3M Marine Instruments 48N User Manual

3M Persomal Safety Division  
3MQUESTemp°Heat Stress Monitors  
Robust Construction  
From the  
Market Leader  
o
UESTemp 48N User Manual  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ii  
Contents  
List of Figures  
Download from Www.Somanuals.com. All Manuals Search And Download.  
iii  
Contents  
LIST OF TABLES  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
1
Introduction  
Up and Running overview  
Introduction  
The QUESTempº 48N offers traditional heat stress monitoring without the aggravation  
of maintaining a wet bulb. Through collaboration with Professor Dr. Thomas Bernard,  
from the College of Public Health at the University of South Florida, mathematical  
models were implemented to create a Waterless Wet Bulb calculation through a  
combination of dry bulb temperature, globe temperature, relative humidity, and air  
flow.  
Figure 1-1: QUESTempº 48N in a thermal environment  
QUESTempº 48N Model  
The QUESTempº 48N model was specifically engineered for the Department of  
Defense Ashore and Afloat Operations in which conditions require hand’s free  
monitoring, in-the-field user RH/WBGT sensor alignments, and snapshot measurement  
logging (called “event logging mode”).  
The QT48N measures and calculates the dry bulb, wet bulb, globe, WBGT outdoors,  
and relative humidity with two different types of data logging modes as a time  
history session or an event logging mode.  
You also have the capability to measure stay times in order to manage work/rest  
regimens. Guidance is based on the screening criteria for heat stress as defined in the  
ACGIH TLV Handbook, U.S. Navy PHEL charts, and Flag Conditions for U.S.  
Navy/Marine Corp. Ashore.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Getting started  
2
Getting Started  
Up and Running overview  
Getting Started  
Up and Running overview  
1. Place the QUESTempº 48N in the work area in a safe location approximately  
3.5 feet off the ground.  
2. Turn the unit On. If the battery voltage displayed during the power-on  
sequence is less than or equal to 6.4 volts, replace or recharge the  
batteries.  
3. Be aware that the sensors require 10 minutes to stabilize to a new  
environment.  
4. In the main menu, View will be selected (an indicator arrow denotes the  
selected menu). Press the I/O Enter key and the measurement screen  
will appear.  
5. Press the Run/Stop key to begin datalogging. Use the Arrow keys to  
set the display to the desired items.  
Placing the QUESTempº 48N on the job site  
The QUESTempº 48N should be placed at a height of 3.5 feet (1.1m) for standing  
individuals or 2 feet (.6m) for seated individuals. Tripod mounting is recommended  
to get the unit away from anything that might block radiant heat or airflow. A 1/4"x  
20 threaded bushing on the bottom of the instrument allows mounting to a  
standard photographic tripod. Do not stand close to the unit during sampling.  
Before data logging, allow ten minutes for the sensors readings to stabilize.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
3
Getting Started  
Keypad Operation  
Keypad Operation  
The unit operates using a keypad with 4 keys. The I/O Enter key responds  
when the key is released while all other keys respond when the key is pressed.  
Keys  
Explanation  
I/O Enter key  
The unit turns on with a single key press. The unit turns off  
by holding the key down while a countdown of 3-2-1 occurs  
in the lower right corner of the display. This key is also used  
to select a mode (such as Setup or View) or enter setup  
changes.  
Pressing and releasing the key while viewing temperatures  
causes the display to view the next available sensor bar  
(indicated in the upper right corner of the display).  
Up Arrow key  
Changes items appearing in the display. Scrolls up.  
Changes items appearing in the display. Scrolls down.  
Down Arrow  
key  
Run/Stop key  
From the menu or view modes, pressing this key starts or  
stops the run mode. Pressing this key will exit the setup,  
print or reset modes.  
Escaping or  
moving back  
one screen  
If you are in the setup, print, reset, or sensor alignment  
screens, you can press Run/Stop key to escape or move  
back one screen.  
Table 1-1: Keypad explained  
I/O Enter key  
Run/Stop key  
Up Arrow key  
Down Arrow key  
Figure 1-2: Keypad keys identified  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
4
Getting Started  
Turning on, logging/capture an event, and turning off  
Turning on, logging/capture an event, and turning off  
To quickly get you started with the QUESTempº 48N, the following section explains  
turning on the instrument, data logging/event logging, and stopping.  
1. Press the I/O Enter key to turn on.  
The first two start-up screens detail the name, type of instrument, and then  
a revision and battery power screen appears. (Ensure the battery level  
does not fall below 6.4V or you will want to replace the battery. See  
“Power options”, page 19 for more details.)  
NOTE: if you have data saved on the instrument, an additional screen stating  
“data in memory” will appear prompting you to reset the instrument, if desired.  
A
B
Figure 1-3: Start-up screens (A indicates model and type  
B- indicates revision and battery power)  
2. In the Index Start-Up screen, press I/O Enter key. (NOTE: if an index is  
selected from the Setup menu, this will appear on this screen.)  
Figure 1-4: Index start-up screen  
3. A Log Rate Start-Up screen will appear, press I/O Enter key. (This screen  
details your log rate time or event logging setup option and the last saved  
logged session or event.) (See Setup, page 8 for more information on event  
logging.)  
Figure 1-5: Log rate start-up screen  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
5
Getting Started  
Turning on, logging/capture an event, and turning off  
4. The main menu (navigational screen) will appear after the start-up screens.  
Press I/O Enter key. (This will open to the measurement screens.)  
Indicator arrow  
Indicates the selected menu option.  
Either press up/down arrows to  
select a new menu or press I/O  
Enter key to select.  
Figure 1-6: Main menu of the QT48N  
5. Press Run/Stop key to either Data Log or log/record an Event and an asterisk  
will appear in far right corner. (NOTE: Data logging or an Event is selected  
via the Setup menu.)  
