| 	
		 Operating Manual   
					Model 34700-2K/17700-2K   
					Recovery/Recycling/Recharging Unit   
				Table of Contents   
					Introduction .............................................................................................. 2   
					Glossary of Terms ................................................................................. 2   
					Setup Instructions.................................................................................... 2   
					Initial Setup............................................................................................. 4   
					Vacuum Pump Initial Fill ......................................................................... 5   
					Installation Routine ................................................................................. 5   
					Operating Guidelines............................................................................... 6   
					Using the Selection Menu....................................................................... 6   
					Change Filter .......................................................................................... 6   
					Recycle................................................................................................... 7   
					Tank Refill............................................................................................... 7   
					Vacuum Oil Time .................................................................................... 7   
					Filter Capacity......................................................................................... 7   
					Basic/Advanced Prompts ....................................................................... 8   
					Selecting a Unit (Metric/English) ............................................................ 8   
					Language Select..................................................................................... 8   
					Change Defaults..................................................................................... 9   
					Using the Control Panel.......................................................................... 9   
					Keypad Functions................................................................................. 10   
					Operating Instructions........................................................................... 11   
					Operating Tips ...................................................................................... 11   
					Recovering Refrigerant......................................................................... 12   
					Evacuating the A/C System.................................................................. 14   
					Replenishing A/C System Oil ............................................................... 16   
					Recharging the A/C System ................................................................. 17   
					Maintenance Instructions...................................................................... 19   
					Replacing the Filter-Drier...................................................................... 19   
					Changing the Vacuum Pump Oil .......................................................... 20   
					Checking for Leaks............................................................................... 21   
					Electrical Protection.............................................................................. 22   
					General Maintenance ........................................................................... 22   
					Replacement Parts List ......................................................................... 22   
					Flow Diagram.......................................................................................... 24   
					Wiring Diagram....................................................................................... 25   
					Limited Warranty.................................................................................... 26   
					U.S. Patents: 4,523,897; 4,688,388 Re 33,212; 4,768,347; 4,805,416; 4,809,520; 4,878,356; 4,938,031;   
					5,005,369; 5,005,375; 5,038,578; 5,042,271; 5,209,653; 5,248,125; Australian Patent: 613,058; Canadian   
					Patents: 1,311,621; 1,311,622; 2,012,620; 2,026,348; European Patent: 0 315 296 Bl; German Patent: 031296   
					Mexican Patent: 16208 OTHER U.S. AND FOREIGN PATENTS PENDING.   
					Mfd. by Robinair, SPX Corporation, Montpelier, OH 43543   
					1 
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Introduction   
					This manual contains important safety procedures concerning the   
					operation, use, and maintenance of this product. Failure to follow the   
					instructions contained in this manual may result in serious injury. If you   
					are unable to understand any of the contents of this manual, please bring   
					it to the attention of your supervisor. Do not operate this equipment   
					unless you have read and understood the contents of this manual.   
					The 34700-2K models are used for R-134a vehicles, whereas the 17700-2K models   
					are used for R-12 vehicles. Both models are designed to be compatible with   
					existing service equipment and standard service procedures.   
					The 34700-2K and the 17700-2K models are UL-listed, single-pass systems that   
					meet the SAE specifications for recycled refrigerant.   
					To validate your warranty, complete the warranty card attached to your unit, and   
					return it within ten days from date of purchase.   
					Glossary of Terms   
					A/C System   
					Unit   
					The air conditioning system being serviced.   
					The refrigerant recovery/recycling/recharging unit.   
					Internal Storage Vessel The refillable refrigerant storage vessel designed   
					specifically for this unit.   
					Source Tank   
					A disposable tank of new refrigerant used to refill   
					the internal storage vessel.   
					Setup Instructions   
					Low Side   
					Gauge   
					High Side   
					Gauge   
					Keypad   
					2 
					5 
					0 
					2 
					0 
					7 
					R-134A   
					R-134A   
					bar   
					0 
					0 
					0 
					VAC-CHARGE   
					CHARGE   
					bar   
					3 
					RECOVER   
					VACUUM   
					kPa   
					kPa   
					0 
					0 
					0 
					5 
					0 
					2 
					3 
					4 
					4 
					1 
					in Hg   
					VAC   
					psi   
					psi   
					START   
					STOP   
					HIGH   
					CLOSED   
					CLOSED   
					1 
					4 
					2 
					5 
					8 
					0 
					3 
					6 
					F1   
					MENU   
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					7 
					9 
					CLEAR   
					ENTER   
					VAC-CHARGE   
					STOP   
					CHARGE   
					RECOVER   
					VACUUM   
					0 
					I 
					START   
					1 2 3   
					4 5 6   
					7 8 9   
					F1   
					MENU   
					Low Side   
					Valve   
					High Side   
					Valve   
					Main Power   
					Switch   
					CLEAR   
					ENTER   
					0 
					INST0925   
					Diagram of the Control Panel   
					2 
					© 2001 Robinair, SPX Corporation   
				Setup Instructions   
					1 
					2 
					Diagram of Unit’s Components—   
					Side View   
					3 
					4 
					HIGH   
					LOW   
					1. 1-800 Phone Number Decal   
					2. High Side Inlet   
					3. Low Side Inlet   
					4. Power Cord with Tag   
					5. Fill Hose   
					6. Tank Strap   
					5 
					6 
					INST0691   
					VAC-CHHARGE   
					CHARGE   
					RECOVER   
					VACUUM   
					8 
					START   
					STOP   
					LOW   
					HIGH   
					CLOSED   
					CLOSED   
					1 
					4 
					2 
					5 
					8 
					0 
					3 
					6 
					F1   
					MENU   
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					7 
					9 
					CLEAR   
					ENTER   
					1 
					9 
					Diagram of Unit’s Components—   
					Internal View   
					10   
					CLOSE   
					OUNCES   
					12   
					MLS   
					340   
					320   
					300   
					280   
					260   
					240   
					220   
					200   
					180   
					160   
					140   
					120   
					100   
					80   
					11   
					10   
					9 
					6 
					8 
					7 
					6 
					5 
					4 
					3 
					2 
					60   
					40   
					1 
					20   
					1. Control Panel Assembly   
					2. Vacuum Pump   
					3. Internal Storage Vessel   
					4. Scale Assembly   
					5. Oil Drain Bottle   
					6. Hose Holder   
					FLOW   
					5 
					2 
					7 
					7. Air Purge Control   
					8. Vacuum Pump Oil Fill   
					9. Sight Glass   
					3 
					10. Oil Drain   
					INST0946   
					4 
					3 
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Setup Instructions   
					KEYPAD FUNCTIONS   
					In addition to the number keys, the keypad contains special keys that accomplish   
					specific operating functions.   
