" Burner  
adapter, rotate this fitting so that the brazed H line is aligned 180 °  
2
(opposite) from the oxidizer Inlet port.  
Figure 41 Tightening the Heater Swivel Nut  
18 Making sure that the Burner inlet fitting does not loosen, use a 1/2" wrench  
on a flat of the tapered union fitting and 9/16" wrench on the 1/4" Swagelok  
100  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
nut of the Burner adapter to tighten the tapered union fitting one-quarter  
turn past finger-tight.  
19 Rotate the quartz heater assembly so that the thermocouple and heater  
leads are in the same plane and pointed in the same direction as the peg on  
the Burner inlet fitting. Turn the splitter fitting so that H inlet port is also  
2
aligned with the peg on the Burner inlet fitting.  
Figure 42 Proper Alignment of the Burner  
20 Tighten the nut on the splitter fitting one-quarter turn past finger-tight  
using a 9/16" wrench on the 1/4" nut and a 7/16" wrench on the flats of the  
splitter fitting.  
21 Carefully bend the H line into position so that the 1/16" Valco nut and  
2
ferrule can be screwed into the side port of the splitter fitting. Tighten the  
connection of the H line to the splitter fitting using a 3/8" wrench on the  
2
vertical flat of the splitter fitting and 1/4" wrench on the Valco nut.  
22 Make sure that no connections have loosened or moved out of alignment, if  
so, reposition or retighten the fittings as needed.  
23 Replace the Burner in the shroud.  
24 Follow the standard system start-up procedure, including column  
placement.  
Operation and Maintenance Manual  
101  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Tube Replacement for the NCD  
The Burner has one quartz combustion tube that requires replacement.  
Generally the tube should be replaced only if sensitivity decreases. The  
“Troubleshooting” chapter provides additional information to assist in  
determining whether tube replacement may be necessary. Refer to Figure 2 on  
page 46 for proper part nomenclature. Follow the instructions below for tube  
replacement.  
1 Turn off power to the GC and the Controller and let the system cool down  
under vacuum.  
2 Turn off power to the vacuum pump.  
3 Lift the Burner out of the shroud. It is recommended to remove the coil,  
noting the position of the ferrule. In some instances, if the coil can be  
uncoiled, it may be convenient to leave it attached to the Burner.  
4 Disconnect the hydrogen and oxidant lines.  
5 Disconnect the power connector that leads to the GC, if necessary.  
6 Loosen and disconnect the union fitting from the top of the Burner.  
7 Remove the tapered union fitting from the bottom of the Burner.  
8 Slide the large quartz tube out of the Burner and remove it from the quartz  
heater assembly.  
9 Insert the new quartz tube into the quartz heater assembly. Position a 1/4"  
ferrule (flat end butted up against the top of the swivel nut) onto the quartz  
combustion tube. With the ferrule positioned against the swivel nut,  
approximately 0.5 cm of the quartz tube should extend outside the nut.  
Figure 43 Proper Ferrule Orientation to the Large Quartz Tube  
102  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Figure 44 Large Quartz Tube Properly Inserted into the Quartz Heater Assembly  
10 Insert the lower burner tube into the center of the quartz tube and finger  
tighten the heater swivel nut onto the tapered union fitting, then tighten an  
additional 1/4 turn making sure not to break the quartz tube.  
11 To begin the final tightening, use a 7/16" wrench and 5/16" wrench to  
tighten the 1/4" Burner adapter one-quarter turn past finger-tight.  
12 Using a 5/8" wrench on the heater swivel nut and a 1/2" wrench on one of  
the flats of the tapered union fitting, tighten the heater swivel nut  
one-quarter turn past finger-tight. Using a 5/16" wrench on the ¼" Burner  
adapter, rotate this fitting so that the brazed H line is aligned 180º  
2
(opposite) from the oxidizer Inlet port.  
13 Making sure that the Burner inlet fitting does not loosen, use a 1/2" wrench  
on a flat of the tapered union fitting and 9/16" wrench on the 1/4" Swagelok  
nut of the Burner adapter to tighten the tapered union fitting one-quarter  
turn past finger-tight.  