Numeric counter  
o Event logging/recording: As you record events, the numeric counter will  
increase by one displaying the number of saved events.  
o Data logging mode: This will display the number of saved data logged  
sessions instead of Events. (The process is identical to event mode.)  
Numeric counter  
Event records or Data logging  
sessions displayed  
Run indicator  
Will appear when you are in “Run  
mode”. For an event, it will appear for  
about 1-2 seconds as an event is  
recorded in the instrument’s history.  
Figure 1-7: Measurement screen (example)  
6. To stop data logging, press and hold Run/Stop key. (NOTE: please skip this  
step for an Event log. An event log will capture the measurements with the  
initial run/stop key press.)  
7. To view different measurements, press the Up or Down Arrow key to toggle  
through the views.  
8. To return to the main menu, press and hold the I/O Enter key (3, 2, 1  
countdown will appear) and the main menu will display.  
To select an option on the main menu, press the Up or Down Arrow until  
an arrow appears directly in front of the appropriate menu selection and  
then press I/O enter key.  
9. To power off, press and hold the I/O Enter key from the main menu.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Sensors  
Sensors  
Globe Thermometer  
6
Sensors  
Globe Thermometer  
The globe thermometer (left position) gives an indication of the radiant heat  
exposure on an individual due to either direct sunlight or hot objects in the  
environment. This is accomplished by placing a temperature sensor inside a  
blackened copper sphere and measuring the temperature rise. The WBGT index  
is based on the response of a 6 inch diameter globe. The QUESTemp uses a 2  
inch diameter globe for a faster response time. The temperature of the 2 inch  
globe is correlated to match that of a 6 inch globe.  
Waterless Wetbulb & Relative Humidity Sensor  
The relative humidity sensor (middle position) is used to calculate the Waterless  
Wetbulb from a combination of dry bulb temperature, humidity and wind speed  
measurements. The waterless wetbulb is used to calculate an estimated WBGT  
value.  
Dry Bulb Thermometer  
The dry bulb thermometer (right position) measures the ambient air  
temperature. This measurement is used in the outdoor WBGT calculation when  
a high solar radiant heat load may be present. The series of white plates  
surrounding the sensor shield it from radiant heat.  
B
C
A
A. Globe thermometer  
B. Relative humidity sensor  
C. Dry bulb thermometer  
Figure 1-8: Sensors identified  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Measurements  
7
Measurements  
WetBulb Globe Temperature (WBGT)  
Measurements  
The QUESTemp48N data logging, with an event logging mode, heat stress  
monitor directly senses three parameters: dry bulb temperature (DB), globe  
temperature (G), and relative humidity (RH).  
It computes the Wet Bulb (WB), the Wet Bulb Globe Temperature (WBGT), and the  
Stay Times Indices for ACGIH, U.S. Navy PHEL charts, and Flag Conditions for U.S.  
Navy/Marine Corp. Ashore.  
WetBulb Globe Temperature (WBGT)  
The WBGT is a type of temperature index which combines the effects of  
temperature, humidity, radiant heat, and air flow and provides a relatively simplistic  
tool to analyze thermal comfort. The resulting WBGT values can then be compared  
to indices of work-rest regimens (stay times) based upon workloads.  
The WBGT calculation is a weighted average of the three temperature sensors using  
the following formulas:  
**NOTE: Navy and Marine calculations are all based on the outdoor computation.  
WBGT (outdoor) = 0.7WB + 0.2G + 0.1DB (denoted as “WBGT” on the display)  
Stay Times/Rest Times (Indicies)  
Stay Times Indices represent how long a worker should be able to safely work  
under heat stress conditions. Select one of four indices for displaying and  
printing from the unit: ACGIH Stay Times, NAVY PHEL’s, or Flag Conditions for  
U.S. Navy/Marine Corp. Ashore. Refer to Appendix B for more information on  
the indices. NOTE: it will appear as “Index” on the measurement screen.  
Remote measurements  
The top sensor bar may be removed from the instrument and used through a  
remote cable. (Please refer to Appendix C for remote cable options.) Shelter  
the instrument and remote the sensor bar if the measured environment is  
expecting heavy rain or if temperatures are above 60C.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Operating  
Operating QUESTempº 48N  
8
View  
Operating QUESTempº 48N  
From a powered on mode, the main menu  
will appear (after the start-up screens are  
displayed.) Use the Up Arrow and Down  
Arrow keys to move the marker in the  
display in front of the desired mode. Pressing  
the I/O Enter key will select the mode.  
Figure 1-9: Main menu example  
View  
Displays the measured data but does not log it. If more than one set of sensors  
is plugged into the unit, they can be displayed by pressing and releasing the I/O  
Enter key. The displayed sensor set is shown in the upper right corner.  
View mode  
Arrow icon Indicates you are in view mode  
Event #  
Last logged event/session  
Figure 1-10: Viewing measured data  
NOTE: To return to the main menu, hold down the I/O Enter key while a  
3, 2, 1 countdown is shown in the lower right corner of the display.  
Setup  
The setup options are from the main menu when Setup is selected. At any time,  
to exit (or escape) a screen, press the Run/Stop key. Table 1-2 identifies the  
setup parameters.  
Settings  
Explanation  
Temperature units  
Selectable: ºF/ºC  
Language setting  
Time and Date  
Selectable: English, Spanish, French, Italian, and German  
Time settings: 24-hour clock  
Day-month-year format  
Log Rate  
1, 2, 5, 10, 15, 30, 60 minutes log rate options  
Event Log: when run/stop is pressed, an event or snapshot  
of the current conditions is logged into memory. The events  
range from 0-99.  
Index/Stay Times  
ACGIH, PHEL, FLAG, or none (denoted “----”)  
Table: 1-2: Settings indentified  
Download from Www.Somanuals.com. All Manuals Search And Download.  
           
Operating:  
Setup  
9
Operating QUESTempº 48N  
Setup  
Selecting Setup parameters  
1. From the main menu, select Setup by pressing the I/O Enter key. Using  
the Up/Down Arrow keys, select the settings in steps 2-6.  
