| USER MANUAL   HELIO2   HELIUM-OxygEn BLEndEr   (DISS and NIST Connections)   Model No. PM5400 Series   PM5500 Series   (shown)   SAVE THESE INSTRUCTIONS   Federal (USA) law restricts this device to sale   by or on the order of a physician.   CAUTION   300 Held Drive   Tel: (+001) 610-262-6090   Northampton, PA 18067 USA   Fax: (+001) 610-262-6080   REAd ALL INSTRUCTIONS BEFORE USING   This manual instructs a Professional to install and operate the HELIO2 Blender.   This is provided for your safety and to prevent damage to the HELIO2 Blender.   If you do not understand this manual, DO NOT USE the HELIO2 Blender and   contact your Provider.   dANGER   This product is not intended as a life-sustaining or life-supporting device.   ExPLANATION OF ABBREVIATIONS   FIHe-O2   Fractional Concentration of Inspired Helium-Oxygen   Fractional Concentration of Inspired Oxygen   Helium-Oxygen   FIO2   Heliox   DISS   NIST   psi   Diameter Indexed Safety System   Non-Interchangeable Screw Thread   Pounds Per Square Inch   lpm   Liters Per Minute   SAFETY INFORMATION - WARNINGS ANd CAUTIONS   Indicates an imminently hazardous situation which,   if not avoided, will result in death or serious injury.   dANGER   Indicates a potentially hazardous situation which, if   not avoided, could result in death or serious injury.   WARNING   Indicates a potentially hazardous situation which, if   not avoided, may result in minor or moderate injury.   CAUTION   Used without the safety alert symbol indicates a   potentially hazardous situation which, if not avoided,   may result in property damage.   CAUTION   CONSULT ACCOMPANYING DOCUMENTS   Symbol for “USE NO OIL”   Symbol indicates the device complies with the   requirements of Directive 93/42/EEC concerning   medical devices and all applicable International   Standards. (On CE marked devices ONLY)   Prescription Required   HELIO2   BLENDEr   HELIUM-OxygEn   2 WARNING   • Only trained, qualified medical personnel under the direct supervision of a   licensed physician should operate the HELIO2 Blender.   • Use this HELIO2 Blender only for its Intended Use as described in this   manual.   • Confirm prescribed dose before administering to patient. Monitor on a   frequent basis. Continuous Monitoring with an alarmed Oxygen Monitor   / Analyzer is recommended.   • • The HELIO2 Blender shall be serviced by a qualified service technician.   Always follow ANSI and CGA standards for Medical Gas Products,   Flowmeters and Oxygen Handling.   • The clinician must verify the Heliox tank concentration prior to utilizing   the Blender.   • An alarmed Oxygen Monitor / Analyzer must be used to verify oxygen   concentration.   • Accuracy of oxygen concentration will be affected if bleed is not activated   at flow settings below 15 lpm for the High Flow Blender, and 3 lpm for   the Low Flow Blender.   • dO NOT obstruct the alarm.   • dO NOT use Blender when alarm is sounding.   • dO NOT use oil in or around the Blender.   • dO NOT occlude or obstruct the bleed port on the auxiliary outlet of   the Blender.   • dO NOT use near any type of flame or flammable/explosive   substances, vapors or atmosphere.   • Oxygen Concentration dial does not rotate 360 degrees. Rotating the   dial beyond the endpoint settings will damage the Blender.   CAUTION   • Turn off gas supplies when HELIO2 Blender is not in use.   • Store the HELIO2 Blender in a clean, dry area when not in use.   • The HELIO2 Blender contains magnetic, ferrous material that may   affect the results of an MRI.   • Ensure all connections are tight and leak free.   • Avoid excessive pressure surges greater than 100 psi (6.9 bar) when   pressuring the Blender inlets.   • DO NOT steam autoclave.   • DO NOT immerse HELIO2 Blender into any liquid.   • DO NOT gas sterilize with (EtO) Ethylene Oxide.   • DO NOT use if dirt or contaminants are present on or around the   Blender or connecting devices.   • DO NOT smoke in an area where oxygen is being administered.   • DO NOT clean with aromatic hydrocarbons.   • Inlet pressure of Device used in conjunction with the HELIO2 Blender must   match inlet pressure of FIO2 Blender.   HELIO2   BLENDEr   HELIUM-OxygEn   3 SPECIFICATIONS   High Flow   Low Flow   Model   PM5580: 80/20 heliox PM5480: 80/20 heliox   PM5570: 70/30 heliox PM5470: 70/30 heliox   15 - 120 lpm   3 - 30 lpm   Primary Outlet   Flow Range   With both supply pressures at 50 psi (3.4 bar) with   BLEED Closed   2 - 100 lpm   0 - 30 lpm   Auxiliary Outlet   Flow Range   With both supply pressures at 50 psi (3.4 bar) with   BLEED Open   13 lpm or less   3 lpm or less   Bleed Flow @   at 50 psi (3.4 bar)   at 50 psi (3.4 bar)   100% FIO2   Maximum   Combined   ≥ 120 lpm   ≥ 30 lpm   Flow (All Outlets)   Bypass Flow   (Loss of Helium or   Oxygen supply)   > 85 lpm   > 45 lpm   Bypass Alarm   Activation   50 psi   (3.45 bar)   60 psi   (4.14 bar) (3.45 bar)   50 psi   60 psi   (4.14 bar)   13-25 psi   16-24 psi   18-22 psi 16-24 psi   0.9-1.7 bar 1.1-1.65 bar 1.2-1.5 bar 1.1-1.65 bar   When pressure differential is 6 psi (0.4 bar)   or less.   Alarm Reset:   Alarm Sound Level:   ≥ to 80 db at 1 ft (0.3 m)   Oxygen Concentration   Adjustment Range:   PM5480   PM5580   20 -   100%   PM5470   PM5570   30 - 100%   30 - 75 psi (2.1 - 5.2 bar) Heliox & Oxygen   within 10 psi (0.69 bar) of each other   Gas Supply Pressure:   Mixed Gas Stability:   ±1% Oxygen   Connection Types:   DISS Type - Heliox & Oxygen Inlets & Outlets   and / or NIST Type - Heliox & Oxygen Inlets   Note: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).   HELIO2   BLENDEr   HELIUM-OxygEn   4 SPECIFICATIONS continued   dimensions: (without fittings)   (12.5 cm)   (5.7 cm)   (10.4 cm)   (1.3 kg)   Depth:   4.9 in   2.3 in   4.1 in   2.9 lbs   3.5 lbs   Width:   Height:   Weight:   (1.6 kg)   Shipping Weight:   59°F to 104°F (15°C to 40°C)   Operating Temperature Range:   Transport / Storage Requirements   -10°F to 140°F (-23°C to 60°C)   Temperature Range:   Humidity:   Max 95% Noncondensing   ± 3% of full scale @ 50 psi (3.4 bar)   FIO2 Accuracy:*   Pressure drop:   Low Flow:   ≤ 2 psi (0.14 bar) at inlet pressures from 30-90 psi   (2.1- 6.2 bar) and at 10 lpm flow rate at 60% FIO2   . High Flow: ≤ 3 psi (0.21 bar) at inlet pressures from 30-90 psi   (2.1- 6.2 bar) and at 30 lpm flow rate at 60% FIO2   . The Helium-Oxygen Blender has been cleaned for Oxygen Service prior to delivery.   The Helium-Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195.   The alarmed Oxygen Monitor / Analyzer should comply with ISO 7767 to   meet CE requirement.   dryness and Composition for inlet gases:   Heliox:   Medical grade 80/20 or 70/30 is required.   Oxygen:   Oxygen supply must meet all requirements of USP   Medical Oxygen.   dew Point: Both inlets should remain 10°F (-12.2°C) or more   below the lowest temperature to which the air   distribution system equipment is exposed. At a   temperature of 25°F (-3.9°C) and a pressure of   90 psi (6.33 kg/cm2) this equates to 2000 mg/m3.   (ONLY for CE   requirements)   * Accuracy of FIO2 will be affected if bleed flow is not engaged at low flows. (At or below 3 lpm for Low Flow and 15 lpm for High Flow)   . * When Heliox tank pressure and oxygen outlet pressures are unbalanced, bleed may need to be engaged at a higher liter flow   to maintain accuracy.   Specifications are subject to change without prior notice.   HELIO2   BLENDEr   HELIUM-OxygEn   5 dIAGRAMS   CAUTION   Missing or illegible labels must be replaced, contact   Precision Medical, Inc.   PM5400 and PM5500 Models   B A E F H G D C HELIO2   BLENDEr   HELIUM-OxygEn   6 COMPONENT dESCRIPTION   ITEM   DESCRIPTION   Oxygen Concentration dial   A Adial used for selecting oxygen concentrations between 20%   -100% or 30% -100%. The FIO2 scale is used for reference   only. The actual FIO2 must be verified with an Alarmed   Oxygen Monitor / Analyzer.   