14 Rotate the quartz heater assembly so that the thermocouple and heater  
leads are in the same plane and pointed in the same direction as the peg on  
the Burner inlet fitting. Turn the splitter fitting so that H inlet port is also  
2
aligned with the peg on the Burner inlet fitting.  
Operation and Maintenance Manual  
103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Figure 45 NCD Tube Replacement Detail  
Figure 46 NCD Tube Replacement Detail  
15 The assembled Burner is now ready for re-installation on the GC, including  
column placement.  
104  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors  
Operation and Maintenance Manual  
7
Troubleshooting  
105  
Agilent Technologies  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Solving Detector Problems  
A basic understanding of the Detector helps one systematically diagnose and  
solve Detector problems. Many symptoms may be caused by more than one  
problem and these are the most difficult to troubleshoot. It should be pointed  
out, however, that analysis of sulfur or nitrogen compounds has traditionally  
been very difficult because of the inherent reactivity and instability of the  
compounds themselves. Often, problems blamed on the Detector actually  
originate from either poor chromatographic technique or other system failures  
(most of these problems are injector related).  
Therefore, the first step in troubleshooting is to isolate the problem: in the  
chromatographic system, the Burner assembly, or in the Detector itself (ozone  
generator, vacuum pump, photomultiplier tube or electronics). Diagnosing the  
location of a problem is often facilitated by returning to Chapter 4 and  
attempting to verify initial conditions. The table at the end of this section lists  
many common problems, their most probable causes and corrective action that  
should be taken.  
As a good practice, keep a maintenance log (as advised in Chapter 6) and use it  
to aid troubleshooting. in the maintenance log, keep track of detector flow  
rates (oxidizer and hydrogen), pressures (burner controller and reaction cell  
in detector), and background signal (the difference between ozone “on” and  
ozone “off”).  
For assistance with troubleshooting Detector problems, contact Agilent with  
the serial number of the unit, the conditions used by the instrument, and any  
recent changes that have been made.  
106  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Power Problems  
The first step in the determination of a power related problem is to verify  
power to the pump, Controller, and Detector itself. The inability to establish  
power may be as trivial as a blown fuse. Fuse requirements and positions on  
the Detector are indicated in see Figure 47 and Table 4.  
Repeated fuse failure is an indication of a more serious situation.  
Detector Fuse  
If the Standby LED does not illuminate when the unit is plugged in, and  
multiple fuses fail on the Detector, disconnect the power cord and contact  
Agilent.  
Vacuum Pump Fuse  
Repeated blown fuses of the vacuum pump indicate a pump oil problem. If the  
vacuum pump has not been in operation for some time or if water accumulated  
in the vacuum pump has not been properly removed, the pump may be difficult  
to start, or may require more power than normal, and this may blow the main  
AC fuse. Disconnect the main power cord from the AC outlet and from the back  
panel of the Detector. Replace the pump oil. Connect the pump power cord to  
an AC outlet in the lab to start it. Allow the pump to operate for 10 to  
15 minutes, after which the pump may be plugged back into the rear panel of  
the Detector. If fuses blow after the pump oil has been changed, the pump has  
been damaged. Contact Agilent for further information.  
The dry piston pump may require a short break-in time when new or after  
seals have been changed. It can be operated for a few hours by plugging it into  
a wall outlet. When turned off, it is necessary to allow the pump pressure to  
reach atmospheric conditions before restarting.  
Also note that the dry piston pump is equipped with a thermal fuse;  
connecting the pump to a power supply that exceeds 110 V 10% will result in  
the pump shutting down.  
Dual Plasma Controller Fuses  
If the Controller will not power on, confirm that the power cord is firmly  
seated at both ends. Also make certain that there is power to the outlet at the  
wall, surge protector, or circuit breaker. If the power cord is seated correctly  
and power is available to the Controller, the fuse in the power entry module  
(just above the power switch) on the back of the Controller may have blown.  
Make sure that the actual line voltage and the setting on the Controller match.  