					START   
					STOP   
					START— Begins, or resumes, a function.   
					STOP— Terminates, or pauses, a function.   
					RECOVER— Activates the recovery sequence.   
					VACUUM— Activates vacuum and automatic recycling sequence.   
					RECOVER   
					VACUUM   
					VAC-CHARGE   
					CHARGE   
					MENU   
					VAC-CHARGE— Activates vacuum and automatic recycling sequence,   
					followed by a charge.   
					CHARGE— Charges A/C system with a programmed amount of   
					refrigerant.   
					MENU— Enters the selection menu.   
					UP/DOWN ARROWS— Scroll through menu items.   
					F1 (Inject Oil)— Injects oil into A/C system (active at end of vacuum).   
					F1   
					VAC-CHHARGE   
					CHARGE   
					RECOVER   
					VACUUM   
					STOP   
					START   
					1 2 3   
					4 5 6   
					7 8 9   
					Diagram of Keypad   
					F1   
					MENU   
					CLEAR   
					ENTER   
					0 
					4 
					© 2001 Robinair, SPX Corporation   
				Setup Instructions   
					INITIAL SETUP   
					CAUTION! R-134a systems have special fittings (per SAE   
					specifications) to avoid cross-contamination with R-12 systems. Do not   
					attempt to adapt your unit for another refrigerant — system failure will   
					result! Read and follow all warnings at the beginning of this manual   
					before operating the unit.   
					CAUTION! Avoid the use of an extension cord, because the   
					extension cord may overheat. However, if you must use an extension   
					cord, use a No. 14 AWG minimum, and keep the cord as short as   
					possible.   
					1. The first time the unit is powered up, it will start in the initial setup mode.   
					If the initial setup must be performed again, it may be selected using the   
					menu function.   
					2. The first step is to select a language. Use the UP and DOWN arrow keys to   
					select the desired language. Press START to save the currently displayed   
					language.   
					IMPORTANT!   
					You must press   
					the MENU key   
					to access all the   
					functions.   
					3. Next select the operating units. Toggle between UNITS ENGLISH and   
					UNITS METRIC using the arrow keys. Press START to save the currently   
					displayed choice.   
					4. Toggle between BASIC/ADVANCED using the ARROW keys. Use the   
					BASIC PROMPT option to receive step-by-step, on-screen prompting   
					through any procedure. Use ADVANCED PROMPT once you know the   
					procedure and no longer need the step-by-step routine. Press START to   
					save the currently displayed choice.   
					NOTE: This manual is written using the BASIC prompt option   
					Diagram of Hose Connections   
					Vacuum Pump Components   
					1. Fill Hose   
					2. Quick-Couplers (34700 Only)   
					1. Oil Filler Tube   
					2. Pump Exhaust   
					3. Oil Fill Port   
					4. Sight Glass   
					5. Oil Drain Fitting   
					6. Inlet   
					1 
					6 
					2 
					HIGH   
					LOW   
					3 
					4 
					OPEN   
					CLOSE   
					1 
					5 
					2 
					INST0692   
					INST0701   
					5 
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Setup Instructions   
					NOTE: The vacuum pump is shipped without oil in the reservoir. Before starting   
					the pump, oil must be added to the pump, or damage to the pump may occur.   
					IMPORTANT!   
					For maximum   
					performance,   
					change the   
					5. Press the START key to begin the oil fill process.   
					6. Attach the flexible tube and cap to the oil bottle, and pour eight ounces of   
					vacuum pump oil into the fill port.   
					vacuum pump   
					oil frequently.   
					7. Press the START key. While the vacuum pump is running, slowly add oil   
					until the level rises to the center of the reservoir's sight glass.   
					8. Press the STOP key, and replace the black plastic plug on the fill port.   
					9. Connect the service hoses, open both panel valves, and press START.   
					10. Connect the fill hose to a full source tank.   
					IMPORTANT!   
					The pump must   
					be running   
					when adding   
					oil.   
					11. Open the tank valve. Invert the tank and install it on the back of the unit,   
					making sure to secure the tank strap.   
					NOTE: If using a refillable tank, install the tank upside down, and connect the fill   
					hose to the vapor valve.   
					12. Press START, and the unit will automatically run a   
					five-minute vacuum to clear all internal air.   
					1 
					RECOVER   
					VACUUM   
					VAC-CHARGE   
					CHARGE   
					START   
					STOP   
					LOW   
					HIGH   
					CLOSED   
					13. After the vacuum is complete, press START to begin   
					CLOSED   
					1 
					4 
					2 
					5 
					8 
					0 
					3 
					6 
					F1   
					MENU   
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					7 
					9 
					CLEAR   
					ENTER   
					filling the internal storage vessel.   
					CLOSE   
					OUNCES   
					12   
					MLS   
					340   
					320   
					300   
					280   
					260   
					240   
					220   
					200   
					180   
					160   
					140   
					120   
					100   
					80   
					11   
					10   
					9 
					2 
					14. The unit stops when a sufficient amount of refrigerant   
					has been transferred to the internal tank, or when the   
					source tank is empty. Press the STOP key to pause the   
					process. Press STOP again to exit, or START to   
					resume. This process takes 15-20 minutes.   
					8 
					7 
					6 
					5 
					4 
					3 
					2 
					60   
					40   
					1 
					20   
					3 
					FLOW   
					1. Oil Fill   
					2. Sight Glass   
					3. Oil Drain   
					NOTE: Add at least 8 lb. (3.6 kg) of refrigerant before   
					stopping the process to ensure enough refrigerant is available   
					for charging.   
					INST0947   
					15. When the fill process is complete, you may press STOP   
					to exit.   
					16. The unit is now ready to operate.   
					NOTE: There is no need to calibrate the scale, because it is   
					calibrated at the factory.   