Press I/O Enter key to select Setup.  
Figure 1-11: Selecting temperature setting and setup  
2. For the Time and Date settings, once selected, press I/O Enter key to  
move to the update field. (An underline will appear under the first field.)  
Press Up/Down Arrows to change the number/month. Press I/O  
Enter key to toggle through each field. Repeat as necessary until all  
values are selected.  
3. For the Temperature setting, either Fahrenheit or Celsius will appear.  
Press the Up/Down Arrow to select and I/O Enter key to change.  
Temperature setting press I/O Enter key  
to switch between values.  
Figure 1-12: Selecting temperature setting  
4. For the Language setting, once selected, press I/O Enter key. To select a  
language, press Up/Down Arrow keys. Once selected, press Run/Stop  
key to return to setup parameters.  
5. For the Log Rate, once selected, press I/O Enter key.  
Repeatedly press I/O Enter keys to select either the Log Rate Event or  
a Log Rate interval (for data logging; e.g. “5”).  
Optional: changing Next Event field. Events are numbered starting  
at 1 and incrementing. The number may be increased to simulate  
skipping events. This may be useful if the events represent locations  
and the starting location is somewhere other than 1.  
Event  
Example of setting Log Rate  
to Event Log  
Next event number  
*To increase, press up arrow key (used  
to simulate skipping an event.)  
Figure 1-13: Selecting log rate with event log  
6. For the Index setting, once selected press I/O Enter key. Press  
Up/Down Arrow keys to select: ---” (which is equivalent for no index),  
ACGIH, PHEL, or Flag.  
7. Exit Setup by pressing the Run/Stop key.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Operating:  
Print, reset, run  
10  
Operating QUESTempº 48N  
Print  
Print  
Allows printing to a parallel or serial printer or to a computer. The QUESTempº  
48N will recognize the cable plugged in and configure itself for serial or parallel.  
If no cable is plugged in, it will default to serial. Press I/O Enter key to begin  
printing. Press Run/Stop key to return to the menu.  
NOTE: if you wish to stop the printing, press I/O Enter key until you return  
to the main menu. When the printer has stopped printing, remove the cable  
from the printer to the instrument.  
Reset  
Resetting enables you to clear the logged data from memory. Press the I/O  
Enter key to enter the Reset mode. Clear the memory by holding down the  
I/O Enter key while the display counts down from three.  
Run  
For Data Logging setup, the run mode begins a session in memory and logs the  
data. For Event Logging, each run/stop is stored as an event and a numeric  
counter displays the number of saved events.  
1. Begin a session by pressing the Run/Stop key from the view mode (or  
measurement view). An asterisk in the lower right corner indicates the run  
mode.  
To toggle through the views, press the Up or Down Arrow.  
Run indicator  
Figure 1-14: Run mode indicator  
2. End the session by pressing the Run/Stop key again. (The session will stop  
recording when the asterisk is no longer displayed.)  
NOTE: If the logging memory is full or if there are no sensors plugged into  
the unit, attempting to enter the Run mode will result in an error message. If  
the memory capacity is exceeded, the asterisk in the lower right corner of the  
display will turn into an “F” and the memory remaining screen will show “0.0”.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Displayed items  
11  
Displayed Items  
Run  
Displayed Items  
For the QUESTemp48N, the number in the upper right corner indicates the  
saved session or the saved event log.  
1indicates the sensor bar placed on (or attached to) the top of the  
instrument. Sensors 2 and 3 are labeled on the side of the unit as  
“Sensor 2”, and “Sensor 3”.  
An *asterisk in the lower right corner indicates that the unit is in the  
run mode and is logging data or saves an event.  
The following measurements can be accessed on the display:  
Screen 1: WET (Wet bulb)  
DRY (Dry bulb)  
Figure 1-15: Wet and Dry screen  
Screen 2: GLOBE and RH  
Figure 1-16: Globe and RH screen  
Screen 3: WBGT  
Figure 1-17: WBGT screen  
Screen 4: Time (24 hour format)  
Date (day, month, year)  
Figure 1-18: Time and Date screen  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
12  
Displayed Items  
Stay Time/Index  
Screen 5: BAT (Battery voltage)  
MEM (Logging memory  
available in days)  
Figure 1-19: Battery & memory  
screen  
Screen 6: Index  
(Navy PHEL stay times)  
Figure 1-20: Navy PHEL stay times  
NOTE: A series of dashes appear in the display if one of the following occur:  
The temperature is outside of its allowable range  
A temperature sensor has failed  
Stay times temperatures are outside of the their defined range  
Stay Time/Index  
The screen(s) displaying stay time data appear different for each of the possible  
indices.  
If ACGIH is selected, the recommended working minutes per hour are shown for  
each of the workload categories Light (L), Moderate (M), Heavy (H), and Very  
Heavy (VH).  
Figure 1-21: Index for ACGIH  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
13  
Data Logging  
Stay Time/Index  
If the Navy PHELs are selected, the recommended working hours are shown  
based on a maximum of eight hours. Three screens are used to display the  
PHELs two at time.  
NOTE: 8:01following one of the PHELs indicates greater than eight hours.  
PHELs stay time is greater than 8 hours.  
Figure 1-22: Navy PHELs screen  
If Flag is selected, the Flag Conditions for Navy/Marines Corp. Ashore warning  
system screen will appear. There are five flag systems, no flag, yellow, green, red,  
and black, which provide heat exposure guidelines for acclimated individuals. (For  
more information, please refer to the “Flag Conditions US Navy/Marine Corp.  
Ashore” on page 26.)  
Navy/Marine Corp’s heat  
condition flag warning example  
Figure 1-23: Flag Conditions Navy/Marine Corp. Ashore  
Data Logging  
Data from each sensor is recorded at the interval set by the logging rate. Every  
time Run/Stop is pressed, a session is either started or ended in memory.  