To replace the fuse in the power entry module, follow these steps:  
1 Unplug the power cord from the back of the Controller.  
2 Use a small flat-head screwdriver to pry open the fuse cover on the power  
entry module on the back of the Controller.  
Operation and Maintenance Manual  
107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
3 Remove the old fuses from the holder and replace with new fuses.  
If the Controller still does not power on after replacing the fuse, contact  
Agilent.  
Consult Table 4 for the appropriate replacement fuse.  
Table 4 Fuses for 100 V, 120 V and 230 V Versions of 355 SCD and 255 NCD  
100/120 V  
230 V  
Detector Back Panel  
Main AC Power  
15 A/250 V  
3AG slo-blo  
T10 A/250 V  
5 x 20mm  
Detector  
Electronics Power Supply (F1) 250 mA/250 V  
3AG slo-blo  
T125 mA/250 V  
5 x 20mm  
Pump (F2)  
15 A/250 V  
3AG slo-blo  
T5 A/250 V  
5 x 20mm  
Ozone Generator (F3)  
Photomultiplier Tube Cooler  
1 A/250 V  
3AG slo-blo  
T100 mA/250 V  
5 x 20mm  
1 A/250V  
3AG slo-blo  
T500 mA/250V  
5 x 20mm  
Dual Plasma Controller Back Panel  
Main AC Power (2 fuses) T2 A/250 V  
1 A/250 V  
5 x 20mm  
5 x 20mm  
F1  
ELECTRONICS  
OZONE  
ELECTRONICS  
T125 mA  
OZONE  
F3  
F2  
T100 mA  
PUMP  
PUMP  
T5 A  
Front View (230 V Unit)  
Side View  
Figure 47 Fuse Positions on the Power Supply Board  
108  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Ozone Generation Problems  
Following the verification of power to the instrumental components, the next  
step in tracing a problem is the determination of ozone production. With the  
ozone toggle off and the vacuum pump on, read the signal output on the front  
of the Detector. Toggle the ozone on. A properly operating Detector will  
typically display a difference in background of 0.2 to 2 mV. If no change is  
observed, a problem most likely lies within the ozone generator itself, the high  
voltage transformer supplying power to the ozone generator, or in the transfer  
line system between the ozone generator and the reaction cell.  
Ozone Generator  
Loss of ozone from the ozone generator may arise from generator fault, or from  
a leak in the corona discharge tube. Typically a pungent ozone odor will be  
noticed emanating from the side of the Detector if the ozone generator is  
leaking.  
Any operation requiring access to the inside of the equipment, could result in injury. To  
avoid potentially dangerous shock, disconnect from power supply before opening the  
equipment.  
WARNING  
WARNING  
High voltages supply the ozone generator. Unplug the instrument.  
Removal of the left side panel (as seen from the front of the Detector) and  
removal of the ozone generator cover panel may reveal corrosion caused by  
escaped ozone. If this is the case, replace the ozone generator. Resistance  
across the leads to the ozone generator should be infinite. Measure the  
resistance with an ohmmeter to verify. If there is a resistance, replace the  
ozone generator.  
High Voltage Transformer  
The high voltage transformer produces 6000 - 8000 volts.  
WARNING  
The following test should only be performed by those with high voltage experience using  
high voltage probes.  
Using high voltage probes, directly measure the secondary output of the ozone  
generator. A properly functioning generator will produce approximately  
7500 volts. Output less than 3000 volts mandates replacement of the high  
voltage transformer.  
Operation and Maintenance Manual  
109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Plugged Restrictor Lines  
Plugged restrictor lines are verified by turning off the air/oxygen regulator  
located inside the door of the Detector and observing little or no change in the  
pressure as indicated on the regulator. The plug can be located in either the  
pre-ozone restrictor or the post-ozone restrictor (see Figure 3). A partially  
plugged restrictor line can result in a non-linear signal, or reduced sensitivity  
which will vary with the flow of ozone. Replace the restrictor lines.  
110  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Response Problems  
Low or no response problems are the most difficult to troubleshoot on the  
Agilent SCD and NCD as they may arise from one source or a combination of  
sources. Primarily, response problems are due to combustion problems, and  
Burner integrity should be investigated first.  