					6 
					© 2001 Robinair, SPX Corporation   
				Operating Guidelines   
					USING THE SELECTION MENU   
					1. Press the MENU button. The top line of the display reads SETUP MENU.   
					2. Use the UP and DOWN arrow keys to scroll through the menu choices   
					displayed on the second line. The menu choices are:   
					1. SELECT LANGUAGE   
					2. SELECT UNITS (ENGLISH/METRIC)   
					3. TANK REFILL   
					7. VACUUM OIL TIME   
					8. CHANGE VACUUM PUMP OIL   
					9. SELECT PROMPTS   
					4. RECYCLE ONLY   
					10. CHANGE DEFAULTS (password protected)   
					11. VERSION X.XX   
					5. FILTER CAPACITY   
					6. CHANGE FILTER   
					3. Press START to make a choice from the menu. Press STOP to pause any   
					process, and STOP a second time to exit any process.   
					CHANGE FILTER   
					The filter-drier removes acid, particulates, and water from the refrigerant. Change   
					the filter-drier after 150 pounds (68 kg) of refrigerant has been filtered. See the   
					REPLACING THE FILTER DRIER section on the following page, as well as the   
					Maintenance Section, for instructions.   
					RECYCLE   
					Manual recycling may be necessary if excessive air and/or moisture is recovered   
					from the A/C system.   
					1. Press the MENU key. Use the arrow keys to select RECYCLE ONLY, and   
					press START to begin.   
					2. Press the START button to start recycling. To pause recycling, press the   
					STOP key. To terminate recycling, press the STOP key again, or press   
					START to resume.   
					TANK REFILL   
					1. Press the MENU key. Use the arrow keys to select TANK REFILL, and press   
					START to begin.   
					2. Connect the fill hose to the source tank.   
					3. Open the tank valve. Invert the tank, install it on the back of the unit, and   
					secure the tank strap.   
					7 
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Operating Guidelines   
					NOTE: If using a refillable tank, install the tank upside down, and connect the fill   
					hose to the vapor valve.   
					4. Press the START key, and the tank automatically refills. The unit stops when   
					a sufficient amount of refrigerant has been transferred to the internal tank, or   
					if the source tank is empty. Press the STOP key to pause the process. Press   
					STOP again to exit, or START to resume.   
					5. When the fill process is complete, press STOP to exit.   
					VACUUM OIL TIME   
					This function displays how long the vacuum pump has run since the last oil   
					change.   
					1. Press the MENU key. Use the arrow keys to select VACUUM OIL TIME, and   
					press START to begin.   
					2. The display reads: OIL TIME = XX:XX This shows how long the pump has   
					run since the last oil change. The time resets to zero after a VACUUM PUMP   
					OIL CHANGE. See page 20 of this manual for details.   
					3. Press STOP to exit.   
					FILTER CAPACITY   
					This function is used to show the operator how many pounds or kilograms of   
					refrigerant have been recovered since the last filter change.   
					1. Press the MENU key. Use the arrow keys to select FILTER CAPACITY, and   
					press START to begin.   
					2. The display reads: FILTERED= XXXlbs(kg). This shows how much   
					refrigerant has passed through the filter. The amount filtered resets to zero   
					after a FILTER CHANGE. See page 19 of this manual for details.   
					3. Press STOP to exit.   
					SELECT PROMPT (BASIC/ADVANCED)   
					Use the BASIC PROMPT option to receive step-by-step, on-screen prompting   
					through any procedure. Use ADVANCED PROMPT once you know the procedure   
					and no longer need the step-by-step routine.   
					1. Press the MENU key. Use the arrow keys to choose SELECT PROMPT, and   
					press START to begin.   
					2. Toggle between BASIC/ADVANCED using the ARROW keys.   
					3. Press START to save the current choice and exit.   
					NOTE: This manual is written for the BASIC PROMPT option.   
					8 
					© 2001 Robinair, SPX Corporation   
				Operating Guidelines   
					SELECTING A UNIT (METRIC/ENGLISH)   
					1. Press the MENU key. Use the arrow keys to choose SELECT UNITS, and   
					press START to begin.   
					2. Toggle between UNITS ENGLISH and UNITS METRIC using the arrow   
					key.   
					3. Press START to save the current choice and exit.   
					LANGUAGE SELECT   
					The operator can choose between English, Spanish, French, Italian, or German.   
					1. Press the MENU key. Use the arrow keys to choose SELECT LANGUAGE,   
					and press START to begin.   
					2. Use the UP and DOWN arrows to scroll through the languages.   
					3. Press START to save the current choice. Press STOP to exit without saving.   
					CHANGE DEFAULTS   
					For service use only.   
					VERSION   
					Displays the current software revision.   
					9 
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Operating Guidelines   
					USING THE CONTROL PANEL   
					The control panel has various components that control specific operating functions.   
					MAIN POWER SWITCH—Supplies electrical power to the control panel.   
					DIGITAL DISPLAY—Used on the visual interface between the operator and   
					the machine.   
					LOW SIDE MANIFOLD GAUGE—Connects to an A/C system and shows the   
					system’s low side pressure.   
					HIGH SIDE MANIFOLD GAUGE—Connects to an A/C system and shows the   
					system’s high side pressure.   
					LOW SIDE VALVE—Controls the low side flow from the A/C system through   
					the unit.   
					HIGH SIDE VALVE—Controls the high side flow from the A/C system through   
					the unit. It has three positions: 1) Recover/Vacuum, 2) Closed, 3) Oil Inject/Charge.   
					1 
					2 
					3 
					4 
					5 
					0 
					2 
					2 
					0 
					7 
					R-134A   
					R-134A   
					bar   
					0 
					0 
					0 
					CHARGE   
					bar   
					3 
					RECOVER   
					VACUUM   
					VAC-CHARGE   
					kPa   
					5 
					kPa   
					0 
					0 
					0 
					0 
					5 
					0 
					4 
					2 
					3 
					4 
					1 
					in Hg   
					VAC   
					psi   
					psi   
					STOP   
					START   
					HIGH   
					CLOSED   
					CLOSED   
					1 2 3   
					4 5 6   
					7 8 9   
					F1   
					MENU   
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					CLEAR   
					ENTER   
					0 
					5 
					0 
					I 
					6 
					7 
					INST0926   
					Diagram of Control Panel   
					1. Low Side Gauge   
					2. High Side Gauge   
					3. Display   
					5. Main Power Switch   
					6. Low Side Valve   
					7. High Side Valve   
					4. Bezel & Keypad Assembly   
					10   
					© 2001 Robinair, SPX Corporation   
				Operating Instructions   
					OPERATING TIPS   
					Follow the SAE-J1991 recommended service procedure for the containment of   
					R-12, and the SAE-J2210 recommended service procedure for the containment of   
					R-134a.   