This dial does not rotate 360°. The dial starts at 20% or   30% and ends at 100%.   Primary Outlet Port   A male DISS oxygen fitting with check valve that delivers flow   when engaged to any controlling device, such as a flowmeter.   B C Auxiliary Outlet Port   Amale DISS oxygen fitting with check valve that delivers flow   when engaged to any controlling device, such as a flowmeter.   This outlet is equipped with a bleed valve that allows the user   to control if the bleed is ON or OFF. With the bleed in the “ON”   position, this outlet delivers accurate oxygen concentrations   in the following flows:   Model   High Flow   Low Flow   Flow Range   2 - 100 lpm   0 - 30 lpm   Auxiliary Bleed Collar   d The collar is used to engage and disengage the bleed. The   bleed is necessary to maintain accurate FIO2 Concentration   below 15 lpm for the High Flow and ≤ 3 lpm for the Low Flow.   To activate the bleed, slide and rotate the knurled collar back   until it contacts the cover. To deactivate the bleed, pull and   rotate collar away from cover until bleed flow valve is closed.   Oxygen Inlet Fitting   Afemale DISS or NIST oxygen fitting with one way valve that   is used to connect an oxygen supply hose.   E F Heliox Inlet Fitting   A male DISS or NIST heliox fitting with one way valve that   is used to connect a heliox supply hose.   Alarm   G An audible alarm that sounds due to an excessive   pressure drop or deletion of either gas supply.   Rear Slide Mount with dove tail.   H HELIO2   BLENDEr   HELIUM-OxygEn   7 PRE-USE TESTING   WARNING   • Read this User Manual before installing or operating the HELIO2 Blender.   • Confirm the concentration of heliox with an alarmed Oxygen Monitor /   Analyzer.   • Confirm contents of heliox cylinder prior to use.   CAUTION   Inspect the HELIO2 Blender for visual damage before use, DO NOT   USE if damaged.   NOTE: The tests listed below should be performed prior to placing the   HELIO2 Blender in service.   Pre-Use Testing consists of:   • Alarm Test   • Reverse Gas Flow Procedure   1. Secure the HELIO2 Blender to a wall or pole bracket in an upright position.   2. Connect the heliox and oxygen supply lines to the appropriate inlet fittings   on the bottom of the HELIO2 Blender.   3. Attach a flowmeter, or other metering device to one of the outlet ports and   verify FIO2 range for accuracy with an alarmed Oxygen Monitor / Analyzer.   Primary Outlets Flow capacity:   • High Flow Blender (PM 5500 Model) 15 lpm to 120 lpm   • Low Flow Blender (PM 5400 Model) 3 lpm to 30 lpm   Auxiliary Outlet Use:   The auxiliary flow outlet maintains the same flow capacity and FIO2 accuracy   as the Primary Outlets with Bleed Valve not engaged. When the bleed flow is   activated, some of the heliox mixture will vent to atmosphere to maintain FIO2   concentration accuracy at the low flow settings.   • High Flow Blender (PM 5500 Model) 15 lpm or less   • Low Flow Blender (PM 5400 Model) 3 lpm or less   4. Attach a supply line to the outlet port of the flowmeter.   ALARM TEST   1. Connect the HELIO2 Blender to respective heliox and oxygen sources,   pressurize the Blender and turn “ON” the flowmeter.   2. Set Oxygen Concentration Dial to 50% FIO2.   3. Disconnect or turn “OFF” the heliox supply line to the HELIO2 Blender. The   Blender should alarm with a loud whistle noise. The whistle indicates the   alarm is operating correctly.   4. Reconnect and activate the heliox supply line to the Blender, the alarm   should stop whistling.   5. Disconnect or turn “OFF” the oxygen supply line to the HELIO2 Blender.   The Blender should alarm with a loud whistle noise. The whistle indicates   the alarm is operating correctly.   6. Reconnect and activate the oxygen supply line to the Blender, the alarm   should stop whistling.   HELIO2   7. If alarm fails to function properly, DO NOT USE.   