It is beyond the scope of this supplement to deal with all possible  
chromatography related problems. The following table lists many of the  
common problems, their most probable causes, and the corrective action that  
should be taken. For further assistance contact Agilent.  
Operation and Maintenance Manual  
111  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Temperature Reading Problems  
Normally, the controller displays the burner temperature set-point. For  
diagnostic purposes, a switch in the controller (JP2) can be positioned to allow  
the display of the actual thermocouple temperature reading. When the actual  
temperature and set-point temperature agree, the heater indicator light on the  
front of the controller is illuminated. Note that there is a slight offset between  
these readings.  
112  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Diagnosing General Problems  
Table 5 Troubleshooting Detector Issues  
Problem  
Possible Cause  
Diagnosis  
Corrective Action  
No Response  
No ozone  
Little or no difference in  
Continue below. See No  
output signal between ozone Ozone in Problem column.  
ON vs. OFF. Typically the  
background signal will be  
0.3-0.8 mV higher with the  
ozone ON.  
No Ozone  
No Ozone  
Blown fuse  
Ozone indicator light  
remains off when ozone  
button is pushed.  
Replace ozone fuse.  
High Voltage Transformer  
and/or ozone generator is  
inoperative.  
No difference in output  
Have ONLY someone with  
signal between ozone ON vs. high voltage experience use  
OFF even though flow a high voltage probe to  
through the ozone generator check the input to the ozone  
is good (about  
25-35 mL/min).  
generator across the two  
wires. It should be  
>6000 volts. If it is not then  
replace high voltage  
transformer. If it is, replace  
ozone generator.  
No Ozone  
Ozone restrictor(s) plugged. Needle on the regulator  
inside front panel does not  
move downward when  
Replace plugged  
restrictor(s).  
regulator is turned fully  
counterclockwise. The  
measured flow is low or  
non-existent.  
No Response  
No Response  
Hydrogen and/or air ran out. Measure flow rates.  
Set correctly.  
Change tubes.  
Broken ceramic tube(s).  
Pressure on Controller  
and/or Detector is too high  
or too low.  
No Response  
Low Response  
Low Response  
Heating element failure.  
Temperature at Controller  
<100 °C.  
Replace heating element.  
Adjust flow rates.  
Improper hydrogen/air flow Measure flow rates.  
rates.  
Detector leaks.  
Note pressures on  
Controller.  
Locate and repair leaks,  
check integrity of ferrules.  
Operation and Maintenance Manual  
113  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Table 5 Troubleshooting Detector Issues (continued)  
Problem  
Possible Cause  
Diagnosis  
Corrective Action  
Low Response  
Contaminated ceramic  
tubes.  
If there does not appear to  
be a leak, then the tubes  
should be inspected.  
Replace tubes.  
Contamination can result  
from column bleed, samples  
which may contain volatile  
metal complexes, and large  
injections of coke forming  
hydrocarbons.  
Wandering Baseline  
Wandering Baseline  
Wandering Baseline  
Poor temperature control.  
Monitor the temperature on Check for a loose connection  
the Controller. It should vary on the heating element or  
by no more than 5 °C.  
thermocouple. Reposition  
the thermocouple in the  
Burner.  
Contamination in one of the Check the difference in the  
Detector gases.  
Change Detector gases after  
output signal between ozone adding in-line traps.  
on and off. The difference  
should be between 0.2-2 mV  
after equilibration.  
Leak in the oxidizer line at  
If using SCD, use a microliter Reseat air line into air inlet  
the oxidizer inlet connection. syringe containing a small  
amount of sulfur compound  
nut, or replace.  
e.g., CS2 to “snoop” for leaks  
while watching mV output.  
Leaks are very evident by a  
large increase in signal  
displayed on the LED.  
Leak at weld of H2 inlet  
fitting.  
Tailing Peaks with  
Non-Equimolar Response  
Severe contamination of  
Detector gases.  
High background signal with Avoid “house” gases,  
respect to ozone off.  
especially air from  
compressors. Clean gases  
should be used with  
appropriate traps.  