					The recovery compressor is not a vacuum pump. The compressor pulls the A/C   
					system to a partial vacuum only. You must use the unit’s vacuum cycle to remove   
					moisture from the A/C system. We recommend a minimum 15-minute vacuum, or   
					follow the system manufacturer's recommendations.   
					This unit is designed to be used with the manifold gauge set built into the   
					control panel.   
					The unit includes a 6 cfm (142 l/m) Robinair high vacuum pump for fast, thorough   
					evacuation. Change the vacuum pump oil after every 10 hours of use.   
					R-134a systems require special oils. Refer to the A/C system manufacturer’s   
					service manuals for oil specifications.   
					Pressing the START and STOP keys together for several seconds will exit any   
					mode and reset the control.   
					NOTE: The following operating instructions are written to be used with the   
					BASIC PROMPTS mode of operation. It is recommended that the BASIC   
					PROMPTS mode is used until the operator becomes very familiar with the   
					operation of the unit. See the OPERATING GUIDELINES section of this manual   
					for instructions on how to select between BASIC PROMPTS and ADVANCED   
					PROMPTS.   
					11   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Operating Instructions   
					RECOVERING REFRIGERANT   
					WARNING   
					Wear safety goggles when working with refrigerant. Read and follow   
					all warnings at the beginning of this manual before operating the unit.   
					1. Connect the power cord to the back of the unit, and plug it into the correct   
					voltage outlet.   
					2. Turn on the MAIN POWER and, if necessary, empty the oil drain bottle   
					located on the right hand side of the unit.   
					3. Press the RECOVER button.   
					4. If 150 pounds (68 kg) or more of refrigerant has been recovered since the last   
					filter-drier change, the display reads FILTER WEIGHT XXX lb (XX kg).   
					Press START.   
					NOTE: Refer to the filter change procedure on page 19 of this manual for details   
					about replacing the filter.   
					5. Connect the high and low side hoses to the A/C system, and open the coupler   
					valves.   
					6. Put the Low Side Valve in the Recover/Vacuum position. Put the High Side   
					Valve in the Recover/Vacuum position. Press START to continue.   
					7. If the system pressure is below 25 psi, the display reads: LOW SYSTEM   
					PRESSURE until the pressure increases or the START button is pressed. You   
					may press STOP to exit at this point.   
					8. If the unit has refrigerant in the low-side plumbing, it begins the clearing   
					process and displays CLEARING IN PROGRESS. If you wish to skip the   
					clearing operation or stop the clearing prematurely, press the START key.   
					Manifold Gauges   
					Recover   
					0 
					2 
					5 
					0 
					2 
					0 
					7 
					R-134A   
					R-134A   
					bar   
					0 
					0 
					0 
					VAC-CHARGE   
					CHARGE   
					bar   
					3 
					RECOVER   
					VACUUM   
					kPa   
					5 
					kPa   
					0 
					0 
					0 
					0 
					5 
					0 
					2 
					3 
					4 
					4 
					1 
					in Hg   
					VAC   
					psi   
					psi   
					STOP   
					START   
					HIGH   
					CLOSED   
					CLOSED   
					1 
					4 
					2 
					5 
					8 
					0 
					3 
					6 
					F1   
					MENU   
					7 
					9 
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					CLEAR   
					ENTER   
					0 
					I 
					INST0928   
					Diagram of Control Panel   
					During Recovery   
					Valves Open   
					12   
					© 2001 Robinair, SPX Corporation   
				Operating Instructions   
					9. When the system has recovered to a vacuum level of approximately 13 in. Hg.,   
					the compressor automatically shuts off.   
					10. The unit then goes into automatic oil drain, and the display reads: OIL   
					DRAINING. Oil draining can require up to 90 seconds to complete.   
					11. After the oil drain is complete, the display alternates between:   
					RECOVERY COMPLETE   
					CHECK OIL BOTTLE   
					RECOVERED XX.XX lbs. (X.XX kg)   
					RECOVERED XX.XX lbs. (X.XX kg)   
					NOTE: The displayed recovered weight can vary depending on ambient conditions,   
					and should not be used as an indicator of scale accuracy.   
					12. Check the oil drain bottle, and note the amount of oil that was removed from   
					the A/C system. This is the amount of oil that must be charged into the A/C   
					system after evacuation is complete.   
					13. To ensure complete recovery of refrigerant, wait 5 minutes, and watch the   
					manifold gauges for a rise in pressure above 0 in. Hg. A pressure rise may   
					occur if there was freezing in the A/C system during recovery. If a rise occurs,   
					press the START button to resume the recovery process. Repeat as needed   
					until the system pressure holds for two minutes, then press STOP to exit.   
					Recovery is now complete. You are now ready to make any repairs to the A/C   
					system, if necessary, or advance to the Evacuation Process.   
					Diagram of the Oil   
					Injection System   
					1. Oil Injector Bottle   
					2. Oil Drain Bottle   
					OUNCES   
					360   
					12   
					1 
					340   
					320   
					300   
					280   
					260   
					240   
					220   
					200   
					180   
					160   
					140   
					120   
					100   
					80   
					11   
					10   
					9 
					8 
					7 
					6 
					5 
					4 
					3 
					2 
					60   
					40   
					1 
					20   
					2 
					INST0699   
					13   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Operating Instructions   
					EVACUATING THE A/C SYSTEM   
					WARNING   
					Wear safety goggles when working with refrigerant. Use only   
					authorized refillable refrigerant tanks. Read and follow all warnings   
					at the beginning of this manual before operating the unit. In addition   
					to the number keys, the keypad contains special keys that   
					accomplish specific operating functions.   
					IMPORTANT!   
					Evacuate the   
					system for at   
					least 15   
					NOTE:   
					minutes to   
					ensure   
					adequate   
					moisture and   
					contaminant   
					removal.   