Each session contains a header with time, date, and summary information.  
Memory Table: Gives the number of logging DAYS.  
Log Rate  
1 sensor  
2 sensors  
3 sensors  
1 min  
11.2  
5.6  
2 min  
22.5  
11.2  
7.5  
5 min  
56.2  
28.1  
10 min  
112.4  
56.2  
15 min  
168.6  
84.3  
30 min  
337.3  
168.6  
112.4  
60 min  
674.5  
337.3  
224.8  
3.7  
18.7  
37.5  
56.2  
Table 1-3: Example of a Memory table  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Electronic Sensor  
check  
14  
Electronic sensor check  
Performing an electronic sensor check  
Electronic sensor check  
A verification module, Quest model 053-923, may be used to check the operation of the  
QUESTemp’s wet bulb, dry bulb, and globe. The purpose is to verify that the electronic  
components are within a specific range with known values and a known source. The  
temperature tolerances should be within +/-0.5oC.  
Example of Verification module  
NOTE: If the sensors are outside of the tolerances, this indicates the sensor alignment  
should be serviced off-site for calibration.  
Performing an electronic sensor check  
1. Ensure the instrument is reading in Celsius prior to your electronic sensor check.  
To change the temperature setting, select Setup from the main menu. Then,  
either Fahrenheit or Celsius will appear on the screen. When selected, press I/O  
enter key to switch between settings. For more information, please refer to page 9,  
step 3.  
2. Remove the top sensor bar, place to the side, and plug in the verification module into  
the center pins of the sensor housing.  
3. Verify the measurement readings on the screen are within +/- 0.5C tolerance to the  
readings printed on the verification module label. Example below:  
a. Wet Bulb (WB): 11.1C  
b. Dry Bulb (DB): 45.3C  
c. Globe (G): 69.2C  
NOTE: Relative humidity (RH) is not valid on this instrument.  
4. Once completed, remove verification module and place sensor bar back on the  
instrument. (Tighten down the two bolts.) Change the Celsius reading back to  
Fahrenheit. (Refer to step 1 a-b.)  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Sensor alignment  
Sensor Alignment  
RH Sensor Alignment  
15  
Sensor Alignment  
For highly accurate measurement readings, you should align your QUESTempº 48N prior to  
data logging or event logging/recording. For the RH sensor, it is recommended to align with  
a High concentration level using the 75% sensor salt, NaCl (sodium chloride), or a Low  
concentration level using the, the 33% sensor salt, MgCl (magnesium chloride). (To order  
sensor alignment salts, refer to an online vendor such as, colepalmer.com.)  
RH Sensor Alignment  
1. To open, navigate to the main menu and select View by pressing I/O Enter key (see  
A). A measurement screen will appear (see B).  
(A)  
(B)  
Figure 1-24: Main menu with view selected (A) & measurement screen (B)  
2. Press and hold I/O Enter key and then press Down Arrow key from the View menu.  
The Alignment screen will appear.  
(RH Sensor) Alignment menu  
“High” is selected when the arrow  
appears in front of the text.  
Note: WBGT is not used in sensor alignment  
Figure 1-25: Alignment screen  
3. Press Up/Down Arrow key to select either High or Low. Then press I/O Enter key.  
4. Remove (or slide) the Sunshield (white globe) from the RH sensor and place it to the  
side. (See Figure 1-26).  
5. Place a latex finger cot (user supplied) over the humidity sensor in order for the sensor  
to stabilize quicker about 15 minutes (or up to 60 minutes without). Note: latex finger cots  
can be found at various hardware stores or online such as colepalmer.com.  
Place unrolled finger cot, with a small  
hole cut in the top, so it is completely  
stretched over the humidity sensor.  
Sunshield:  
Slide off RH  
sensor to  
remove.  
Figure 1-26: Sunshield & placement of finger cot prior to RH alignment  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
Sensor Alignment  
RH Sensor Alignment  
16  
6. Place the Salt container, with the salt alignment cap removed, over the RH sensor  
(middle sensor). Allow level to stabilize for at least 15 minutes. (Refer to Figure 1-25.)  
About sensor alignment: if readings are within +/-0.5%, a change in sensor alignment  
is not required. Skip to step 7 to return to main menu.  
NOTE: For a High alignment, use the 75% sensor alignment salt (NaCL) and for a Low  
alignment, use the 33% alignment salt (MgCl).  
Alignment salt (remove rubber cap and place over RH  
sensor with the sunshield removed)  
Once Alignment salt is placed, allow  
levels to stabilize (about 15 mins).  
Low RH Alignment screen  
Adjust % to 33% (using up/down arrow  
keys) and press Enter (after level has  
stabilized) to store sensor alignment level.  
Figure 1-27: RH alignment with Low % example  
7. In either the High/Low RH alignment screen, adjust the level to the percentage displayed  
on the salt container by using the Up/Down Arrow keys. Press I/O Enter key to save.  
(The instrument will state either “Successful” or “Failure” if it passed/did not pass. For  
“Failure” repeat the steps above.)  
Example of high RH percentage prior to  
sensor alignment. Press up/down arrows  
to adjust level after salt has stabilized.  
Press Enter to save  
the alignment.  
Sensor alignment with High % RH level  
Figure 1-28: RH alignment with High% screen  
8. To return to the main menu, press Run/Stop key.  
9. Verify the RH sensor is +/-0.5% of the specific percentage. If not repeat the steps above.  
10. Remove the alignment salt and finger cot. Replace the Sunshield over the RH sensor  
when completed with the sensor alignment.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Printing  
Printing  
Data/Communications  
17  
Printing  
Printing options include to a RS-232 port or to a parallel printer. Serial transmission requires  
Quest cable #54-715. Parallel transmission requires Quest cable #56-875. With the  
applicable cable plugged in, select PRINT from the menu and press the I/O Enter key to  
enter the PRINT mode. Begin printing by pressing the I/O Enter key. Press the key again  
to abort the printing.  