Tailing Peaks  
Poor column connection.  
Verify column position at  
inlet and outlet. Look for  
discoloration of column at  
Detector side which  
Reinstall column.  
indicates column in  
combustion zone.  
Tailing Peaks  
Cracked tubes.  
Confirm pressure and  
vacuum ranges. Inspect  
columns and ferrules.  
Replace tubes as needed.  
Replace thermocouple.  
Controller Reads >1000 °C  
Thermocouple open.  
Check electrical resistance  
between thermocouple pins.  
114  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table 5 Troubleshooting Detector Issues (continued)  
Problem Possible Cause  
Diagnosis  
Corrective Action  
Burner Fault Cycles. (off and Cracked tube.  
ON)  
Pressure increases  
>525 torr, Burner faults.  
Cycle repeated as  
temperature and pressure  
reset.  
Replace broken tubes.  
Burner Fault Cycles (off and Leak in Burner.  
ON)  
Pressure increases  
>600 torr, Burner faults.  
Cycle repeated as  
temperature and pressure  
reset.  
Locate and repair leaks,  
check integrity of ferrules.  
Table 6 Troubleshooting Pump Issues  
Problem  
Possible Cause  
Diagnosis  
Corrective Action  
Pump is noisy  
High-pressure dry piston.  
Pressure in the reaction cell Change sleeve/seal.  
is unacceptably high and the  
pump is getting audibly  
noisy.  
Pump Doesn’t Start  
Pump Doesn't Start  
Fuses Blow on Startup  
Water in Pump  
Dry piston seals have  
recently been changed.  
Pump fuses blow  
immediately after changing time of a few minutes after  
the seals and sleeves.  
Dry piston requires break-in  
seals are changed.  
Pump switch off (dry piston Locate pump switch on  
pump must start from  
atmospheric pressure).  
Turn ON switch.  
pump and verify position.  
Inspect oil for integrity.  
Emulsified oil (worn seals  
and head on dry piston  
pump).  
Change pump oil, and plug  
unit into wall to run for 10-15  
minutes.  
Cracked coalescing filter.  
Milky yellow oil in the pump Change coalescing filter and  
window.  
pump oil.  
Reaction Cell Pressure High chemical trap clogged.  
Reaction Cell Pressure High Leak in ozone generator.  
Remove trap from the  
vacuum line and confirm  
expected pressure readings.  
Change chemical trap.  
Air regulator falls rapidly to  
zero when turned off.  
Replace ozone generator.  
Pump Loses Oil  
Gurgle Sound  
Ballast Open.  
Oil level drops.  
Reset ballast  
See Pump specific sections.  
High level of oil in  
Coalescing Filter  
Plugged oil return restrictor No visible movement of oil in Change filter and clear  
the return line. restrictor.  
Operation and Maintenance Manual  
115  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Table 7 Troubleshooting Burner Issues  
Problem  
Possible Cause  
Diagnosis  
Corrective Action  
Fuses Blow on Startup  
Shorted heater element.  
Inspect for exposed wires or Take Burner top off and  
broken insulation around  
wire leads. Measure  
resistance leads, should not  
be 0.  
reseat insulation around  
heater leads.  
Low Sensitivity  
Leaks at ferrules or fittings. For SCD, use a microliter  
syringe containing a small  
Inspect fittings for scoring  
and replace if necessary.  
amount of sulfur compound Replace ferrules if worn.  
e.g., CS2 to “snoop” for leaks Hand tighten all fittings  
while watching mV output.  
Leaks are very evident by a  
large increase in signal  
displayed on the LED.  
mounted onto vespel  
ferrules, and wrench tighten  
all connections with Valco  
ferrules.  
For NCD, a high background  
could be observed from  
atmospheric nitrogen.  
Low Sensitivity  
Low Sensitivity  
Improper gas flows.  
Measure gas flows.  
Reset gas flows.  
Broken ceramic tubes or  
ferrules  
Remove and inspect tubes  
and ferrules for chips and/or  
breaks.  
Replace tubes or ferrules.  
High Pressure  
Low Sensitivity  
Improper column  
positioning.  