					• If any oil was drained from the system during recovery, DO NOT use the VAC-   
					CHARGE feature. The oil must be replenished into the A/C system, which is not   
					possible when the VAC-CHARGE function is used.   
					• If the vacuum pump has been run more than 10 hours since the last oil change,   
					the display reads: VACUUM OIL TIME XX:XX. Press the STOP key to change   
					the vacuum pump oil, or press the START key to continue. Instructions for   
					changing the vacuum pump oil are located in the maintenance section of this   
					manual. NOTE: Vacuum pump oil should be changed after every 10 hours of   
					use to maintain maximum performance and endurance levels.   
					• If the system being evacuated contains a pressure over 25 psi at any point during   
					the evacuation, the display reads PRESSURE EXISTS. This message indicates   
					that the A/C system contains refrigerant; press any key to continue. Press the   
					RECOVERY key to recover any refrigerant in the system (See RECOVERING   
					REFRIGERANT, page 12). After recovery is complete, return to evacuating the   
					A/C system.   
					IMPORTANT!   
					If the vacuum   
					pump has run   
					for 10 or more   
					hours without   
					an oil change,   
					the message   
					VACUUM OIL   
					TIME XX:XX   
					appears on the   
					display.   
					VAC-CHARGE   
					1. Press the VAC-CHARGE key to select the VAC-CHARGE feature.   
					2. Ensure the service hoses are connected and both panel valves are in the   
					VACUUM/ RECOVER position. Press START.   
					Vacuum   
					Vac-Charge   
					Change the   
					pump oil   
					6 
					0 
					2 
					5 
					0 
					The Control Panel   
					During Evacuation   
					0 
					6 
					2 
					5 
					0 
					0 
					0 
					6 
					following the   
					procedures in   
					the   
					MAINTENANCE   
					INSTRUCTIONS.   
					0 
					0 
					5 
					0 
					7 
					R-134A   
					R-134A   
					bar   
					8 
					0 
					0 
					0 
					0 
					0 
					0 
					VAC-CCHARGE   
					CHARGE   
					bar   
					3 
					RECOVER   
					VACUUM   
					8 
					kPa   
					5 
					kPa   
					0 
					0 
					0 
					0 
					5 
					0 
					4 
					2 
					3 
					4 
					1 
					in Hg   
					VAC   
					psi   
					psi   
					3 
					STOP   
					START   
					HIGH   
					CLOSED   
					CLOSED   
					1 
					4 
					2 
					5 
					8 
					0 
					3 
					6 
					F1   
					MENU   
					7 
					9 
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					CLEAR   
					ENTER   
					0 
					I 
					INST0929   
					14   
					© 2001 Robinair, SPX Corporation   
				Operating Instructions   
					NOTE: It is not necessary to change the High side panel valve from vacuum to   
					charge when performing the VAC-CHARGE function.   
					3. Press the START key to charge the default amount of refrigerant, or use the   
					number keys to enter the desired charge weight. Then press the START key.   
					4. If the weight entered leaves less than 3 lbs (1.36 kg) of refrigerant in the   
					internal storage vessel, the VAC-CHARGE process does not begin, and the   
					display reads INSUFFICIENT REFRIG. At this point, refrigerant must be   
					added to the internal storage vessel. See page 6 of this manual for internal   
					storage vessel refill instructions, and then return to Step 1 of EVACUATING   
					the A/C system.   
					IMPORTANT!   
					Evacuate the   
					A/C system for   
					at least 15   
					minutes to   
					ensure   
					adequate   
					moisture and   
					contaminant   
					removal.   
					5. If the internal storage vessel contains a sufficient amount of refrigerant, press   
					the START key to accept the default evacuation time of 15:00 minutes, or   
					enter the desired vacuum time by using the number keys. Then press the   
					START key.   
					6. The unit automatically charges the A/C system after the specified vacuum   
					time has elapsed.   
					7. Advance to Step 4 of RECHARGING the A/C SYSTEM in this manual to   
					complete the charging process.   
					VACUUM   
					1. Press the VACUUM key.   
					2. Ensure the service hoses are connected and panel valves are in the correct   
					position. Press START.   
					3. Press the START key to accept the default evacuation time of 15:00 minutes,   
					or enter the desired vacuum time by using the number keys, and press the   
					START key.   
					4. The unit evacuates the A/C system and stops when the specified time has   
					elapsed. Pressing the STOP key will pause the process. Press START to   
					resume, or STOP again to exit.   
					5. You are now ready to replenish the A/C system oil (if necessary), or to   
					recharge the system with refrigerant.   
					15   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Operating Instructions   
					REPLENISHING A/C SYSTEM OIL   
					CAUTION! To prevent air from entering the A/C system, never let the   
					oil level drop below the pickup tube while charging or replenishing.   
					Before charging the A/C system, you must replenish any oil removed from the A/C   
					system during the recovery process. Charge only the amount of oil that was   
					removed from the A/C system during recovery. Check the oil drain bottle to   
					determine the amount of oil that was removed during recovery. Empty the oil drain   
					bottle before recovering the next A/C system to prevent an inaccurate oil charge.   
					NOTE: If no oil was removed from the A/C system during recovery, DO NOT   
					charge any oil into the A/C system.   
					THE VACUUM CYCLE MUST RUN COMPLETELY OR THE OIL INJECT   
					WILL NOT OPERATE.   
					1. Select the correct oil for the A/C system being serviced. Refer to the vehicle   
					manufacturer's service manual.   
					2. Adjust the o-ring around the oil injector bottle to the required oil charge level.   
					For example, if the bottle's oil level is at 4 ounces, and you need 1/2 ounce of   
					oil to replenish the A/C system, place the o-ring at the 3 1/2 ounce level.   
					3. Reattach the oil injector bottle to the unit.   
					4. Close the Low Side manifold valve. Put the High Side valve in the Oil Inject/   
					Charge position.   
					5. Press the F1 (INJECT OIL) button once and release it when it “beeps.” Press   
					the F1 button again and hold it until the oil level reaches the o-ring.   
					6. Press the STOP button after the oil charge is complete to recharge the A/C   
					system with refrigerant. You must recharge the AC system with refrigerant at   
					this time to ensure all of the oil is delivered.   