Data/Communications  
3MTM Detection Management Software DMS is recommended for downloading, storing,  
graphing, and printing reports (see below). Communications programs, such as Window’s  
Hyperterminal may also be used to capture the printout into a file. The baud rate is fixed at  
9600.  
Figure 1-29: Sample DMS event data report  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
Printing  
Parallel  
18  
Parallel  
Data can be sent directly to parallel printers that accept direct ASCII test input without  
special drivers. Make sure the printer is powered on and is online, ready to accept  
data, prior to printing.  
QUEST TECHNOLOGIES, a 3M company  
HEAT STRESS REPORT  
Page 1  
EVENT: 1  
Sensor: 1  
Page 2  
Degrees Fahrenheit  
Stay Times: ACGIH, Acclimated, WBGT, clo correction = 1.0 °C  
File Name _________________________ QUESTempº48N Rev 1.00  
Serial # TK09090909  
Employee _________________________  
TIME WBGT DRY GLOBE RH HI FLOW L VH  
M
H
11:08 67.9 82.4 90.7  
11:09 68.1 82.6 91.3  
13  
12  
0
0
0.5 60 60 60 60  
0.5 60 60 60 60  
Facility _________________________ Session (3)  
Start: 21-NOV-09 11:07:32  
Department ________________________ Stop: 21-NOV-09 11:10:15  
EVENT: 2  
Page 3  
Sensor: 1  
Degrees Fahrenheit  
Job  
_________________________ Printed: 21-NOV-09 11:16:00  
Stay Times: ACGIH, Acclimated, WBGT, clo correction = 1.0 °C  
TIME WBGT DRY GLOBE RH HI L VH  
11:08 79.3 98.9 104.5 15 0 60 45 30 15  
11:09 80.2 99.2 105.6 15 0 60 45 30 15  
Comments/Notes______________________________________________  
____________________________________________________________  
M
H
Logging Interval: 1 minutes  
Degrees Fahrenheit  
EVENT: 3  
Sensor: 1  
Page 4  
MAXIMUM LEVELS, Sensor 1  
WBGT  
68.3 21-NOV-09 11:10:08  
82.7 21-NOV-09 11:09:56  
91.4 21-NOV-09 11:10:12  
Degrees Fahrenheit  
Stay Times: ACGIH, Acclimated, WBGTi, clo correction = 1.0 °C  
DRY BULB  
GLOBE  
REL HUMIDITY 14% 21-FEB-08 11:07:32  
FLOW (m/s) 0.6 21-FEB-08 11:09:08  
TIME WBGT DRY GLOBE RH HI L VH  
M
H
11:08 68.1 88.0 92.7  
11:09 68.4 88.3 92.9  
11  
11  
0
0
60 60 60 60  
60 60 60 60  
MAXIMUM LEVELS, Sensor 2  
WBGT  
DRY BULB  
GLOBE  
80.5 21-FEB-08 11:10:11  
99.2 21-FEB-08 11:09:07  
106.1 21-FEB-08 11:10:06  
EVENT: 4  
Page 5  
Sensor: WBGT(W-AVG) = .50*WBGT(1) + .25*WBGT(2) + .25*WBGT(3)  
Degrees Fahrenheit  
HEAT INDEX  
0
00-XXX-00 00:00:00  
REL HUMIDITY 15% 21-FEB-08 11:07:32  
Stay Times: ACGIH, Acclimated, WBGT, clo correction = 1.0 C WBGT  
WBGTo  
MAXIMUM LEVELS, Sensor 3  
WBGT  
DRY BULB  
GLOBE  
68.6 21-FEB-08 11:09:56  
88.6 21-FEB-08 11:10:08  
93.0 21-FEB-08 11:10:03  
TIME W-AVG W-AVG  
L
M
H
VH  
-----  
11:08 71.5  
11:09 71.8  
-----  
-----  
70.8  
71.1  
--- --- --- ---  
60 60 60 60  
60 60 60 45  
HEAT INDEX  
REL HUMIDITY 11% 21-FEB-08 11:07:32  
0
00-XXX-00 00:00:00  
MAXIMUM LEVELS, Sensor(WEIGHTED AVERAGE)  
WBGT  
71.4 21-FEB-08 11:10:14  
Figure 1-30: Sample printouts  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Power options  
Powering  
9-Volt Alkaline Battery Replacement  
19  
Powering  
The QUESTempº 48N uses a 9-volt alkaline battery. A door on the back of the  
instrument allows the user access to the 9-volt battery.  
The 2-position switch located in the battery compartment is  
set by the user to the 9-volt battery.  
Figure 1-31: 9-volt battery  
9-Volt Alkaline Battery Replacement  
WARNING: Replace batteries only in a non-hazardous environment.  
The 9-volt battery should be replaced should be recharged when the voltage drops  
below 6.4 volts. The battery voltage is displayed when the instrument is turned on.  
While turned on, the battery voltage can be displayed at any time by pressing the up or  
down arrow keys to move through the display until the battery voltage screen appears.  
If, while operating, the battery voltage drops below 6.4 volts, the display will  
automatically switch to the display showing the battery voltage along with a low battery  
message. After a low battery occurs, the unit will continue to operate for  
approximately 8 hours. When the battery voltage falls to 6.2 volts or below, the unit  
will automatically turn off.  
Replace only with an approved 9-volt alkaline battery.  
Approved 9-Volt Batteries  
Eveready: Energizer 522, EN22, 6LR61  
Duracell: MN1604  
Panasonic: 6LR61, 6AM6X  
Rayovac: A1604  
UltraLife: U9V  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
Appendix A  
Appendix A: Specifications  
Approved 9-Volt Batteries  
20  
Appendix A: Specifications  
Measurements  
Globe Temperature, Dry Bulb Temperature, Wet Bulb Temperature, % Relative  
Humidity, and WBGT Index  
Temperatures given in Celsius or Fahrenheit  
Index (displayed stay time with selected index): ACGIH TLV, U.S. Navy PHEL  
Charts, Flag Conditions for U.S. Navy/Marine Corp. Ashore  
Data Logging mode and Event logging mode  
Data logging mode records and prints all measurements at user selected interval of 1,  
2, 5, 10, 15, 30, or 60 minutes.  