BLENDEr   HELIUM-OxygEn   8 REVERSE GAS FLOW PROCEdURE   (CE Requirements ONLY)   1. Assure bleed flow valve is not engaged. Disconnect the oxygen hose   from the gas source. Remove all outlet connections from the HELIO2   Blender to ensure that there is no outlet flow.   2. Place the free end of the oxygen supply hose under water. While   gradually increasing the heliox supply pressure from 30-75 psi   (2.07-5.17 bar) check for leakage past the oxygen inlet check valve.   3. Replace the Duckbill Check Valve in the oxygen inlet, if leakage is   >100 ml/min. Reference HELIO2 Blender Service Manual (P/N 506124).   4. Repeat steps 1-3 to check for leakage past the heliox inlet check   valve.   OPERATING INSTRUCTIONS   CAUTION   Inspect the HELIO2 Blender for visual damage before use,   DO NOT USE if damaged.   1. Secure Blender to wall or pole mount bracket.   2. Connect heliox and oxygen supply lines from HELIO2 Blender to   heliox cylinder and oxygen supply.   3. Connect oxygen flowmeter to Blender outlet.   4. Adjust the Oxygen Concentration Dial to the prescribed   concentration. The balance of the concentration will be helium   exiting the flowmeter.   NOTE: The Oxygen Concentration Dial does not rotate 360°. dO   NOT force dial beyond the oxygen concentration endpoints   as this will damage the Blender.   5. Confirm the flow of heliox mixture to the patient.   6. Actual heliox concentration to the patient may vary due to   entrainment of room air via the patient interface device.   CAUTION   • Refer to “Oxygen Flowmeter Conversion Chart” for corrected   heliox flows.   • An oxygen flowmeter should be used on the outlets of the HELIO2   Blender along with the corresponding flow conversion chart.   • Actual flow from an oxygen flowmeter utilized to deliver heliox   are higher than read on the flowmeter.   HELIO2   BLENDEr   HELIUM-OxygEn   9 7. Confirm the concentration of heliox with an Oxygen Monitor /   Analyzer. If necessary activate the bleed flow valve to maintain   FIO2 accuracy.   8. To activate the bleed, turn and rotate the knurled collar back until it   contacts the cover.   9. To deactivate the bleed, pull and rotate the collar away from the   cover until bleed flow valve is closed.   10. Turn “OFF” the heliox and oxygen supply or disconnect when the   HELIO2 Blender is not in use.   CLEANING   CAUTION   • dO NOT steam autoclave.   • dO NOT immerse the HELIO2 Blender into any liquid.   • dO NOT use any strong solvent or abrasive cleaners.   • dO NOT gas sterilize with (EtO) Ethylene Oxide.   • dO NOT clean with aromatic hydrocarbons.   1. Disconnect all gas connections and equipment before cleaning.   2. Clean exterior surfaces with a cloth dampened with mild detergent   and water.   3. Wipe dry with a clean cloth.   MAINTENANCE   The following maintenance on the HELIO2 Blender must be performed   by a trained service technician:   • The alarm should be tested prior to being placed into clinical   service and periodically there after.   • Every year conduct the Operational Verification Procedure (OVP).   * A detailed description of the OVP tests can be found in the   Blender Service Manual (P/N 506124), and available on the   • Every 2 years the HELIO2 Blender should be serviced.   PM5400 (P/N 506125)   PM5500 (P/N 506212)   • Refer to the HELIO2 Blender Service Manual (P/N 506124) for   complete details regarding further maintenance and testing.   