					6 
					0 
					2 
					5 
					0 
					7 
					0 
					0 
					0 
					3 
					5 
					0 
					0 
					0 
					0 
					8 
					2 
					0 
					0 
					0 
					2 
					0 
					0 
					0 
					0 
					4 
					0 
					5 
					0 
					0 
					1 
					0 
					0 
					3 
					3 
					6 
					0 
					2 
					5 
					0 
					5 
					1 
					0 
					0 
					3 
					Inject   
					Oil   
					Button   
					0 
					6 
					0 
					2 
					0 
					0 
					2 
					0 
					OUNCES   
					12   
					1 
					360   
					340   
					0 
					0 
					5 
					0 
					7 
					R-134A   
					11   
					10   
					9 
					320   
					300   
					280   
					260   
					240   
					220   
					200   
					180   
					160   
					140   
					120   
					100   
					80   
					0 
					R-134A   
					bar   
					0 
					0 
					3 
					1 
					8 
					R 
					0 
					0 
					0 
					0 
					0 
					0 
					0 
					1 
					VAC-CCHARGE   
					CHARGE   
					0 
					bar   
					kPa   
					8 
					3 
					RECOVER   
					8 
					VACUUM   
					8 
					D 
					5 
					kPa   
					0 
					A 
					0 
					7 
					T 
					0 
					0 
					R 
					E 
					0 
					0 
					0 
					5 
					0 
					4 
					2 
					3 
					4 
					1 
					6 
					in Hg   
					VAC   
					0 
					psi   
					5 
					psi   
					3 
					0 
					4 
					3 
					STOP   
					START   
					2 
					60   
					40   
					1 
					20   
					HIGH   
					CLOSED   
					CLOSED   
					1 
					4 
					2 
					5 
					8 
					0 
					3 
					6 
					Oil   
					F1   
					MENU   
					7 
					9 
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					Drain   
					Bottle   
					CLEAR   
					ENTER   
					Oil   
					Inject   
					Bottle   
					0 
					I 
					INST0930   
					INST0699   
					Diagram of the Oil Injection System   
					Control Panel   
					16   
					© 2001 Robinair, SPX Corporation   
				Operating Instructions   
					RECHARGING THE A/C SYSTEM   
					WARNING   
					Wear safety goggles when working with refrigerant. Use only   
					authorized refillable refrigerant tanks. Disconnect hoses with   
					extreme caution! All hoses may contain liquid refrigerant under   
					pressure. Read and follow all warnings at the beginning of this   
					manual before operating the unit.   
					IMPORTANT!   
					Evacuate the   
					A/C system for   
					at least 15   
					minutes for   
					adequate   
					moisture and   
					contaminant   
					removal.   
					1. Press the CHARGE button. (If an oil inject has been performed, the   
					CHARGE key does not have to be pressed.)   
					2. Put the Low Side Valve in the Closed position. Put the High Side Valve in   
					the Oil Inject/Charge position. Press START to continue.   
					3. Accept either the default weight by pressing START, or type in a weight   
					with the number keys, and press START.   
					4. If the weight entered will leave less than 3 lbs (1.36 kg) of refrigerant in the   
					refrigerant tank, the charge function will not start, and the display reads:   
					INSUFFICIENT REFRIG.   
					PRESS ANY KEY TO EXIT   
					See the Operating Guidelines section of the manual for refill instructions.   
					5. Upon entering a valid charge weight, the display reads:   
					CHARGE IN PROGRESS   
					CHARGED= X.XX lbs. (X.XX kg)   
					6. If, during the charge cycle, the weight fails to charge 0.05 lbs (0.02 kg) in 30   
					seconds, the unit intermittently beeps while the display alternates between:   
					CHARGING HAS SLOWED   
					PRESS START TO RETRY   
					CHARGE HAS SLOWED   
					OR STOP TO EXIT   
					7. Pressing the START button when the charging is slowed causes the charge   
					to resume. If charging does not complete, see the SLOW CHARGE   
					PROCEDURE below.   
					8. When the charge is complete the display will show   
					CHARGE COMPLETE   
					X.XXlb (kg) CHARGED   
					9. For R-134a systems, close the high and low side coupler valves. Remove the   
					service hoses from the A/C system.   
					The A/C system is now ready for use.   
					17   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Operating Instructions   
					SLOW CHARGE PROCEDURE   
					WARNING   
					Before starting the vehicle's engine, verify that it is in PARK or   
					NEUTRAL, with the emergency brake ON.   
					Never run a vehicle without adequate ventilation in the work area.   
					CAUTION! Close the high side manifold valve before starting the   
					vehicle A/C system.   
					1. Close the High Side Valve. Put the Low Side Valve in the Recover/Vacuum   
					position.   
					2. Start the vehicle, and set the AC system to its maximum setting.   
					3. Press START. The unit charges out of the low side inlet only, allowing the   
					vehicle's compressor to pull the refrigerant into the A/C system.   
					4. When the unit is finished charging, the display reads:   
					CHARGE COMPLETE   
					X.XX lb (kg) CHARGED   
					5. Close the LOW SIDE manifold valve.   
					6. Turn off the vehicle's engine.   
					7. For R-134a systems, close the high and low side coupler valves. Remove the   
					service hoses from the A/C system.   
					The A/C system is now ready for use.   
					18   
					© 2001 Robinair, SPX Corporation   
				Maintenance Instructions   
					REPLACING THE FILTER-DRIER   
					Order part no. 34724 for a replacement filter-drier. The filter-   
					drier on this unit is designed to trap acid and particulates, and   
					is formulated to remove water from the refrigerant. You must   
					change the filter-drier to ensure adequate moisture and   
					contaminant removal.   
					Typically, you can recycle up to 150 pounds (68 kilograms)   
					of refrigerant between filter changes.   
					CAUTION! For best results, use Robinair filter-   
					driers (part no. 34724). All performance tests and   
					claims are based on using this specially blended   
					filter-drier. Use of another may affect performance   
					results.   
					1. Press the MENU button.   
					INST0478   
					2. Scroll through the menu to CHANGE FILTER and   
					Filter-Drier   
					press START.   
					3. Press START again, and the unit will begin clearing the   
					filter.   
					4. When clearing is complete, the display reads:   
					TURN UNIT OFF AND REPLACE FILTER   
					5. Turn off the main power, and unplug the machine.   
					MAIN POWER BREAKER   
					CONTROL BREAKER   
					6. Open the unit door, and replace the old filter with   
					the new filter.   