Event logging mode (1 second record) 128K bytes of data memory.  
Languages  
English, French, Spanish, Italian, German  
Sensors  
Dry Bulb Sensor- 1000 Ohm Platinum RTD  
Accuracy and Ranges: +/-0.5 from 0°C to 120°C (+/-0.9°F from 32°F  
to 248°F)  
Waterless Wet Bulb (Humidity) Sensor  
Accuracy and Ranges: Expanded measurement uncertainty of 1.1°C  
(k=2) between 0°C and 80°C (32°F and 176°F)  
Globe Sensor- 1000 Ohm Platinum RTD  
Accuracy and Ranges: +/-0.5 from 0°C to 120°C (+/-0.9°F from 32°F  
to 248°F)  
Relative Humidity Sensor - Integrated circuit with capacitive polymer sensor  
Accuracy and Ranges: +/-5% from 20 to 95% (non-condensing)  
Operating Temperature Range  
Sensor Assembly: -5C to +100C  
Electronics: -5C to 60C  
Housing  
Designed water resistant to a light rain or mist. If rain is frequent, best practice  
would be to remote the sensor bar and keep the instrument sheltered.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
21  
Appendix A: Specifications  
Approved 9-Volt Batteries  
Size  
Height 9.2in (23.5cm); Width 7.2in (18.3mm); Depth 3.0in (7.5mm)  
Dimensions include mounted sensor assembly  
Weight  
2.6 lbs. (1.2 kg) with mounted sensor assembly  
Remote Sensor Bar  
Using a cable, the top sensor bar can be setup for remote measurements, up to  
200 feet (61m).  
Power  
9V alkaline  
Battery Life  
9V alkaline: 80 hours  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Block model  
22  
Appendix A: Specifications  
QUESTemp 48N Block Model  
QUESTemp° 48N Block Model  
Sensors  
1, 2, & 3  
Gain  
Multiplexer  
Battery  
Power  
Controller  
Keypad  
LCD  
A/D  
Microprocessor  
RS232  
Printer  
Data  
Memory  
Program  
Memory  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix B  
23  
Appendix B  
ACGIH  
Appendix B: Heat Exposure Tables  
ACGIH  
Screening Criteria for Heat Stress Exposure. WBGT values in C. NOTE: according  
to the ACGIH’s guidelines, the temperature values represent a work and rest process  
which is explained in the standards. Please refer to the ACGIH TLVs and BEIs for  
specific details.  
Work and recovery  
(TLV )  
Light  
Moderate Heavy  
Very Heavy  
75% to 100%  
50% to 75%  
25% to 50%  
0% to 25%  
31.0  
28.0  
29.0  
30.0  
31.5  
26.0*  
27.5  
29.0  
30.5  
23.5*  
25.5*  
28.0  
31.0  
32.0  
32.5  
30.0  
Work and recovery  
(Action Limit)  
75% to 100%  
50% to 75%  
Light  
Moderate Heavy  
Very Heavy  
28.0  
28.5  
29.5  
30.0  
25.0  
26.0  
27.0  
29.0  
22.5*  
24.0  
25.5  
28.0  
20.0*  
22.5*  
24.5  
25% to 50%  
0% to 25%  
27.0  
*Values not specified by ACGIH have been estimated for continuity.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
24  
Appendix B  
ACGIH Clothing Corrections  
ACGIH Clothing Corrections  
The following clothing corrections are in degrees Celsius. When a clothing correction  
is entered into the setup portion of the QUESTemp48N, the value is added to the  
WBGT only for looking up the stay times. The WBGT value displayed by the unit  
does not reflect corrections.  
Clothing type  
Clothing correction (Addition to WBGT (ºC)  
Work clothes (long sleeve shirt and pants)  
Cloth (woven material) coveralls  
Double-layer woven clothing  
SMS polypropylene coveralls  
Polyolefin coveralls  
0º  
0º  
3º  
0.5º  
1º  
Limited-use vapor-barrier coveralls  
11º  
Cited from "American Conference of Governmental Industrial Hygienists - Threshold  
Limit Values and Biological Exposure Indices for 2008"; Reprinted with permission  
from ACGIH  
United States Navy  
Physiological Heat Exposure Limits (PHELs) Time Table  
(Without the presence of fuel combustion gases/fuel vapors)  
The recommended working hours are shown based on a maximum of eight hours.  
Naval personnel will follow a category, I - VI, based upon their function.  