HELIO2   BLENDEr   HELIUM-OxygEn   10   OxYGEN FLOWMETER CONVERSIONS   Oxygen   Flowmeter   Corrected Heliox Flow (LPM) at Various FIO2 Settings   Setting   20%   1.8   30%   1.6   40%   1.4   50%   1.3   60%   1.2   70%   1.18   2.4   80%   1.15   2.3   90%   1.02   2.0   100%   1.0   1 2 3.6   3.2   2.8   2.6   2.4   2.0   3 5.4   4.8   4.2   3.9   3.6   3.5   3.5   3.1   3.0   4 7.2   6.4   5.6   5.2   4.8   4.7   4.6   4.1   4.0   5 9.0   8.0   7.0   6.5   6.0   5.9   5.8   5.1   5.0   6 10.8   12.6   14.4   16.2   18.0   19.8   21.6   23.4   25.2   27.0   28.8   30.6   32.4   34.2   36.0   37.8   39.6   41.4   43.2   45.0   46.8   48.6   50.4   52.2   54.0   55.8   57.6   59.4   61.2   63.0   64.8   66.6   68.4   70.2   72.0   9.6   8.4   7.8   7.2   7.1   6.9   6.1   6.0   7 11.2   12.8   14.4   16.0   17.6   19.2   20.8   22.4   24.0   25.6   27.2   28.8   30.4   32.0   33.6   35.2   36.8   38.4   40.0   41.6   43.2   44.8   46.4   48.0   49.6   51.2   52.8   54.4   56.0   57.6   59.2   60.8   62.4   64.0   9.8   9.1   8.4   8.3   8.1   7.1   7.0   8 11.2   12.6   14.0   15.4   16.8   18.2   19.6   21.0   22.4   23.8   25.2   26.6   28.0   29.4   30.8   32.2   33.6   35.0   36.4   37.8   39.2   40.6   42.0   43.4   44.8   46.2   47.6   49.0   50.4   51.8   53.2   54.6   56.0   10.4   11.7   13.0   14.3   15.6   16.9   18.2   19.5   20.8   22.1   23.4   24.7   26.0   27.3   28.6   29.9   31.2   32.5   33.8   35.1   36.4   37.7   39.0   40.3   41.6   42.9   44.2   45.5   46.8   48.1   49.4   50.7   52.0   9.6   9.4   9.2   8.2   8.0   9 10.8   12.0   13.2   14.4   15.6   16.8   18.0   19.2   20.4   21.6   22.8   24.0   25.2   26.4   27.6   28.8   30.0   31.2   32.4   33.6   34.8   36.0   37.2   38.4   39.6   40.8   42.0   43.2   44.4   45.6   46.8   48.0   10.6   11.8   13.0   14.2   15.3   16.5   17.7   18.9   20.1   21.2   22.4   23.6   24.8   26.0   27.1   28.3   29.5   30.7   31.9   33.0   34.2   35.4   36.6   37.8   38.9   40.1   41.3   42.5   43.7   44.8   46.0   47.2   10.4   11.5   12.7   13.8   15.0   16.1   17.3   18.4   19.6   20.7   21.9   23.0   24.2   25.3   26.5   27.6   28.8   29.9   31.1   32.2   33.4   34.5   35.7   36.8   38.0   39.1   40.3   41.4   42.6   43.7   44.9   46.0   9.2   9.0   10   11   12   13   14   15   16   17   18   19   20   21   22   23   24   25   26   27   28   29   30   31   32   33   34   35   36   37   38   39   40   10.2   11.2   12.2   13.3   14.3   15.3   16.3   17.3   18.4   19.4   20.4   21.4   22.4   23.5   24.5   25.5   26.5   27.5   28.6   29.6   30.6   31.6   32.6   33.7   34.7   35.7   36.7   37.7   38.8   39.8   40.8   10.0   11.0   12.0   13.0   14.0   15.0   16.0   17.0   18.0   19.0   20.0   21.0   22.0   23.0   24.0   25.0   26.0   27.0   28.0   29.0   30.0   31.0   32.0   33.0   34.0   35.0   36.0   37.0   38.0   39.0   40.0   HELIO2   BLENDEr   HELIUM-OxygEn   11   TECHNICAL dESCRIPTION   For a complete Technical Description of the HELIO2 Blender and   list of Replacement Parts, reference the HELIO2 Blender Service   Manual (P/N 506124) available on the Internet;   RETURNS   Returned products require a Returned Goods Authorization   (RGA) number, contact Precision Medical, Inc. All returns must   be packaged in sealed containers to prevent damage. Precision   Medical, Inc. will not be responsible for goods damaged in transit.   Refer to Precision Medical, Inc. Return Policy available on the   dISPOSAL INSTRUCTIONS   This device and its packaging contain no hazardous materials.   No special precautions need to be taken when disposing the   device and/or its packaging.   Please Recycle   HELIO2   BLENDEr   HELIUM-OxygEn   12   TROUBLESHOOTING   If the HELIO2 Blender fails to function, consult the Troubleshooting   Guide below.   If problem cannot be solved by using Troubleshooting Guide, refer to the   HELIO2 Blender Service Manual (P/N 506124) available on the Internet;   Problem   Probable Cause   Remedy   1. •HIgH flow model,   flow requirement   below 15 lpm.   1. Use auxiliary outlet &   engage bleed   Oxygen   concentration   discrepancy   between Blender   setting and   alarmed Oxygen   Monitor / Analyzer   (greater than 3%)   • LOW flow model,   flow requirement   below 3 lpm.   2. Recalibrate alarmed   Oxygen Monitor /   Analyzer or Verify   with second alarmed   Oxygen Monitor /   Analyzer   2. Alarmed Oxygen   Monitor / Analyzer   