					OUNCES   
					12   
					MLS   
					340   
					320   
					300   
					280   
					260   
					240   
					220   
					200   
					180   
					160   
					140   
					120   
					100   
					80   
					11   
					10   
					9 
					8 
					7. Close the unit door, plug in the machine, and turn   
					on the Main Power.   
					7 
					6 
					5 
					4 
					3 
					2 
					60   
					40   
					1 
					20   
					The filter change is now complete.   
					Filter   
					Drier   
					INST0948   
					Location of the Filter-Drier   
					19   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Maintenance Instructions   
					CHANGING THE VACUUM PUMP OIL   
					For maximum vacuum pump performance, change the vacuum pump oil after   
					every 10 hours of operation.   
					1. Turn on the MAIN POWER switch.   
					2. Press the MENU button.   
					IMPORTANT!   
					Review current   
					local, state, and   
					federal statutes,   
					cases, laws, and   
					regulations to   
					determine the   
					current status and   
					appropriate   
					disposal method   
					for pump oil. It is   
					the responsibility   
					of the user to   
					NOTE: Do not connect the service hoses to a vehicle.   
					3. Use the arrow keys to select CHANGE VACUUM PUMP OIL, and press   
					START.   
					4. Press START again to begin.   
					5. The vacuum pump will run for two minutes. Allow the vacuum pump to run   
					until it automatically stops.   
					determine if a   
					material is a   
					hazardous waste   
					at the time of   
					disposal. Ensure   
					that you are in   
					compliance with   
					all applicable laws   
					and regulations.   
					6. Remove the black plastic plug on the oil fill port of the vacuum pump.   
					7. Remove the oil drain cap from the vacuum pump, and drain the oil into a   
					suitable container for correct disposal.   
					8. Replace the oil drain cap.   
					9. Attach the flexible tube and cap to the oil bottle, and pour eight ounces of   
					vacuum pump oil into the fill port.   
					10. Press the START key. While the vacuum pump is   
					running, slowly add oil until the level rises to the center of   
					the reservoir's sight glass.   
					1 
					11. Press the STOP key and replace the black plastic plug on   
					VAC-CHARGE   
					CHARGE   
					RECOVER   
					VACUUM   
					START   
					STOP   
					LOW   
					HIGH   
					CLOSED   
					CLOSED   
					1 
					4 
					7 
					2 
					5 
					8 
					0 
					3 
					6 
					9 
					F1   
					MENU   
					the fill port.   
					RECOVER   
					VACUUM   
					RECOVER   
					VACUUM   
					OIL INJECT   
					CHARGE   
					CLEAR   
					ENTER   
					12. The unit is now ready to operate.   
					CLOSE   
					OUNCES   
					12   
					MLS   
					340   
					320   
					300   
					280   
					260   
					240   
					220   
					200   
					180   
					160   
					140   
					120   
					100   
					80   
					11   
					10   
					9 
					2 
					8 
					7 
					6 
					5 
					4 
					3 
					2 
					60   
					40   
					1 
					20   
					3 
					FLOW   
					1. Oil Fill   
					2. Sight Glass   
					3. Oil Drain   
					INST0947   
					20   
					© 2001 Robinair, SPX Corporation   
				Maintenance Instructions   
					1 
					6 
					2 
					3 
					4 
					OPEN   
					CLOSE   
					5 
					Diagram of Vacuum Pump   
					1. Oil Filler Tube   
					2. Pump Exhaust   
					3. Oil Fill Port   
					4. Sight Glass   
					5. Oil Drain Fitting   
					6. Inlet   
					CHECKING FOR LEAKS   
					Every three months, or as specified by local or state laws, you should check the   
					unit for leaks.   
					IMPORTANT!   
					Inspect the unit   
					periodically for   
					leaks. The   
					manufacturer   
					does not   
					1. Turn off the MAIN POWER switch, and disconnect the power cord from the   
					outlet.   
					2. Open the rear door. Remove the top cover and front panel.   
					reimburse for   
					lost refrigerant.   
					3. Use a leak detector to probe all connections for refrigerant leaks. Tighten   
					fittings if a leak is indicated.   
					4. Reassemble the body panels and close the rear door.   
					21   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Replacement Parts List   
					ELECTRICAL PROTECTION   
					The unit is equipped with two circuit breakers. If either circuit breaker trips, the   
					unit will not function correctly and may lose all power. Press the circuit breaker   
					button to reset. The circuit breakers are on the back of the unit.   
					GENERAL MAINTENANCE   
					1. On a regular basis, wipe off the unit with a clean cloth to remove grease, dust,   
					or other dirt.   
					2. Periodically check the internal components for leaks—over time, fittings can   
					loosen as the unit is moved. Open the unit door panel, and trace lines with a   
					leak detector. Also, check connections on the back of the unit. Tighten any   
					loose fittings or connections you may find.   
					REPLACEMENT PARTS LIST   
					The following is a list of replacement parts and accessories you may need to   
					service or maintain your unit.   
					We suggest you keep several filter-driers on hand so you will always be able to   
					change them and complete any recycling job that is in progress.   
					Premium High Vacuum Pump Oil is also available in handy quart containers, or in   
					convenient gallon containers:   
					Quart (shipped 12 quarts per case) No. 13203   
					Gallon (shipped 4 gallons per case) No. 13204   
					Because of ongoing product improvements,   
					we reserve the right to change design, specifications,   
					and materials without notice.   