PHEL Curves (Total Exposure Time in Hours: Minutes)  
WBGT(F)  
I
II  
III  
IV  
V
VI  
80.0  
81.0  
82.0  
83.0  
84.0  
85.0  
86.0  
87.0  
88.0  
89.0  
90.0  
91.0  
>8:00  
>8:00  
>8:00  
>8:00  
8:00  
8:00  
7:45  
7:05  
6:30  
5:55  
5:25  
5:00  
4:35  
>8:00  
>8:00  
8:00  
7:45  
7:05  
6:30  
5:55  
5:25  
4:55  
4:30  
4:10  
3:50  
8:00  
8:00  
7:05  
6:25  
5:55  
5:20  
4:55  
4:30  
4:05  
3:45  
3:25  
3:10  
6:35  
4:30  
>8:00  
>8:00  
>8:00  
>8:00  
8:00  
6:35  
5:25  
4:55  
4:30  
4:05  
3:45  
3:25  
3:10  
2:50  
2:40  
2:25  
4:30  
3:40  
3:20  
3:05  
2:50  
2:35  
2:20  
2:10  
2:00  
1:50  
1:40  
8:00  
7:25  
6:45  
6:10  
5:40  
5:15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
25  
Appendix B  
United States Navy  
WBGT(F)  
I
II  
III  
IV  
V
VI  
92.0  
4:50  
4:25  
4:05  
3:45  
3:25  
3:10  
2:55  
2:40  
2:30  
2:20  
2:10  
2:00  
1:50  
1:40  
1:35  
1:30  
1:20  
1:15  
1:10  
1:05  
1:00  
0:55  
0:55  
0:50  
0:45  
0:45  
0:40  
0:35  
0:35  
0:35  
0:30  
0:30  
0:25  
4:10  
3:50  
3:35  
3:15  
3:00  
2:45  
2:35  
2:20  
2:10  
2:00  
1:50  
1:45  
1:35  
1:30  
1:25  
1:15  
1:10  
1:05  
1:00  
1:00  
0:55  
0:50  
0:45  
0:45  
0:40  
0:40  
0:35  
0:35  
0:30  
0:30  
0:25  
0:25  
0:25  
3:30  
3:15  
3:00  
2:45  
2:30  
2:20  
2:10  
2:00  
1:50  
1:40  
1:35  
1:25  
1:20  
1:15  
1:10  
1:05  
1:00  
0:55  
0:50  
0:50  
0:45  
0:40  
0:40  
0:35  
0:35  
0:30  
0:30  
0:25  
0:25  
0:25  
0:20  
0:20  
0:20  
2:55  
2:40  
2:25  
2:15  
2:05  
1:55  
1:45  
1:40  
1:30  
1:25  
1:15  
1:10  
1:05  
1:00  
0:55  
0:50  
0:50  
0:45  
0:40  
0:40  
0:35  
0:35  
0:30  
0:30  
0:25  
0:25  
0:25  
0:20  
0:20  
0:20  
0:15  
0:15  
0:15  
2:15  
2:00  
1:50  
1:45  
1:35  
1:25  
1:20  
1:15  
1:10  
1:05  
1:00  
0:55  
0:50  
0:45  
0:45  
0:40  
0:35  
0:35  
0:30  
0:30  
0:25  
0:25  
0:25  
0:20  
0:20  
0:20  
0:15  
0:15  
0:15  
0:15  
0:15  
0:10  
0:10  
1:30  
1:25  
1:15  
1:10  
1:05  
1:00  
0:55  
0:50  
0:45  
0:45  
0:40  
0:35  
0:35  
0:30  
0:30  
0:25  
0:25  
0:25  
0:20  
0:20  
0:20  
0:15  
0:15  
0:15  
0:15  
0:10  
0:10  
0:10  
0:10  
0:10  
0:10  
0:10  
0:05  
93.0  
94.0  
95.0  
96.0  
97.0  
98.0  
99.0  
100.0  
101.0  
102.0  
103.0  
104.0  
105.0  
106.0  
107.0  
108.0  
109.0  
110.0  
111.0  
112.0  
113.0  
114.0  
115.0  
116.0  
117.0  
118.0  
119.0  
120.0  
121.0  
122.0  
123.0  
124.0  
Download from Www.Somanuals.com. All Manuals Search And Download.  
26  
Appendix B  
Flag Conditions for U.S. Navy/Marine Corp. Ashore  
Flag Conditions for U.S. Navy/Marine Corp. Ashore  
The following chart details the heat stress monitoring Flag Conditions for  
U.S. Navy/Marine Corp. Ashore in Degrees Fahrenheit and Celsius.  
No Flag  
Green  
80.0-  
84.9  
Yellow  
85.0-  
87.9  
Red  
Black  
>90  
Flag Degrees F  
Flag Degrees C  
<80.0  
88.0-  
89.9  
<26.7  
26.7-  
29.4  
29.4-  
31.1  
31.1-  
32.2  
>32.2  
* Rest means minimal physical activity (sitting or standing) and should be  
accomplished in the shade if possible.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
Appendix C  
27  
Appendix C: Accessories (sold separately)  
Appendix C: Accessories (sold separately)  
Replacement Relative humidity sensor (one included)  
Sensor array with 2-inch Globe and removable RH sensor (one included) 57-909  
Sensor alignment salt at 75% [ NaCl (sodium chloride)] (one included) 100-452  
Sensor alignment salt at 33% [ MgCl (magnesium chloride)] (one included) 100-451  
57-317  
Verification module (one included)  
Neck Strap (one included)  
QT-3X series replacement battery cover  
QT-3X series replacement dry bulb shield assembly-  
consists of (5) shields and (1) cap/cover.  
Remote sensor cable - 6 Foot  
Remote sensor cable - 25 Foot  
Remote sensor cable - 100 Foot  
Remote sensor cable - 200 Foot  
Serial/Computer interface cable (RS-232 to PC) 9 Pin  
RS232 to USB converter (one included)  
Parallel printer interface cable  
Storage case (one included)  
QT-48N’s Owner’s manual (one included)  
53-923  
057-333  
QT3X-BC  
QT3X-DBS  
53-924  
53-925  
53-926  
53-927  
54-715  
53-810  
56-875  
53-922  
57-330  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
Appendix D  
28  
Appendix D: Software  
Appendix D: Software  
The QUESTemp48N has the flexibility to be set up and controlled through  
computer software. The programmable start and stop time feature is only  
accessible through the computer. The instrument also has the capability of  
sending live data while measuring. These features are best utilized using 3MTM  
Detection Management Software DMS.  
DMS quick overview  
The focus of this section is to briefly introduce the following QSP-II topics:  
Downloading your data, setting up parameters, and viewing your data in charts,  
graphs, and reports. (For further details on DMS, please refer to the online Help  
and select Contents.)  
Communicating and Downloading in DMS  
In order to download, review the data, and setup parameters in DMS software,  
this will require connecting the QT48N to a computer. The steps below explain  
connecting and communicating to DMS.  