Inspect for charred Detector After cutting off charred end  
end of column.  
of column, measure column  
length 108-109 mm from  
ferrule (or 114-115 mm from  
bottom of nut) and re-attach  
column.  
High Background  
116  
Operation and Maintenance Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Index  
Fluidics overview, 42  
Fuses  
C
S
replacing, 107, 108  
Calibration of Flow Sensor, 88  
Ceramic Tube Replacement, 98  
Column Insertion Distance, 61  
Schematic  
for NCD, 36  
for NCD (nitrosamine mode), 37  
for SCD, 35  
I
Sensitivity  
Installation  
D
minimum detection limit, 89  
NCD specifications, 20  
SCD specifications, 20  
Sensor Calibration, 88  
Shut-down procedure, 77  
Signal Output Cables  
installation, 58  
column connections, 61  
Dry piston pump, 52  
Dual Plasma Burner, 59  
Dual Plasma Controller, 60  
Edwards oil-sealed pump, 46  
gases, 42, 43  
initial connections, 46  
overview, 40  
oxidant supply, 57  
power requirements, 41  
required tools, 45  
selecting a location, 41  
signal output cables, 58  
transfer line, 62  
Detection Limits, 76  
Detector  
cleaning, 83  
display output controls, 65  
initial check, 69  
operating conditions, 75  
operation, 73  
power controls, 64  
schematic, NCD, 36  
schematic, NCD (nitrosamine mode), 37  
schematic, SCD, 35  
shut down, 77  
Special Operating Modes  
nitrosamine, 78  
nonhydrocarbon gaseous samples, 78  
Specifications, 20  
Start-up Procedure, 72  
System schematic, 42  
System Stabilization, 73  
System Startup, initial, 67  
signal controls, 65  
start-up procedure, 72  
Dry Piston Pump  
Vacuum Pump, 46  
Welch dry piston pump, 52  
installation, 52  
Dual Plasma Burner  
T
M
assembling for NCD, 95  
assembling for SCD, 90  
cross-section illustration for NCD, 27  
cross-section illustration for SCD, 26  
installing, 59  
Temperature  
set-point vs. thermocouple reading, 112  
Theory of Operation, 23  
Troubleshooting  
Maintenance  
changing oil mist filter, 84  
cleaning the detector, 83  
pump, 82  
burner issues, 116  
detector, 106  
detector issues, 113  
general problems, 113  
ozone, 109  
power and fuses, 107, 108  
pump issues, 115  
overview, 25  
Minimum Detection Limit (MDL), 89  
tube replacement for NCD, 102  
tube replacement for SCD, 98  
Dual Plasma Controller  
controls and settings, 66  
front panel illustration, 66  
installing, 60  
O
Operating Conditions, 75  
Ozone Generator  
overview, 28  
table of problems and causes, 113  
Tube Replacement  
operating conditions, 75  
overview, 24  
troubleshooting, 109  
for NCD, 102  
rear panel illustration, 60  
for SCD, 98  
P
E
V
Pressure Regulator, use of, 43  
Edwards oil-sealed pump  
ballast position, 51  
installation, 46  
Vacuum Pump  
overview, 32  
Q
Quartz Tube Replacement, 102  
W
F
Welch Dry Piston Pump  
installation, 52  
R
FID Adapter  
overview, 33  
Flow Sensor Calibration, 88  
operation notes, 53  
Recorder test, 68  
Operation and Maintenance Manual  
117  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Agilent Technologies  
© Agilent Technologies, Inc.  
Printed in USA, June 2007  
Download from Www.Somanuals.com. All Manuals Search And Download.  

3M Snowshoes 747 User Manual
Acer Laptop 1300 Series User Manual
Advantech Computer Hardware CP 150 User Manual
Alpine Car Stereo System CDA 9853R User Manual
Alpine Car Stereo System TME M770 User Manual
Altinex Switch MT104 100 103 User Manual
Ask Proxima Projector DP5600 User Manual
Asus Computer Hardware V8170 User Manual
Audiovox Automobile Alarm APS25CH User Manual
Avenview Stereo System DVI C5 R User Manual