					22   
					© 2001 Robinair, SPX Corporation   
				Replacement Parts List   
					34700-2K   
					17700-2K   
					R-12   
					Replacement   
					Part Number   
					R-134a   
					Replacement   
					Part Number   
					Component   
					96" Red Hose   
					63096   
					62096   
					68396A   
					68296A   
					96" Blue Hose   
					Fan   
					RA17416   
					34724   
					RA17416   
					34724   
					Filter-Drier   
					Compressor   
					RA19458   
					RA15425   
					RA19427   
					RA40994   
					RA18752   
					RA19429   
					RA19592   
					RA17419   
					RA19603   
					RA19768   
					RA19742   
					RA19741   
					18190A   
					RA19458   
					RA15425   
					RA19427   
					RA40994   
					RA18752   
					RA19429   
					RA19592   
					RA17419   
					RA19603   
					RA19768   
					RA19742   
					RA19741   
					18190A   
					Vacuum Pump   
					High Pressure Switch   
					Main Power Switch   
					Vacuum Switch   
					Pump Protection Switch   
					Automatic Expansion Valve   
					Oil Catch Bottle   
					Scale Assembly   
					Control Module   
					High Side Gauge   
					Low Side Gauge   
					Low Side Coupler   
					High Side Coupler   
					Automatic Air Purge   
					Solenoid Rebuild Kit   
					Castors   
					18191A   
					18191A   
					RA19743   
					RA19258   
					RA19631   
					RA19612   
					RA19744   
					RA19258   
					RA19631   
					RA19612   
					ISV (Internal Storage Vessel)   
					23   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Flow Diagram   
					I 
					24   
					© 2001 Robinair, SPX Corporation   
				Wiring Diagram   
					25   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Limited Warranty   
					Robinair Limited Warranty Statement   
					Rev. July 11, 2003   
					This product is warranted to be free from defects in workmanship, materials, and   
					components for a period of one year from date of purchase. All parts and labor required to   
					repair defective products covered under the warranty will be at no charge. The following   
					restrictions apply:   
					1. The limited warranty applies to the original purchaser only.   
					2. The warranty applies to the product in normal usage situations only, as described in the   
					Operating Manual. The product must also be serviced and maintained as specified.   
					3. If the product fails, it will be repaired or replaced at the option of the manufacturer.   
					4. Transportation charges for warranty service will be reimbursed by the factory upon   
					verification of the warranty claim and submission of a freight bill for normal ground   
					service. Approval from the manufacturer must be obtained prior to shipping to an   
					authorized service center.   
					5. Warranty service claims are subject to authorized inspection for product defect(s).   
					6. The manufacturer shall not be responsible for any additional costs associated with a   
					product failure including, but not limited to, loss of work time, loss of refrigerant,   
					cross-contamination of refrigerant, and unauthorized shipping and/or labor charges.   
					7. All warranty service claims must be made within the specified warranty period. Proof-   
					of-purchase date must be supplied to the manufacturer.   
					8. Use of recovery/recycling equipment with unauthorized refrigerants, sealants, or dyes   
					will void the warranty.   
					• Authorized refrigerants are listed on the equipment or are available through the   
					Technical Service Department.   
					• The manufacturer prohibits the use of the recovery/recycling equipment on air   
					conditioning (A/C) systems containing leak sealants, either of a seal-swelling or   
					aerobic nature.   
					• The manufacturer prohibits the use of dyes injected through the oil injection device   
					on the recovery/recycling equipment.   
					This Limited Warranty does NOT apply if:   
					• The product, or product part, is broken by accident.   
					• The product is misused, tampered with, or modified.   
					• The product is used for recovering or recycling any substance other than the specified   
					refrigerant type. This includes, but is not limited to, materials and chemicals used to seal   
					leaks in A/C systems.   
					• The product is equipped with an oil injection device that has been used to inject dye.   
					The manufacturer only endorses the use of separate dye injection devices, and does not   
					support the use of the oil injection feature for this purpose.   
					Note: Refillable refrigerant tanks are reusable.   
					26   
					© 2001 Robinair, SPX Corporation   
				Notes   
					27   
					34700-2K/17700-2K Cool-Tech Recovery/Recycling/Recharging Unit   
				Notes   
					28   
					© 2001 Robinair, SPX Corporation   
				CONVERSION   
					TABLE   
					Visit our web site at   
					
					or   
					OZ.   
					LBS.   
					0.5   
					1.0   
					1.5   
					2.0   
					2.5   
					3.0   
					3.5   
					4.0   
					4.5   
					5.0   
					5.5   
					6.0   
					6.5   
					7.0   
					7.5   
					8.0   
					8.5   
					0.03   
					0.06   
					0.09   
					0.13   
					0.16   
					0.19   
					0.22   
					0.25   
					0.28   
					0.31   
					0.34   
					0.38   
					0.41   
					0.44   
					0.47   
					0.50   
					0.53   
					0.56   
					0.59   
					0.63   
					0.69   
					0.69   
					0.72   
					0.75   
					0.78   
					0.81   
					0.84   
					0.88   
					0.91   
					0.94   
					0.97   
					1 lb.   
					% 
					Call our Toll-Free   
					Technical Support Line at   
					800-822-5561   
					in the continental U.S. or Canada.   
					In all other locations, contact your local distributor. To help us serve   
					you better, please be prepared to provide the model number, serial   
					number, and date of purchase of your unit.   
					To validate your warranty, you must complete the warranty card   
					attached to your unit, and return it within ten days from date of   
					purchase.   
					9.0   
					9.5   
					NATIONWIDE NETWORK OF AUTHORIZED SERVICE CENTERS   
					10.0   
					10.5   
					11.0   
					11.5   
					12.0   
					12.5   
					13.0   
					13.5   
					14.0   
					14.5   
					15.0   
					15.5   
					16.0   
					If your unit needs repair or replacement parts, contact the service   
					center in your area. For help in locating a service center, call the   
					toll-free technical support line.   
					Due to ongoing product improvements,   
					we reserve the right to change design,   
					specifications, and materials without notice.   
					The347002K/177002Kisdesignedtomeetallapplicableagencycertifications,including   
					Underwriter'sLaboratories, Inc., SAEStandards, andCUL. Correctmaintenanceofthis   
					equipment will provide accurate A/C service for many years.   
					Certain state and local jurisdictions dictate that using this equipment to sell refrigerant   
					by weight may not be permitted. We recommend charging for any A/C service by the   
					job performed.   
					This weight scale provides a means of metering the amount of refrigerant needed for   
					optimum A/C system performance as recommended by OEM manufacturers.   
					SPX Corporation   
					655 Eisenhower Drive   
					Owatonna, MN 55060-0995 USA   
					Tech Services:1-800-822-5561   
					Fax: 1-800-822-7805   
					Customer Service: 1-800-533-6127   
					Fax: 1-800-322-2890   
					
					124080 (Rev. C, 7-09-03)   
					347002K/177002K Cool-Tech   
					© SPX Corporation   
				 |