1. Using a 3M Quest cable, part number 054-715, plug the cable into the  
computer and plug the opposite end into the side jack of the QT48 data  
port.  
QT48 data port  
Connected and powered on  
Connected to serial port  
Figure 1-33: Communicating with the QT48N and DMS  
2. From the start page of DMS, select  
communication panel will appear.  
button and the instrument  
3. Select Heat Stress and then select the Model Type by clicking on  
QT48N. (See Figure 1-34).  
Download from Www.Somanuals.com. All Manuals Search And Download.  
       
29  
Appendix D: Software  
Data Finder and quick reports  
4. Click on the  
button in the Instrument Communications page while  
Data finder checkbox is checked (see 3 below).  
Note: The data Finder page will appear. See next section.  
3
1
2
Download  
Data Finder checkbox. When  
checked, this will automatically  
open to the downloaded data page.  
Double-click to  
select/view  
QT44/46 data in  
charts and graphs  
2
3
4
1
Figure 1-34: Communicating with the QT48N and DMS  
Data Finder and quick reports  
In the Data finder page, the data you downloaded is stored by instrument family and  
then organized by models. Note: when working in this page, click on the Most  
Recent button if you are looking for your recently downloaded data.  
1. In the data finder page, click on the downloaded data and select either analyze or  
print report.  
Double-click to  
select/view  
QT44/46 data in  
charts and graphs  
1
Figure 1-35: Data Finder Page  
Download from Www.Somanuals.com. All Manuals Search And Download.  
     
30  
Appendix D: Software  
Data Finder and quick reports  
Quick Report: select the  
button to generate a report with basic  
information and a summary table of your downloaded data.  
Figure 1-36: Quick Report  
Analyze: allows you to add charts/graphs and customize how they will appear  
in a report. Tip: add the panels by using the add a panel and dragging them  
into the appropriate location.  
Press  
button and it will generate a report based on the where the  
charts/graphs (panels) are placed.  
Figure 1-37: Analyze button opens to the panel layout page  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
31 Appendix D: Quest Service  
Contacting Quest Technologies  
Customer service  
Contacting 3M Instrumentation  
Should your 3M equipment need to be returned for repair or for recalibration, please  
contact the service department at the following number or access the online form via  
the website. For technical issues, please contact Technical Support.  
Service Department and Technical Support: 1 (800) 245-0779.  
Fax: 1 (262) 567-4047. Office hours are 8:00 a.m. to 5:00 p.m. United States Central.  
International customers  
Contact your local, factory-authorized distributor from whom the product was  
purchased. You can obtain the name and contact information of your local factory-  
authorized distributor from Quest by using the e-mail, telephone, or fax information  
given under “Contacting 3M” above.  
Calibration  
The QUESTemp48N and 3M field calibrator devices should be examined regularly by  
the factory. An annual calibration is recommended. (Please see Service Department  
above.)  
Warranty  
3M warrants our instruments to be free from defects in materials and workmanship  
for one year under normal conditions of use and service. For United States  
customers, we will replace or repair (our option) defective instruments at no charge,  
excluding batteries, abuse, misuse, alterations, physical damage, or instruments  
previously repaired by other than 3M. Microphones, sensors, printers, and chart  
recorders may have shorter or longer warranty periods. This warranty states our total  
obligation in place of any other warranties expressed or implied. Our warranty does  
not include any liability or obligation directly resulting from any defective instrument  
or product or any associated damages, injuries, or property loss, including loss of use  
or measurement data.  
For warranty outside the United States, a minimum of one year warranty applies  
subject to the same limitation and exceptions as above with service provided or  
arranged through the authorized 3M distributor or our 3M European Service  
Laboratory. Foreign purchasers should contact the local 3M authorized sales agent  
for detail.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
         
About Us  
3M Detection Solutions is a world class manufacturer of rugged, reliable instrumentation and software  
systems that help monitor and evaluate occupational and environmental health and safety hazards, including  
noise dosimetry, sound level monitoring, heat stress, indoor air quality and select toxic/combustible gases.  
The 3M Detection brand of instrumentation is used by safety and industrial hygiene professionals to help  
comply with applicable occupational standards and regulations.  
About 3M Personal Safety  
3M offers a comprehensive, diverse portfolio of Personal Safety solutions providing respiratory  
protection, hearing protection, fall protection, reflective materials for high visibility, protective clothing,  
protective eyewear, head and face protection, welding helmets, and other adjacent products and  
solutions such as tactical safety equipment, detection, monitoring equipment, active communications  
equipment and compliance management. In 2012, 3M celebrated 40 years of safety leadership –  
recognizing the company’s respiratory and hearing protection solutions introduced in 1972. Visit  
Personal Safety Division  
3M Detection Solutions  
1060 Corporate Center Drive  
Oconomowoc, WI 53066  
ISO 9001 Registered Company  
ISO 17025 Accredited Calibration Lab  
Customer Service: 262-567-9157  
Toll Free: 800-245-0779  
3M is a trademark of 3M Company  
used under license in Canada.  
Please recycle. Printed in USA.  
© 2013 3M  
All rights reserved.  
057-330 Rev.E. 2/13  
Download from Www.Somanuals.com. All Manuals Search And Download.  

3M Oxygen Equipment 600107046 User Manual
Aastra Telecom Telephone 1600IP User Manual
Advantech Computer Hardware AIMB 210 User Manual
Agilent Technologies Electronic Keyboard 03336 90011 User Manual
Akai Musical Instrument Ewi4000s User Manual
AOSafety Welding System Heat Reflective Aluminum Infused Headgear User Manual
Audiovox Baby Monitor DPF710K User Manual
Audiovox Remote Starter PRO 9051 User Manual
Avocent Stereo Amplifier DA1200P PC VGA User Manual
Beckett Water Gardening Water Dispenser PF357 User Manual