inaccurate   3. Remove obstruction   3. Low flow bleed   obstructed   4. Check gas sources   with calibrated alarmed   Oxygen Monitor /   4. Gas supply   contaminated or   heliox concentration   incorrect   Analyzer to confirm   oxygen is 100% and   verify heliox tank content   5. Isolate Blender. Check   oxygen concentration at   Blender Outlets   6. Assure heliox and   oxygen inlets pressures   are within 10 psi   5. Downstream device   causing back flow   or restricted flow   6. Supply pressure   imbalanced   1. Gas sources turned 1. Turn gas sources “ON”   “OFF”   No flow at   Blender outlets   2. Gas sources not   connected   2. Connect gas sources   1. Difference between   oxygen and heliox   inlet pressures   1. Correct pressure   Alarm sounding   difference until heliox   and oxygen pressures   are within specification   greater than specified   HELIO2   BLENDEr   HELIUM-OxygEn   13   LIMITEd WARRANTY   ANd   LIMITATION OF LIABILITY   PrecisionMedical,Inc.warrantsthattheHeliO2Blender,(theProduct),will   be free of defects in workmanship and/or material for the following period:   Two (2) years from shipment   Should any failure to conform to this warranty appear within the applicable   period, Precision Medical, Inc. shall, upon written notification thereof and   substantiation that the goods have been stored, installed, maintained and   operated in accordance with Precision Medical, Inc.’s instructions and   standard industry practice, and that no modifications, substitutions, or   alterations have been made to the goods, correct such defect by suitable   repair or replacement at its own expense.   ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.   The representatives of Precision Medical, Inc. or any retailers are not   authorized to make oral warranties about the merchandise described in   this contract, and any such statements shall not be relied upon and are   not part of the contract for sale. Thus, this writing is a final, complete and   exclusive statement of the terms of that contract.   THIS WARRANTYIS ExCLUSIVEAND IS IN LIEU OFANYWARRANTY   OF MERCHANTABILITY, FITNESS FORAPARTICULAR PURPOSE OR   OTHERWARRANTYOFQUALITY,WHETHERExPRESSORIMPLIED.   Precision Medical, Inc. shall not under any circumstances be liable for   special, incidental or consequential damages including but not limited to   lost profits, lost sales, or injury to person or property. Correction of non-   conformities as provided above shall constitute fulfillment of all liabilities   of Precision Medical, Inc. whether based on contract, negligence, strict   tort or otherwise. Precision Medical, Inc. reserves the right to discontinue   manufacture of any product or change product materials, designs, or   specifications without notice.   Precision Medical, Inc. reserves the right to correct clerical or   typographical errors without penalty.   HELIO2   BLENDEr   HELIUM-OxygEn   14   dECLARATION OF CONFORMITY   Manufacturer:   Precision Medical, Inc.   300 Held Drive, Northampton, PA 18067, USA   CONTACT: Quality Manager   Phone: 610-262-6090   Authorized European Representative: Emergo Europe (European Office)   Molenstraat 15   2513 BH, The Hague   The Netherlands   Phone: +31 (0) 70.345.8570   Fax: +31 (0) 70.346.7299   Product:   Gas Mixers for Medical Use (Blender)   PM5470EN, PM5480EN, PM5570EN, PM5580EN   IIb   Model(s):   MDD Class:   Classification criteria:   Clause 3.2 Rule 11 of Annex IX of MDD   As delivered, the object of the declaration described above is in conformity with the   requirements of MDD 93/42/EEC Annex II.3 and the following documents:   Document   Edition   EN 980   2008   EN 1041   2008   ISO 11198   ISO 14971   ISO 15001   1995   2007 2nd Ed   2004   ISO 7000   2004   Notified Body:   TÜV Rheinland Products Safety GmbH   EC Certificate No.: HD60019110 0001   ISO 13485 Certified   505963 Rev3 (E) 04/05/10 Printed in USA   |